Illustrated instructions for mechanical quality assurance of a medical linear accelerator

PURPOSE: The purpose of this study was to develop and test a set of illustrated instructions for effective training for mechanical quality assurance (QA) of medical linear accelerators (linac). METHODS: Illustrated instructions were created for mechanical QA and underwent several steps of review, te...

Descripción completa

Detalles Bibliográficos
Autores principales: Court, Laurence, Wang, He, Aten, David, Brown, Derek, MacGregor, Hannelie, du Toit, Monique, Chi, Melinda, Gao, Song, Yock, Adam, Aristophanous, Michalis, Balter, Peter
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978554/
https://www.ncbi.nlm.nih.gov/pubmed/29500846
http://dx.doi.org/10.1002/acm2.12265
_version_ 1783327533916225536
author Court, Laurence
Wang, He
Aten, David
Brown, Derek
MacGregor, Hannelie
du Toit, Monique
Chi, Melinda
Gao, Song
Yock, Adam
Aristophanous, Michalis
Balter, Peter
author_facet Court, Laurence
Wang, He
Aten, David
Brown, Derek
MacGregor, Hannelie
du Toit, Monique
Chi, Melinda
Gao, Song
Yock, Adam
Aristophanous, Michalis
Balter, Peter
author_sort Court, Laurence
collection PubMed
description PURPOSE: The purpose of this study was to develop and test a set of illustrated instructions for effective training for mechanical quality assurance (QA) of medical linear accelerators (linac). METHODS: Illustrated instructions were created for mechanical QA and underwent several steps of review, testing, and refinement. Eleven testers with no recent QA experience were then recruited from our radiotherapy department (one student, two computational scientists, and eight dosimetrists). This group was selected because they have experience of radiation therapy but no preconceived ideas about how to do QA. The following parameters were progressively decalibrated on a Varian C‐series linac: Group A = gantry angle, ceiling laser position, X1 jaw position, couch longitudinal position, physical graticule position (five testers); Group B = Group A + wall laser position, couch lateral and vertical position, collimator angle (three testers); Group C = Group B + couch angle, wall laser angle, and optical distance indicator (three testers). Testers were taught how to use the linac and then used the instructions to try to identify these errors. An experienced physicist observed each session, giving support on machine operation as necessary. RESULTS: Testers were able to follow the instructions. They determined gantry, collimator, and couch angle errors within 0.4°, 0.3°, and 0.9° of the actual changed values, respectively. Laser positions were determined within 1 mm and jaw positions within 2 mm. Couch position errors were determined within 2 mm and 3 mm for lateral/longitudinal and vertical errors, respectively. Accessory‐positioning errors were determined within 1 mm. Optical distance indicator errors were determined within 2 mm when comparing with distance sticks and 6 mm when using blocks, indicating that distance sticks should be the preferred approach for inexperienced staff. CONCLUSIONS: Inexperienced users were able to follow these instructions and catch errors within the criteria suggested by AAPM TG‐142 for linacs used for intensity‐modulated radiation therapy. These instructions are, therefore, suitable for QA training.
format Online
Article
Text
id pubmed-5978554
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-59785542018-06-01 Illustrated instructions for mechanical quality assurance of a medical linear accelerator Court, Laurence Wang, He Aten, David Brown, Derek MacGregor, Hannelie du Toit, Monique Chi, Melinda Gao, Song Yock, Adam Aristophanous, Michalis Balter, Peter J Appl Clin Med Phys Education PURPOSE: The purpose of this study was to develop and test a set of illustrated instructions for effective training for mechanical quality assurance (QA) of medical linear accelerators (linac). METHODS: Illustrated instructions were created for mechanical QA and underwent several steps of review, testing, and refinement. Eleven testers with no recent QA experience were then recruited from our radiotherapy department (one student, two computational scientists, and eight dosimetrists). This group was selected because they have experience of radiation therapy but no preconceived ideas about how to do QA. The following parameters were progressively decalibrated on a Varian C‐series linac: Group A = gantry angle, ceiling laser position, X1 jaw position, couch longitudinal position, physical graticule position (five testers); Group B = Group A + wall laser position, couch lateral and vertical position, collimator angle (three testers); Group C = Group B + couch angle, wall laser angle, and optical distance indicator (three testers). Testers were taught how to use the linac and then used the instructions to try to identify these errors. An experienced physicist observed each session, giving support on machine operation as necessary. RESULTS: Testers were able to follow the instructions. They determined gantry, collimator, and couch angle errors within 0.4°, 0.3°, and 0.9° of the actual changed values, respectively. Laser positions were determined within 1 mm and jaw positions within 2 mm. Couch position errors were determined within 2 mm and 3 mm for lateral/longitudinal and vertical errors, respectively. Accessory‐positioning errors were determined within 1 mm. Optical distance indicator errors were determined within 2 mm when comparing with distance sticks and 6 mm when using blocks, indicating that distance sticks should be the preferred approach for inexperienced staff. CONCLUSIONS: Inexperienced users were able to follow these instructions and catch errors within the criteria suggested by AAPM TG‐142 for linacs used for intensity‐modulated radiation therapy. These instructions are, therefore, suitable for QA training. John Wiley and Sons Inc. 2018-03-03 /pmc/articles/PMC5978554/ /pubmed/29500846 http://dx.doi.org/10.1002/acm2.12265 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Education
Court, Laurence
Wang, He
Aten, David
Brown, Derek
MacGregor, Hannelie
du Toit, Monique
Chi, Melinda
Gao, Song
Yock, Adam
Aristophanous, Michalis
Balter, Peter
Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title_full Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title_fullStr Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title_full_unstemmed Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title_short Illustrated instructions for mechanical quality assurance of a medical linear accelerator
title_sort illustrated instructions for mechanical quality assurance of a medical linear accelerator
topic Education
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978554/
https://www.ncbi.nlm.nih.gov/pubmed/29500846
http://dx.doi.org/10.1002/acm2.12265
work_keys_str_mv AT courtlaurence illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT wanghe illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT atendavid illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT brownderek illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT macgregorhannelie illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT dutoitmonique illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT chimelinda illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT gaosong illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT yockadam illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT aristophanousmichalis illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator
AT balterpeter illustratedinstructionsformechanicalqualityassuranceofamedicallinearaccelerator