Is Ahmed Glaucoma Valve Consistent in Performance?

PURPOSE: To investigate the mechanism of action and consistency in flow characteristics of the Ahmed glaucoma valve (AGV) under simulated physiological conditions in-vitro and to evaluate whether resistance during priming has any effect on performance of the device. METHODS: Each newly opened AGV de...

Descripción completa

Detalles Bibliográficos
Autores principales: Choudhari, Nikhil S., Badakere, Swathi Vallabh, Richhariya, Ashutosh, Chittajallu, Sai Naga Sri Harsha, Senthil, Sirisha, Garudadri, Chandra Sekhar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6016431/
https://www.ncbi.nlm.nih.gov/pubmed/29946493
http://dx.doi.org/10.1167/tvst.7.3.19
_version_ 1783334570037346304
author Choudhari, Nikhil S.
Badakere, Swathi Vallabh
Richhariya, Ashutosh
Chittajallu, Sai Naga Sri Harsha
Senthil, Sirisha
Garudadri, Chandra Sekhar
author_facet Choudhari, Nikhil S.
Badakere, Swathi Vallabh
Richhariya, Ashutosh
Chittajallu, Sai Naga Sri Harsha
Senthil, Sirisha
Garudadri, Chandra Sekhar
author_sort Choudhari, Nikhil S.
collection PubMed
description PURPOSE: To investigate the mechanism of action and consistency in flow characteristics of the Ahmed glaucoma valve (AGV) under simulated physiological conditions in-vitro and to evaluate whether resistance during priming has any effect on performance of the device. METHODS: Each newly opened AGV device was connected to a digital manometer and was primed with normal saline. The device was then placed in a saline bath and connected to an open manometer, a digital manometer, and an infusion pump. Saline was infused at a rate of 3 μL/min for 24 hours. Digital manometer readings were recorded at 4 Hz. RESULTS: Data obtained from 9 devices are presented as medians (ranges). The priming pressure was 1130 (835, 1625) mm Hg. Pressure versus time curves showed two distinct phases; transient and steady phases. The transient phase peak pressure was 24 (13, 45) mm Hg. In the steady phase, opening and closing pressures were 13 (10, 17) and 7 (4, 9) mm Hg, respectively; the valve leaflets briefly opened every 73.9 (51, 76.6) minutes and the fluctuation of pressure (difference between opening and closing pressures) was 6 (3, 9) mm Hg. The Spearman correlation coefficient between priming and opening and priming and closing pressure was ρ = −0.13 (P = 0.72) and ρ = −0.36 (P = 0.33), respectively. CONCLUSIONS: The device showed functionality like a valve. The resistance during priming did not affect opening and closing pressures of the AGV. This study showed variable in vitro performance of the AGV. TRANSLATIONAL RELEVANCE: These laboratory findings might, at least partly, explain the variability in the clinical outcome of the device.
format Online
Article
Text
id pubmed-6016431
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-60164312018-06-26 Is Ahmed Glaucoma Valve Consistent in Performance? Choudhari, Nikhil S. Badakere, Swathi Vallabh Richhariya, Ashutosh Chittajallu, Sai Naga Sri Harsha Senthil, Sirisha Garudadri, Chandra Sekhar Transl Vis Sci Technol Articles PURPOSE: To investigate the mechanism of action and consistency in flow characteristics of the Ahmed glaucoma valve (AGV) under simulated physiological conditions in-vitro and to evaluate whether resistance during priming has any effect on performance of the device. METHODS: Each newly opened AGV device was connected to a digital manometer and was primed with normal saline. The device was then placed in a saline bath and connected to an open manometer, a digital manometer, and an infusion pump. Saline was infused at a rate of 3 μL/min for 24 hours. Digital manometer readings were recorded at 4 Hz. RESULTS: Data obtained from 9 devices are presented as medians (ranges). The priming pressure was 1130 (835, 1625) mm Hg. Pressure versus time curves showed two distinct phases; transient and steady phases. The transient phase peak pressure was 24 (13, 45) mm Hg. In the steady phase, opening and closing pressures were 13 (10, 17) and 7 (4, 9) mm Hg, respectively; the valve leaflets briefly opened every 73.9 (51, 76.6) minutes and the fluctuation of pressure (difference between opening and closing pressures) was 6 (3, 9) mm Hg. The Spearman correlation coefficient between priming and opening and priming and closing pressure was ρ = −0.13 (P = 0.72) and ρ = −0.36 (P = 0.33), respectively. CONCLUSIONS: The device showed functionality like a valve. The resistance during priming did not affect opening and closing pressures of the AGV. This study showed variable in vitro performance of the AGV. TRANSLATIONAL RELEVANCE: These laboratory findings might, at least partly, explain the variability in the clinical outcome of the device. The Association for Research in Vision and Ophthalmology 2018-06-22 /pmc/articles/PMC6016431/ /pubmed/29946493 http://dx.doi.org/10.1167/tvst.7.3.19 Text en Copyright 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Articles
Choudhari, Nikhil S.
Badakere, Swathi Vallabh
Richhariya, Ashutosh
Chittajallu, Sai Naga Sri Harsha
Senthil, Sirisha
Garudadri, Chandra Sekhar
Is Ahmed Glaucoma Valve Consistent in Performance?
title Is Ahmed Glaucoma Valve Consistent in Performance?
title_full Is Ahmed Glaucoma Valve Consistent in Performance?
title_fullStr Is Ahmed Glaucoma Valve Consistent in Performance?
title_full_unstemmed Is Ahmed Glaucoma Valve Consistent in Performance?
title_short Is Ahmed Glaucoma Valve Consistent in Performance?
title_sort is ahmed glaucoma valve consistent in performance?
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6016431/
https://www.ncbi.nlm.nih.gov/pubmed/29946493
http://dx.doi.org/10.1167/tvst.7.3.19
work_keys_str_mv AT choudharinikhils isahmedglaucomavalveconsistentinperformance
AT badakereswathivallabh isahmedglaucomavalveconsistentinperformance
AT richhariyaashutosh isahmedglaucomavalveconsistentinperformance
AT chittajallusainagasriharsha isahmedglaucomavalveconsistentinperformance
AT senthilsirisha isahmedglaucomavalveconsistentinperformance
AT garudadrichandrasekhar isahmedglaucomavalveconsistentinperformance