Measurement and Evaluation of Quantitative Performance of PET/CT Images before a Multicenter Clinical Trial
To ensure the reliability of the planned multi-center clinical trial, we assessed the consistence and comparability of the quantitative parameters of the eight PET/CT units that will be used in this trial. PET/CT images were scanned using a PET NEMA image quality phantom (Biodex) on the eight units...