Endovascular retrieval of a CardioMEMS heart failure system
As the creation and utilization of new implantable devices increases, so does the need for interventionalists to devise unique retrieval mechanisms. This report describes the first endovascular retrieval of a CardioMEMS heart failure monitoring device. A 20-mm gooseneck snare was utilized in conjunc...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6000058/ https://www.ncbi.nlm.nih.gov/pubmed/29904478 http://dx.doi.org/10.1016/j.radcr.2018.01.013 |
Sumario: | As the creation and utilization of new implantable devices increases, so does the need for interventionalists to devise unique retrieval mechanisms. This report describes the first endovascular retrieval of a CardioMEMS heart failure monitoring device. A 20-mm gooseneck snare was utilized in conjunction with a 9-French sheath and Envoy catheter for retrieval. The patient suffered no immediate postprocedural complications but died 5 days after the procedure from multiorgan failure secondary to sepsis. |
---|