Cargando…

Analysis method for PCBs in reclaimed oil using a fast-GC triple stage quadrupole mass spectrometer with the 13-component quantitation method

It is necessary for companies supplying reclaimed oil to analyze polychlorinated biphenyls (PCBs), because there is a possibility of the presence of contaminants due to trace-level PCBs in the reclaimed oil. However, common analysis methods of PCBs are time-consuming and complicated. Fast-GC triple...

Descripción completa

Detalles Bibliográficos
Autores principales: Takakuwa, Hiroshi, Miura, Takashi, Matsumura, Toru, Hashi, Norie, Kubota, Michiyo, Kutsukake, Hiroshi, Muramatsu, Kazuhiro, Kasamatsu, Takashi, Okuda, Masahiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6015118/
https://www.ncbi.nlm.nih.gov/pubmed/29204936
http://dx.doi.org/10.1007/s11356-017-0533-x
Descripción
Sumario:It is necessary for companies supplying reclaimed oil to analyze polychlorinated biphenyls (PCBs), because there is a possibility of the presence of contaminants due to trace-level PCBs in the reclaimed oil. However, common analysis methods of PCBs are time-consuming and complicated. Fast-GC triple stage quadrupole mass spectrometer with the 13-component quantitation method is an official method for analyzing PCBs in insulating oil in Japan. This method is extremely fast and simplified. The purpose of this study involves an investigation of the aforementioned fast and simple method for potential use in the analysis of reclaimed oil. Furthermore, it was attempted to combine the method with sample preparation involving only hexane dilution. The effect of sample dilutions corresponding to 100, 300, and 500 times was evaluated for reducing the matrix effect. The matrix effect was suppressed at a dilution ratio equal to or exceeding 300 times. Calibration curves of four points, namely 0.01, 0.05, 0.1, and 0.5 ng/mL, (ignored origin) by using an internal standard method were prepared for the 13 components. The square of regression coefficient (R(2)) values of all calibration curves exceeded 0.997. This method was adopted for the analysis of reclaimed oil containing 0.5 μg/mL PCBs, which corresponds to the judgment criteria, and accurate quantitation (accuracy value, 94.0–102%) and good repeatability (%RSD, 3.6%) were obtained. Furthermore, the required sensitivity was maintained even when 800 samples were analyzed without a cleaning ion source and an exchanging analysis column.