Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study

PURPOSE: The purpose of this study was to evaluate the repeatability and matching accuracy between two identical intraoral spectrophotometers. MATERIALS AND METHODS: The maxillary right central incisor, canine, and mandibular left central incisor of each of 30 patients were measured using 2 identica...

Descripción completa

Detalles Bibliográficos
Autor principal: Kim, Hee-Kyung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Prosthodontics 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6004350/
https://www.ncbi.nlm.nih.gov/pubmed/29930796
http://dx.doi.org/10.4047/jap.2018.10.3.252
_version_ 1783332502506569728
author Kim, Hee-Kyung
author_facet Kim, Hee-Kyung
author_sort Kim, Hee-Kyung
collection PubMed
description PURPOSE: The purpose of this study was to evaluate the repeatability and matching accuracy between two identical intraoral spectrophotometers. MATERIALS AND METHODS: The maxillary right central incisor, canine, and mandibular left central incisor of each of 30 patients were measured using 2 identical intraoral spectrophotometers with different serial numbers (EasyShade V). The color of each shade tab from 3 shade guides (VITA 3D-Master) was also determined with both devices. All measurements were performed by a single operator. Statistical analyses were performed to verify the repeatability, accuracy, and the differences between the devices with paired t-tests, one-way ANOVA, and intra-class correlation coefficients (ICCs) (α=.05). RESULTS: A high level of measurement repeatability (ICC>0.90) among L(*), a(*), and b(*) color components was observed within and between devices (P<.001). Intra-device matching agreement rates were 80.00% and 81.11%, respectively, while inter-device matching agreement rate was 51.85%. ANOVA revealed no significant different color values within each device, while paired t-test provided significant different color values between both devices. The CIEDE2000 color differences between both devices were 2.28±1.61 ΔE(00) for in-vivo readings. Regarding the clinical matching accuracy of both devices, ΔE(00) values between teeth and matching shade tabs were 3.05±1.19 and 2.86±1.02, respectively. CONCLUSION: Although two EasyShade V devices with different serial numbers show high repeatability of CIE L(*), a(*), and b(*) measurements, they could provide different color values and shade for the same tooth.
format Online
Article
Text
id pubmed-6004350
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher The Korean Academy of Prosthodontics
record_format MEDLINE/PubMed
spelling pubmed-60043502018-06-21 Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study Kim, Hee-Kyung J Adv Prosthodont Original Article PURPOSE: The purpose of this study was to evaluate the repeatability and matching accuracy between two identical intraoral spectrophotometers. MATERIALS AND METHODS: The maxillary right central incisor, canine, and mandibular left central incisor of each of 30 patients were measured using 2 identical intraoral spectrophotometers with different serial numbers (EasyShade V). The color of each shade tab from 3 shade guides (VITA 3D-Master) was also determined with both devices. All measurements were performed by a single operator. Statistical analyses were performed to verify the repeatability, accuracy, and the differences between the devices with paired t-tests, one-way ANOVA, and intra-class correlation coefficients (ICCs) (α=.05). RESULTS: A high level of measurement repeatability (ICC>0.90) among L(*), a(*), and b(*) color components was observed within and between devices (P<.001). Intra-device matching agreement rates were 80.00% and 81.11%, respectively, while inter-device matching agreement rate was 51.85%. ANOVA revealed no significant different color values within each device, while paired t-test provided significant different color values between both devices. The CIEDE2000 color differences between both devices were 2.28±1.61 ΔE(00) for in-vivo readings. Regarding the clinical matching accuracy of both devices, ΔE(00) values between teeth and matching shade tabs were 3.05±1.19 and 2.86±1.02, respectively. CONCLUSION: Although two EasyShade V devices with different serial numbers show high repeatability of CIE L(*), a(*), and b(*) measurements, they could provide different color values and shade for the same tooth. The Korean Academy of Prosthodontics 2018-06 2018-06-12 /pmc/articles/PMC6004350/ /pubmed/29930796 http://dx.doi.org/10.4047/jap.2018.10.3.252 Text en © 2018 The Korean Academy of Prosthodontics http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kim, Hee-Kyung
Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title_full Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title_fullStr Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title_full_unstemmed Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title_short Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
title_sort evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6004350/
https://www.ncbi.nlm.nih.gov/pubmed/29930796
http://dx.doi.org/10.4047/jap.2018.10.3.252
work_keys_str_mv AT kimheekyung evaluationoftherepeatabilityandmatchingaccuracybetweentwoidenticalintraoralspectrophotometersaninvivoandinvitrostudy