Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes

PURPOSE: The target detectability of cone beam computed tomography (CBCT) performed in image‐guided radiation therapy (IGRT) was investigated to achieve sufficient image quality for patient positioning over a course of treatment session while maintaining radiation exposure from CBCT imaging as low a...

Descripción completa

Detalles Bibliográficos
Autores principales: Yang, Ching‐Ching, Yu, Pei‐Chieh, Ruan, Jau‐Ming, Chen, Yu‐Cheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978665/
https://www.ncbi.nlm.nih.gov/pubmed/29516610
http://dx.doi.org/10.1002/acm2.12306
_version_ 1783327543460364288
author Yang, Ching‐Ching
Yu, Pei‐Chieh
Ruan, Jau‐Ming
Chen, Yu‐Cheng
author_facet Yang, Ching‐Ching
Yu, Pei‐Chieh
Ruan, Jau‐Ming
Chen, Yu‐Cheng
author_sort Yang, Ching‐Ching
collection PubMed
description PURPOSE: The target detectability of cone beam computed tomography (CBCT) performed in image‐guided radiation therapy (IGRT) was investigated to achieve sufficient image quality for patient positioning over a course of treatment session while maintaining radiation exposure from CBCT imaging as low as reasonably achievable (ALARA). METHODS: Body CBCT scans operated in half‐fan mode were acquired with three different protocols: CBCT(lowD), CBCT(midD), and CBCT(highD), which resulted in weighted CT dose index (CTDI(w)) of 0.36, 1.43, and 2.78 cGy, respectively. An electron density phantom that is 18 cm in diameter was covered by four layers of 2.5‐cm‐thick bolus to simulate patients of different body sizes. Multivariate analysis was used to examine the impact of body size, radiation exposure, and tissue type on the target detectability of CBCT imaging, quantified as contrast‐to‐noise ratio (CNR). RESULTS: CBCT(midD) allows sufficient target detection of adipose, breast, muscle, liver in a background of water for normal‐weight adults with cross‐sectional diameter less than 28 cm, while CBCT(highD) is suitable for adult patients with larger body sizes or body mass index over 25 kg/m(2). Once the cross‐sectional diameter of adult patients is larger than 35 cm, the CTDI(w) of CBCT scans should be higher than 2.78 cGy to achieve required CNR. As for pediatric and adolescent patients with cross‐sectional diameter less than 25 cm, CBCT(lowD) is able to produce images with sufficient target detection. CONCLUSION: The target detectability of soft tissues in default CBCT scans may not be sufficient for overweight or obese adults. Contrary, pediatric and adolescent patients would receive unnecessarily high radiation exposure from default CBCT scans. Therefore, the selection of acquisition parameters for CBCT scans optimized according to patient body size was proposed to ensure sufficient image quality for daily patient positioning in radiation therapy while achieving the ALARA principle.
format Online
Article
Text
id pubmed-5978665
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-59786652018-06-01 Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes Yang, Ching‐Ching Yu, Pei‐Chieh Ruan, Jau‐Ming Chen, Yu‐Cheng J Appl Clin Med Phys Medical Imaging PURPOSE: The target detectability of cone beam computed tomography (CBCT) performed in image‐guided radiation therapy (IGRT) was investigated to achieve sufficient image quality for patient positioning over a course of treatment session while maintaining radiation exposure from CBCT imaging as low as reasonably achievable (ALARA). METHODS: Body CBCT scans operated in half‐fan mode were acquired with three different protocols: CBCT(lowD), CBCT(midD), and CBCT(highD), which resulted in weighted CT dose index (CTDI(w)) of 0.36, 1.43, and 2.78 cGy, respectively. An electron density phantom that is 18 cm in diameter was covered by four layers of 2.5‐cm‐thick bolus to simulate patients of different body sizes. Multivariate analysis was used to examine the impact of body size, radiation exposure, and tissue type on the target detectability of CBCT imaging, quantified as contrast‐to‐noise ratio (CNR). RESULTS: CBCT(midD) allows sufficient target detection of adipose, breast, muscle, liver in a background of water for normal‐weight adults with cross‐sectional diameter less than 28 cm, while CBCT(highD) is suitable for adult patients with larger body sizes or body mass index over 25 kg/m(2). Once the cross‐sectional diameter of adult patients is larger than 35 cm, the CTDI(w) of CBCT scans should be higher than 2.78 cGy to achieve required CNR. As for pediatric and adolescent patients with cross‐sectional diameter less than 25 cm, CBCT(lowD) is able to produce images with sufficient target detection. CONCLUSION: The target detectability of soft tissues in default CBCT scans may not be sufficient for overweight or obese adults. Contrary, pediatric and adolescent patients would receive unnecessarily high radiation exposure from default CBCT scans. Therefore, the selection of acquisition parameters for CBCT scans optimized according to patient body size was proposed to ensure sufficient image quality for daily patient positioning in radiation therapy while achieving the ALARA principle. John Wiley and Sons Inc. 2018-03-08 /pmc/articles/PMC5978665/ /pubmed/29516610 http://dx.doi.org/10.1002/acm2.12306 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Yang, Ching‐Ching
Yu, Pei‐Chieh
Ruan, Jau‐Ming
Chen, Yu‐Cheng
Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title_full Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title_fullStr Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title_full_unstemmed Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title_short Optimizing the target detectability of cone beam CT performed in image‐guided radiation therapy for patients of different body sizes
title_sort optimizing the target detectability of cone beam ct performed in image‐guided radiation therapy for patients of different body sizes
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978665/
https://www.ncbi.nlm.nih.gov/pubmed/29516610
http://dx.doi.org/10.1002/acm2.12306
work_keys_str_mv AT yangchingching optimizingthetargetdetectabilityofconebeamctperformedinimageguidedradiationtherapyforpatientsofdifferentbodysizes
AT yupeichieh optimizingthetargetdetectabilityofconebeamctperformedinimageguidedradiationtherapyforpatientsofdifferentbodysizes
AT ruanjauming optimizingthetargetdetectabilityofconebeamctperformedinimageguidedradiationtherapyforpatientsofdifferentbodysizes
AT chenyucheng optimizingthetargetdetectabilityofconebeamctperformedinimageguidedradiationtherapyforpatientsofdifferentbodysizes