Comparison of Filtered Back Projection, Hybrid Iterative Reconstruction, Model-Based Iterative Reconstruction, and Virtual Monoenergetic Reconstruction Images at Both Low- and Standard-Dose Settings in Measurement of Emphysema Volume and Airway Wall Thickness: A CT Phantom Study

OBJECTIVE: To evaluate the accuracy of emphysema volume (EV) and airway measurements (AMs) produced by various iterative reconstruction (IR) algorithms and virtual monoenergetic images (VME) at both low- and standard-dose settings. MATERIALS AND METHODS: Computed tomography (CT) images were obtained...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Cherry, Lee, Ki Yeol, Shin, Chol, Kang, Eun-Young, Oh, Yu-Whan, Ha, Moin, Ko, Chang Sub, Cha, Jaehyung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Radiology 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6005943/
https://www.ncbi.nlm.nih.gov/pubmed/29962888
http://dx.doi.org/10.3348/kjr.2018.19.4.809
Descripción
Sumario:OBJECTIVE: To evaluate the accuracy of emphysema volume (EV) and airway measurements (AMs) produced by various iterative reconstruction (IR) algorithms and virtual monoenergetic images (VME) at both low- and standard-dose settings. MATERIALS AND METHODS: Computed tomography (CT) images were obtained on phantom at both low- (30 mAs at 120 kVp) and standard-doses (100 mAs at 120 kVp). Each CT scan was reconstructed using filtered back projection, hybrid IR (iDose(4); Philips Healthcare), model-based IR (IMR-R1, IMR-ST1, IMR-SP1; Philips Healthcare), and VME at 70 keV (VME70). The EV of each air column and wall area percentage (WA%) of each airway tube were measured in all algorithms. Absolute percentage measurement errors of EV (APE(vol)) and AM (APE(WA%)) were then calculated. RESULTS: Emphysema volume was most accurately measured in IMR-R1 (APE(vol) in low-dose, 0.053 ± 0.002; APE(vol) in standard-dose, 0.047 ± 0.003; all p < 0.001) and AM was the most accurate in IMR-SP1 on both low- and standard-doses CT (APE(WA%) in low-dose, 0.067 ± 0.002; APE(WA%) in standard-dose, 0.06 ± 0.003; all p < 0.001). There were no significant differences in the APE(vol) of IMR-R1 between low- and standard-doses (all p > 0.05). VME70 showed a significantly higher APE(vol) than iDose(4), IMR-R1, and IMR-ST1 (all p < 0.004). VME70 also showed a significantly higher APE(WA%) compared with the other algorithms (all p < 0.001). CONCLUSION: IMR was the most accurate technique for measurement of both EV and airway wall thickness. However, VME70 did not show a significantly better accuracy compared with other algorithms.