Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa
PURPOSE: To examine repeatability and reproducibility of ellipsoid zone (EZ) width measurements in patients with retinitis pigmentosa (RP) using a longitudinal reflectivity profile (LRP) analysis. METHODS: We examined Bioptigen optical coherence tomography (OCT) scans from 48 subjects with RP or Ush...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5989764/ https://www.ncbi.nlm.nih.gov/pubmed/29881650 http://dx.doi.org/10.1167/tvst.7.3.13 |
_version_ | 1783329516473548800 |
---|---|
author | Strampe, Margaret R. Huckenpahler, Alison L. Higgins, Brian P. Tarima, Sergey Visotcky, Alexis Stepien, Kimberly E. Kay, Christine N. Carroll, Joseph |
author_facet | Strampe, Margaret R. Huckenpahler, Alison L. Higgins, Brian P. Tarima, Sergey Visotcky, Alexis Stepien, Kimberly E. Kay, Christine N. Carroll, Joseph |
author_sort | Strampe, Margaret R. |
collection | PubMed |
description | PURPOSE: To examine repeatability and reproducibility of ellipsoid zone (EZ) width measurements in patients with retinitis pigmentosa (RP) using a longitudinal reflectivity profile (LRP) analysis. METHODS: We examined Bioptigen optical coherence tomography (OCT) scans from 48 subjects with RP or Usher syndrome. Nominal scan lengths were 6, 7, or 10 mm, and the lateral scale of each scan was calculated using axial length measurements. LRPs were generated from OCT line scans, and the peak corresponding to EZ was manually identified using ImageJ. The locations at which the EZ peak disappeared were used to calculate EZ width. Each scan was analyzed twice by each of two observers, who were masked to their previous measurements and those of the other observer. RESULTS: On average, horizontal width (HW) was significantly greater than vertical width (VW), and there was high interocular symmetry for both HW and VW. We observed excellent intraobserver repeatability with intraclass correlation coefficients (ICCs) ranging from 0.996 to 0.998 for HW and VW measurements. Interobserver reproducibility was also excellent for both HW (ICC = 0.989; 95% confidence interval [CI] = 0.983–0.995) and VW (ICC = 0.991; 95% CI = 0.985–0.996), with no significant bias observed between observers. CONCLUSIONS: EZ width can be measured using LRPs with excellent repeatability and reproducibility. Our observation of greater HW than VW is consistent with previous observations in RP, though the reason for this anisotropy remains unclear. TRANSLATIONAL RELEVANCE: We describe repeatability and reproducibility of a method for measuring EZ width in patients with RP or Usher syndrome. This approach could facilitate measurement of retinal band thickness and/or intensity. |
format | Online Article Text |
id | pubmed-5989764 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | The Association for Research in Vision and Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-59897642018-06-07 Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa Strampe, Margaret R. Huckenpahler, Alison L. Higgins, Brian P. Tarima, Sergey Visotcky, Alexis Stepien, Kimberly E. Kay, Christine N. Carroll, Joseph Transl Vis Sci Technol Articles PURPOSE: To examine repeatability and reproducibility of ellipsoid zone (EZ) width measurements in patients with retinitis pigmentosa (RP) using a longitudinal reflectivity profile (LRP) analysis. METHODS: We examined Bioptigen optical coherence tomography (OCT) scans from 48 subjects with RP or Usher syndrome. Nominal scan lengths were 6, 7, or 10 mm, and the lateral scale of each scan was calculated using axial length measurements. LRPs were generated from OCT line scans, and the peak corresponding to EZ was manually identified using ImageJ. The locations at which the EZ peak disappeared were used to calculate EZ width. Each scan was analyzed twice by each of two observers, who were masked to their previous measurements and those of the other observer. RESULTS: On average, horizontal width (HW) was significantly greater than vertical width (VW), and there was high interocular symmetry for both HW and VW. We observed excellent intraobserver repeatability with intraclass correlation coefficients (ICCs) ranging from 0.996 to 0.998 for HW and VW measurements. Interobserver reproducibility was also excellent for both HW (ICC = 0.989; 95% confidence interval [CI] = 0.983–0.995) and VW (ICC = 0.991; 95% CI = 0.985–0.996), with no significant bias observed between observers. CONCLUSIONS: EZ width can be measured using LRPs with excellent repeatability and reproducibility. Our observation of greater HW than VW is consistent with previous observations in RP, though the reason for this anisotropy remains unclear. TRANSLATIONAL RELEVANCE: We describe repeatability and reproducibility of a method for measuring EZ width in patients with RP or Usher syndrome. This approach could facilitate measurement of retinal band thickness and/or intensity. The Association for Research in Vision and Ophthalmology 2018-06-04 /pmc/articles/PMC5989764/ /pubmed/29881650 http://dx.doi.org/10.1167/tvst.7.3.13 Text en Copyright 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. |
spellingShingle | Articles Strampe, Margaret R. Huckenpahler, Alison L. Higgins, Brian P. Tarima, Sergey Visotcky, Alexis Stepien, Kimberly E. Kay, Christine N. Carroll, Joseph Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title | Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title_full | Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title_fullStr | Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title_full_unstemmed | Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title_short | Intraobserver Repeatability and Interobserver Reproducibility of Ellipsoid Zone Measurements in Retinitis Pigmentosa |
title_sort | intraobserver repeatability and interobserver reproducibility of ellipsoid zone measurements in retinitis pigmentosa |
topic | Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5989764/ https://www.ncbi.nlm.nih.gov/pubmed/29881650 http://dx.doi.org/10.1167/tvst.7.3.13 |
work_keys_str_mv | AT strampemargaretr intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT huckenpahleralisonl intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT higginsbrianp intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT tarimasergey intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT visotckyalexis intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT stepienkimberlye intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT kaychristinen intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa AT carrolljoseph intraobserverrepeatabilityandinterobserverreproducibilityofellipsoidzonemeasurementsinretinitispigmentosa |