An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978980/ https://www.ncbi.nlm.nih.gov/pubmed/29520938 http://dx.doi.org/10.1002/acm2.12289 |
_version_ | 1783327595443519488 |
---|---|
author | Stanley, Dennis N. Rasmussen, Karl Kirby, Neil Papanikolaou, Nikos Gutiérrez, Alonso N. |
author_facet | Stanley, Dennis N. Rasmussen, Karl Kirby, Neil Papanikolaou, Nikos Gutiérrez, Alonso N. |
author_sort | Stanley, Dennis N. |
collection | PubMed |
description | INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone‐beam computed tomography (CBCT) mode, planar radiographic kV mode, and the radiographic MV mode of an Elekta VersaHD. MATERIALS AND METHODS: The CATPHAN, QckV‐1, and QC‐3 phantoms were used to evaluate the image quality parameters. The planar radiographic images were analyzed in PIPSpro™ with spatial resolution (f30, f40, f50), contrast to noise ratio (CNR) and noise being recorded. For XVI CBCT, Head and Neck Small20 (S20) and Pelvis Medium20 (M20) standard acquisition modes were evaluated for uniformity, noise, spatial resolution, and HU constancy. Dose and kVp for the XVI were recorded using the Unfors RaySafe Xi system with the R/F low detector for the kV planar radiographic mode. For each metric, values were normalized to the mean and the standard deviations were recorded. RESULTS: A total of 30 measurements were performed on a single Elekta VersaHD linear accelerator over an 18‐month period without significant adjustment or recalibration to the XVI or iViewGT systems during the evaluated time frame. For the planar radiographic spatial resolution, the normalized standard deviation values of the f30, f40, and f50 were 0.004, 0.003, and 0.003 and 0.015, 0.009, and 0.017 for kV and MV, respectively. The average recorded dose for kV was 67.96 μGy. The standard deviations of the evaluated metrics for the S20 acquisition were 0.083(f30), 0.058(f40), 0.056(f50), 0.021(Water/poly‐HU constancy), 0.029(uniformity) and 0.028(noise). The standard deviations for the M20 acquisition were 0.093(f30), 0.043(f40), 0.037(f50), 0.016(Water/poly‐HU constancy), 0.010(uniformity) and 0.011(Noise). CONCLUSION: A study was performed to assess the stability of the basic image quality parameters recommended by TG‐142 for the Elekta XVI and iViewGT imaging systems. The two systems show consistent imaging and dosimetric properties over the evaluated time frame. |
format | Online Article Text |
id | pubmed-5978980 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-59789802018-06-01 An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems Stanley, Dennis N. Rasmussen, Karl Kirby, Neil Papanikolaou, Nikos Gutiérrez, Alonso N. J Appl Clin Med Phys Radiation Oncology Physics INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone‐beam computed tomography (CBCT) mode, planar radiographic kV mode, and the radiographic MV mode of an Elekta VersaHD. MATERIALS AND METHODS: The CATPHAN, QckV‐1, and QC‐3 phantoms were used to evaluate the image quality parameters. The planar radiographic images were analyzed in PIPSpro™ with spatial resolution (f30, f40, f50), contrast to noise ratio (CNR) and noise being recorded. For XVI CBCT, Head and Neck Small20 (S20) and Pelvis Medium20 (M20) standard acquisition modes were evaluated for uniformity, noise, spatial resolution, and HU constancy. Dose and kVp for the XVI were recorded using the Unfors RaySafe Xi system with the R/F low detector for the kV planar radiographic mode. For each metric, values were normalized to the mean and the standard deviations were recorded. RESULTS: A total of 30 measurements were performed on a single Elekta VersaHD linear accelerator over an 18‐month period without significant adjustment or recalibration to the XVI or iViewGT systems during the evaluated time frame. For the planar radiographic spatial resolution, the normalized standard deviation values of the f30, f40, and f50 were 0.004, 0.003, and 0.003 and 0.015, 0.009, and 0.017 for kV and MV, respectively. The average recorded dose for kV was 67.96 μGy. The standard deviations of the evaluated metrics for the S20 acquisition were 0.083(f30), 0.058(f40), 0.056(f50), 0.021(Water/poly‐HU constancy), 0.029(uniformity) and 0.028(noise). The standard deviations for the M20 acquisition were 0.093(f30), 0.043(f40), 0.037(f50), 0.016(Water/poly‐HU constancy), 0.010(uniformity) and 0.011(Noise). CONCLUSION: A study was performed to assess the stability of the basic image quality parameters recommended by TG‐142 for the Elekta XVI and iViewGT imaging systems. The two systems show consistent imaging and dosimetric properties over the evaluated time frame. John Wiley and Sons Inc. 2018-03-09 /pmc/articles/PMC5978980/ /pubmed/29520938 http://dx.doi.org/10.1002/acm2.12289 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Stanley, Dennis N. Rasmussen, Karl Kirby, Neil Papanikolaou, Nikos Gutiérrez, Alonso N. An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title | An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title_full | An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title_fullStr | An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title_full_unstemmed | An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title_short | An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems |
title_sort | evaluation of the stability of image quality parameters of elekta x‐ray volume imager and iviewgt imaging systems |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978980/ https://www.ncbi.nlm.nih.gov/pubmed/29520938 http://dx.doi.org/10.1002/acm2.12289 |
work_keys_str_mv | AT stanleydennisn anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT rasmussenkarl anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT kirbyneil anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT papanikolaounikos anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT gutierrezalonson anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT stanleydennisn evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT rasmussenkarl evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT kirbyneil evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT papanikolaounikos evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems AT gutierrezalonson evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems |