Cargando…

Comparison of Revision Rates Due to Aseptic Loosening between High-Flex and Conventional Knee Prostheses

PURPOSE: The purpose was to evaluate and compare the revision rate due to aseptic loosening between a high-flex prosthesis and a conventional prosthesis. MATERIALS AND METHODS: Two thousand seventy-eight knees (1,377 patients) with at least 2 years of follow-up after total knee arthroplasty were rev...

Descripción completa

Detalles Bibliográficos
Autores principales: Choi, Young-Joon, Lee, Ki Won, Ha, Jung-Ki, Bae, Joo-Yul, Lee, Suk Kyu, Kim, Sang-Bum, Seo, Dong-Kyo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Knee Society 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5990236/
https://www.ncbi.nlm.nih.gov/pubmed/29843201
http://dx.doi.org/10.5792/ksrr.17.071
Descripción
Sumario:PURPOSE: The purpose was to evaluate and compare the revision rate due to aseptic loosening between a high-flex prosthesis and a conventional prosthesis. MATERIALS AND METHODS: Two thousand seventy-eight knees (1,377 patients) with at least 2 years of follow-up after total knee arthroplasty were reviewed. Two types of implants were selected (LPS-Flex and LPS, Zimmer) to compare revision and survival rates and sites of loosened prosthesis component. RESULTS: The revision rate of the LPS-Flex (4.9%) was significantly higher than that of the conventional prosthesis (0.6%) (p<0.001). The 5-, 10-, and 15-year survival rates were 98.9%, 96.2% and 92.0%, respectively, for the LPS-Flex and 99.8%, 98.5% and 93.5%, respectively, for the LPS. The survival rate of the high-flex prosthesis was significantly lower than that of the conventional prosthesis, especially in the mid-term period (range, 5 to 10 years; p=0.002). The loosening rate of the femoral component was significantly higher in the LPS-Flex prosthesis (p=0.001). CONCLUSIONS: The LPS-Flex had a higher revision rate due to aseptic loosening than the LPS prosthesis in the large population series with a long follow-up. The LPS-Flex should be used carefully considering the risk of femoral component aseptic loosening in the mid-term (range, 5 to 10 years) follow-up period after initial operation.