Commissioning an in‐room mobile CT for adaptive proton therapy with a compact proton system
PURPOSE: To describe the commissioning of AIRO mobile CT system (AIRO) for adaptive proton therapy on a compact double scattering proton therapy system. METHODS: A Gammex phantom was scanned with varying plug patterns, table heights, and mAs on a CT simulator (CT Sim) and on the AIRO. AIRO‐specific...
Autores principales: | Oliver, Jasmine A., Zeidan, Omar, Meeks, Sanford L., Shah, Amish P., Pukala, Jason, Kelly, Patrick, Ramakrishna, Naren R., Willoughby, Twyla R. |
---|---|
Formato: | Online Artículo Texto |