Cargando…

Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty

Failures unique to posterior cruciate-substituting total knee prostheses rarely include polyethylene post fractures but have been described. We report a case involving a fracture of the distal insert cone of a rotating-platform (RP) polyethylene liner in a primary total knee arthroplasty. This case...

Descripción completa

Detalles Bibliográficos
Autores principales: Angerame, Marc R., Jennings, Jason M., Dennis, Douglas A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5994790/
https://www.ncbi.nlm.nih.gov/pubmed/29896543
http://dx.doi.org/10.1016/j.artd.2017.10.002
_version_ 1783330504659959808
author Angerame, Marc R.
Jennings, Jason M.
Dennis, Douglas A.
author_facet Angerame, Marc R.
Jennings, Jason M.
Dennis, Douglas A.
author_sort Angerame, Marc R.
collection PubMed
description Failures unique to posterior cruciate-substituting total knee prostheses rarely include polyethylene post fractures but have been described. We report a case involving a fracture of the distal insert cone of a rotating-platform (RP) polyethylene liner in a primary total knee arthroplasty. This case highlights a 67-year-old male presenting with new-onset knee pain and recurrent effusions with osteolysis 11 years following placement of a posterior-stabilized, RP total knee arthroplasty. At the time of revision surgery, the polyethylene insert cone was found to be fractured just below the junction between cone and the body of the insert. Liner exchange, synovectomy, and osteolytic-defect curettage and cement packing were performed. One year following revision surgery, the patient is without pain and has returned to function without limitations. Clinicians must be aware of this possible failure with RP prostheses in the setting of pain with a stable knee, recurrent aseptic effusions, and osteolysis.
format Online
Article
Text
id pubmed-5994790
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-59947902018-06-12 Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty Angerame, Marc R. Jennings, Jason M. Dennis, Douglas A. Arthroplast Today Case Report Failures unique to posterior cruciate-substituting total knee prostheses rarely include polyethylene post fractures but have been described. We report a case involving a fracture of the distal insert cone of a rotating-platform (RP) polyethylene liner in a primary total knee arthroplasty. This case highlights a 67-year-old male presenting with new-onset knee pain and recurrent effusions with osteolysis 11 years following placement of a posterior-stabilized, RP total knee arthroplasty. At the time of revision surgery, the polyethylene insert cone was found to be fractured just below the junction between cone and the body of the insert. Liner exchange, synovectomy, and osteolytic-defect curettage and cement packing were performed. One year following revision surgery, the patient is without pain and has returned to function without limitations. Clinicians must be aware of this possible failure with RP prostheses in the setting of pain with a stable knee, recurrent aseptic effusions, and osteolysis. Elsevier 2017-11-07 /pmc/articles/PMC5994790/ /pubmed/29896543 http://dx.doi.org/10.1016/j.artd.2017.10.002 Text en © 2017 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Angerame, Marc R.
Jennings, Jason M.
Dennis, Douglas A.
Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title_full Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title_fullStr Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title_full_unstemmed Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title_short Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
title_sort fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5994790/
https://www.ncbi.nlm.nih.gov/pubmed/29896543
http://dx.doi.org/10.1016/j.artd.2017.10.002
work_keys_str_mv AT angeramemarcr fractureoftheinsertconeofapolyethylenelinerinafailedposteriorstabilizedrotatingplatformtotalkneearthroplasty
AT jenningsjasonm fractureoftheinsertconeofapolyethylenelinerinafailedposteriorstabilizedrotatingplatformtotalkneearthroplasty
AT dennisdouglasa fractureoftheinsertconeofapolyethylenelinerinafailedposteriorstabilizedrotatingplatformtotalkneearthroplasty