Intensity output and effectiveness of light curing units in dental offices

BACKGROUND: The aims of the study were measuring the light intensity of light curing units used in Qazvin’s dental offices, determining the relationship between the clinical age of these units and their light intensity, and identifying the reasons for repairing them. MATERIAL AND METHODS: In this cr...

Descripción completa

Detalles Bibliográficos
Autores principales: Omidi, Baharan-Ranjbar, Gosili, Armin, Jaber-Ansari, Mona, Mahdkhah, Ailin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medicina Oral S.L. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6012498/
https://www.ncbi.nlm.nih.gov/pubmed/29946413
http://dx.doi.org/10.4317/jced.54756
_version_ 1783333934125285376
author Omidi, Baharan-Ranjbar
Gosili, Armin
Jaber-Ansari, Mona
Mahdkhah, Ailin
author_facet Omidi, Baharan-Ranjbar
Gosili, Armin
Jaber-Ansari, Mona
Mahdkhah, Ailin
author_sort Omidi, Baharan-Ranjbar
collection PubMed
description BACKGROUND: The aims of the study were measuring the light intensity of light curing units used in Qazvin’s dental offices, determining the relationship between the clinical age of these units and their light intensity, and identifying the reasons for repairing them. MATERIAL AND METHODS: In this cross-sectional study, the output intensity of 95 light curing devices was evaluated using a radiometer. The average output intensity was divided up into four categories (less than 200, 200-299, 300-500, and more than 500 mW/cm2). In addition, a questionnaire was designed to obtain information mainly about the type, clinical age, and frequency of maintenance of the units and the reasons for fixing them. Data were analyzed using Kolmogorov-Smirnov, chi-squared, and t-tests (p< 0.05) on SPSS 24. RESULTS: A total of 95 light curing units were examined, with 61 (64.2%) of them being of the LED type and 34 (35.8%) of the QTH type. While average light intensity in LED units was significantly higher than in QTH devices, the two device types were not significantly different regarding desirable light intensity (i.e., ≥ 300 mw/cm2). A negative correlation was observed between clinical age and light intensity. In addition, bulb replacement in QTH devices was over three times as much as in LED units. Also, repairing QTHs was more than twice as much frequent as fixing LEDs. The most common reason for repair was the breakage of the tip of the device. CONCLUSIONS: The light intensity of LED units is significantly higher than that of QTH devices, and the frequency of repairing in QTHs was significantly more than in LEDs. Furthermore, light intensity decreases with aging, and dentists should regularly monitor the conditions of light units. Key words:Light curing unit, radiometer, light intensity, dental equipment, dental offices.
format Online
Article
Text
id pubmed-6012498
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medicina Oral S.L.
record_format MEDLINE/PubMed
spelling pubmed-60124982018-06-26 Intensity output and effectiveness of light curing units in dental offices Omidi, Baharan-Ranjbar Gosili, Armin Jaber-Ansari, Mona Mahdkhah, Ailin J Clin Exp Dent Research BACKGROUND: The aims of the study were measuring the light intensity of light curing units used in Qazvin’s dental offices, determining the relationship between the clinical age of these units and their light intensity, and identifying the reasons for repairing them. MATERIAL AND METHODS: In this cross-sectional study, the output intensity of 95 light curing devices was evaluated using a radiometer. The average output intensity was divided up into four categories (less than 200, 200-299, 300-500, and more than 500 mW/cm2). In addition, a questionnaire was designed to obtain information mainly about the type, clinical age, and frequency of maintenance of the units and the reasons for fixing them. Data were analyzed using Kolmogorov-Smirnov, chi-squared, and t-tests (p< 0.05) on SPSS 24. RESULTS: A total of 95 light curing units were examined, with 61 (64.2%) of them being of the LED type and 34 (35.8%) of the QTH type. While average light intensity in LED units was significantly higher than in QTH devices, the two device types were not significantly different regarding desirable light intensity (i.e., ≥ 300 mw/cm2). A negative correlation was observed between clinical age and light intensity. In addition, bulb replacement in QTH devices was over three times as much as in LED units. Also, repairing QTHs was more than twice as much frequent as fixing LEDs. The most common reason for repair was the breakage of the tip of the device. CONCLUSIONS: The light intensity of LED units is significantly higher than that of QTH devices, and the frequency of repairing in QTHs was significantly more than in LEDs. Furthermore, light intensity decreases with aging, and dentists should regularly monitor the conditions of light units. Key words:Light curing unit, radiometer, light intensity, dental equipment, dental offices. Medicina Oral S.L. 2018-06-01 /pmc/articles/PMC6012498/ /pubmed/29946413 http://dx.doi.org/10.4317/jced.54756 Text en Copyright: © 2018 Medicina Oral S.L. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Omidi, Baharan-Ranjbar
Gosili, Armin
Jaber-Ansari, Mona
Mahdkhah, Ailin
Intensity output and effectiveness of light curing units in dental offices
title Intensity output and effectiveness of light curing units in dental offices
title_full Intensity output and effectiveness of light curing units in dental offices
title_fullStr Intensity output and effectiveness of light curing units in dental offices
title_full_unstemmed Intensity output and effectiveness of light curing units in dental offices
title_short Intensity output and effectiveness of light curing units in dental offices
title_sort intensity output and effectiveness of light curing units in dental offices
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6012498/
https://www.ncbi.nlm.nih.gov/pubmed/29946413
http://dx.doi.org/10.4317/jced.54756
work_keys_str_mv AT omidibaharanranjbar intensityoutputandeffectivenessoflightcuringunitsindentaloffices
AT gosiliarmin intensityoutputandeffectivenessoflightcuringunitsindentaloffices
AT jaberansarimona intensityoutputandeffectivenessoflightcuringunitsindentaloffices
AT mahdkhahailin intensityoutputandeffectivenessoflightcuringunitsindentaloffices