An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems

INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone...

Descripción completa

Detalles Bibliográficos
Autores principales: Stanley, Dennis N., Rasmussen, Karl, Kirby, Neil, Papanikolaou, Nikos, Gutiérrez, Alonso N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978980/
https://www.ncbi.nlm.nih.gov/pubmed/29520938
http://dx.doi.org/10.1002/acm2.12289
_version_ 1783327595443519488
author Stanley, Dennis N.
Rasmussen, Karl
Kirby, Neil
Papanikolaou, Nikos
Gutiérrez, Alonso N.
author_facet Stanley, Dennis N.
Rasmussen, Karl
Kirby, Neil
Papanikolaou, Nikos
Gutiérrez, Alonso N.
author_sort Stanley, Dennis N.
collection PubMed
description INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone‐beam computed tomography (CBCT) mode, planar radiographic kV mode, and the radiographic MV mode of an Elekta VersaHD. MATERIALS AND METHODS: The CATPHAN, QckV‐1, and QC‐3 phantoms were used to evaluate the image quality parameters. The planar radiographic images were analyzed in PIPSpro™ with spatial resolution (f30, f40, f50), contrast to noise ratio (CNR) and noise being recorded. For XVI CBCT, Head and Neck Small20 (S20) and Pelvis Medium20 (M20) standard acquisition modes were evaluated for uniformity, noise, spatial resolution, and HU constancy. Dose and kVp for the XVI were recorded using the Unfors RaySafe Xi system with the R/F low detector for the kV planar radiographic mode. For each metric, values were normalized to the mean and the standard deviations were recorded. RESULTS: A total of 30 measurements were performed on a single Elekta VersaHD linear accelerator over an 18‐month period without significant adjustment or recalibration to the XVI or iViewGT systems during the evaluated time frame. For the planar radiographic spatial resolution, the normalized standard deviation values of the f30, f40, and f50 were 0.004, 0.003, and 0.003 and 0.015, 0.009, and 0.017 for kV and MV, respectively. The average recorded dose for kV was 67.96 μGy. The standard deviations of the evaluated metrics for the S20 acquisition were 0.083(f30), 0.058(f40), 0.056(f50), 0.021(Water/poly‐HU constancy), 0.029(uniformity) and 0.028(noise). The standard deviations for the M20 acquisition were 0.093(f30), 0.043(f40), 0.037(f50), 0.016(Water/poly‐HU constancy), 0.010(uniformity) and 0.011(Noise). CONCLUSION: A study was performed to assess the stability of the basic image quality parameters recommended by TG‐142 for the Elekta XVI and iViewGT imaging systems. The two systems show consistent imaging and dosimetric properties over the evaluated time frame.
format Online
Article
Text
id pubmed-5978980
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-59789802018-06-01 An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems Stanley, Dennis N. Rasmussen, Karl Kirby, Neil Papanikolaou, Nikos Gutiérrez, Alonso N. J Appl Clin Med Phys Radiation Oncology Physics INTRODUCTION: A robust image quality assurance and analysis methodology for image‐guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long‐term stability of selected image parameters was assessed and evaluated for the cone‐beam computed tomography (CBCT) mode, planar radiographic kV mode, and the radiographic MV mode of an Elekta VersaHD. MATERIALS AND METHODS: The CATPHAN, QckV‐1, and QC‐3 phantoms were used to evaluate the image quality parameters. The planar radiographic images were analyzed in PIPSpro™ with spatial resolution (f30, f40, f50), contrast to noise ratio (CNR) and noise being recorded. For XVI CBCT, Head and Neck Small20 (S20) and Pelvis Medium20 (M20) standard acquisition modes were evaluated for uniformity, noise, spatial resolution, and HU constancy. Dose and kVp for the XVI were recorded using the Unfors RaySafe Xi system with the R/F low detector for the kV planar radiographic mode. For each metric, values were normalized to the mean and the standard deviations were recorded. RESULTS: A total of 30 measurements were performed on a single Elekta VersaHD linear accelerator over an 18‐month period without significant adjustment or recalibration to the XVI or iViewGT systems during the evaluated time frame. For the planar radiographic spatial resolution, the normalized standard deviation values of the f30, f40, and f50 were 0.004, 0.003, and 0.003 and 0.015, 0.009, and 0.017 for kV and MV, respectively. The average recorded dose for kV was 67.96 μGy. The standard deviations of the evaluated metrics for the S20 acquisition were 0.083(f30), 0.058(f40), 0.056(f50), 0.021(Water/poly‐HU constancy), 0.029(uniformity) and 0.028(noise). The standard deviations for the M20 acquisition were 0.093(f30), 0.043(f40), 0.037(f50), 0.016(Water/poly‐HU constancy), 0.010(uniformity) and 0.011(Noise). CONCLUSION: A study was performed to assess the stability of the basic image quality parameters recommended by TG‐142 for the Elekta XVI and iViewGT imaging systems. The two systems show consistent imaging and dosimetric properties over the evaluated time frame. John Wiley and Sons Inc. 2018-03-09 /pmc/articles/PMC5978980/ /pubmed/29520938 http://dx.doi.org/10.1002/acm2.12289 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Stanley, Dennis N.
Rasmussen, Karl
Kirby, Neil
Papanikolaou, Nikos
Gutiérrez, Alonso N.
An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title_full An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title_fullStr An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title_full_unstemmed An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title_short An evaluation of the stability of image quality parameters of Elekta X‐ray volume imager and iViewGT imaging systems
title_sort evaluation of the stability of image quality parameters of elekta x‐ray volume imager and iviewgt imaging systems
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978980/
https://www.ncbi.nlm.nih.gov/pubmed/29520938
http://dx.doi.org/10.1002/acm2.12289
work_keys_str_mv AT stanleydennisn anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT rasmussenkarl anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT kirbyneil anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT papanikolaounikos anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT gutierrezalonson anevaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT stanleydennisn evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT rasmussenkarl evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT kirbyneil evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT papanikolaounikos evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems
AT gutierrezalonson evaluationofthestabilityofimagequalityparametersofelektaxrayvolumeimagerandiviewgtimagingsystems