Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis

Dose reduction techniques have been studied in medical imaging. We propose shutter scan acquisition for region of interest (ROI) imaging to reduce the patient exposure dose received from a digital tomosynthesis system. A prototype chest digital tomosynthesis (CDT) system (LISTEM, Wonju, Korea) and t...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Dohyeon, Jo, Byungdu, Lee, Donghoon, Lee, Haenghwa, Choi, Sunghoon, Kim, Hyemi, Chao, Zhen, Choi, Seungyeon, Kim, Hee‐Joung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978707/
https://www.ncbi.nlm.nih.gov/pubmed/29493082
http://dx.doi.org/10.1002/acm2.12294
_version_ 1783327548663398400
author Kim, Dohyeon
Jo, Byungdu
Lee, Donghoon
Lee, Haenghwa
Choi, Sunghoon
Kim, Hyemi
Chao, Zhen
Choi, Seungyeon
Kim, Hee‐Joung
author_facet Kim, Dohyeon
Jo, Byungdu
Lee, Donghoon
Lee, Haenghwa
Choi, Sunghoon
Kim, Hyemi
Chao, Zhen
Choi, Seungyeon
Kim, Hee‐Joung
author_sort Kim, Dohyeon
collection PubMed
description Dose reduction techniques have been studied in medical imaging. We propose shutter scan acquisition for region of interest (ROI) imaging to reduce the patient exposure dose received from a digital tomosynthesis system. A prototype chest digital tomosynthesis (CDT) system (LISTEM, Wonju, Korea) and the LUNGMAN phantom (Kyoto Kagaku, Japan) with lung nodules 8, 10, and 12 mm in size were used for this study. A total of 41 projections with shutter scan acquisition consisted of 21 truncated projections and 20 non‐truncated projections. For comparison, 41 projections using conventional full view scan acquisition were also acquired. Truncated projections obtained by shutter scan acquisition were corrected by proposed image processing procedure to remove the truncation artifacts. The image quality was evaluated using the contrast to noise ratio (CNR), coefficient of variation (COV), and figure of merit (FOM). We measured the dose area product (DAP) value to verify the dose reduction using shutter scan acquisition. The ROI of the reconstructed image from shutter scan acquisition showed enhanced contrast. The results showed that CNR values of 8 and 12 mm lung nodules increased by 6.38% and 21.21%, respectively, and the CNR value of 10 mm lung nodule decreased by 3.63%. COV values of the lung nodules were lower in a shutter scan image than in a full view scan image. FOM values of 8, 10, and 12 mm lung nodules increased by 3.06, 2.25, and 2.33 times, respectively. This study compared the proposed shutter scan and conventional full view scan acquisition. In conclusion, using a shutter scan acquisition method resulted in enhanced contrast images within the ROI and higher FOM values. The patient exposure dose of the proposed shutter scan acquisition method can be reduced by limiting the field of view (FOV) to focus on the ROI.
format Online
Article
Text
id pubmed-5978707
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-59787072018-06-01 Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis Kim, Dohyeon Jo, Byungdu Lee, Donghoon Lee, Haenghwa Choi, Sunghoon Kim, Hyemi Chao, Zhen Choi, Seungyeon Kim, Hee‐Joung J Appl Clin Med Phys Medical Imaging Dose reduction techniques have been studied in medical imaging. We propose shutter scan acquisition for region of interest (ROI) imaging to reduce the patient exposure dose received from a digital tomosynthesis system. A prototype chest digital tomosynthesis (CDT) system (LISTEM, Wonju, Korea) and the LUNGMAN phantom (Kyoto Kagaku, Japan) with lung nodules 8, 10, and 12 mm in size were used for this study. A total of 41 projections with shutter scan acquisition consisted of 21 truncated projections and 20 non‐truncated projections. For comparison, 41 projections using conventional full view scan acquisition were also acquired. Truncated projections obtained by shutter scan acquisition were corrected by proposed image processing procedure to remove the truncation artifacts. The image quality was evaluated using the contrast to noise ratio (CNR), coefficient of variation (COV), and figure of merit (FOM). We measured the dose area product (DAP) value to verify the dose reduction using shutter scan acquisition. The ROI of the reconstructed image from shutter scan acquisition showed enhanced contrast. The results showed that CNR values of 8 and 12 mm lung nodules increased by 6.38% and 21.21%, respectively, and the CNR value of 10 mm lung nodule decreased by 3.63%. COV values of the lung nodules were lower in a shutter scan image than in a full view scan image. FOM values of 8, 10, and 12 mm lung nodules increased by 3.06, 2.25, and 2.33 times, respectively. This study compared the proposed shutter scan and conventional full view scan acquisition. In conclusion, using a shutter scan acquisition method resulted in enhanced contrast images within the ROI and higher FOM values. The patient exposure dose of the proposed shutter scan acquisition method can be reduced by limiting the field of view (FOV) to focus on the ROI. John Wiley and Sons Inc. 2018-03-01 /pmc/articles/PMC5978707/ /pubmed/29493082 http://dx.doi.org/10.1002/acm2.12294 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Kim, Dohyeon
Jo, Byungdu
Lee, Donghoon
Lee, Haenghwa
Choi, Sunghoon
Kim, Hyemi
Chao, Zhen
Choi, Seungyeon
Kim, Hee‐Joung
Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title_full Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title_fullStr Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title_full_unstemmed Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title_short Feasibility study of shutter scan acquisition for region of interest (ROI) digital tomosynthesis
title_sort feasibility study of shutter scan acquisition for region of interest (roi) digital tomosynthesis
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978707/
https://www.ncbi.nlm.nih.gov/pubmed/29493082
http://dx.doi.org/10.1002/acm2.12294
work_keys_str_mv AT kimdohyeon feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT jobyungdu feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT leedonghoon feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT leehaenghwa feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT choisunghoon feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT kimhyemi feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT chaozhen feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT choiseungyeon feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis
AT kimheejoung feasibilitystudyofshutterscanacquisitionforregionofinterestroidigitaltomosynthesis