Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans

The delivery consistency of a Varian Edge linear accelerator over the entire course of treatment for nasopharynx carcinoma (NPC) and prostate cancer intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment plans was investigated using four different approach...

Descripción completa

Detalles Bibliográficos
Autores principales: Woon, Wui Ann, Ravindran, Paul B., Ekayanake, Piyasiri, Lim, Yivonne Yih Fang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6020617/
https://www.ncbi.nlm.nih.gov/pubmed/29962690
http://dx.doi.org/10.4103/jmp.JMP_109_17
_version_ 1783335336163672064
author Woon, Wui Ann
Ravindran, Paul B.
Ekayanake, Piyasiri
Lim, Yivonne Yih Fang
author_facet Woon, Wui Ann
Ravindran, Paul B.
Ekayanake, Piyasiri
Lim, Yivonne Yih Fang
author_sort Woon, Wui Ann
collection PubMed
description The delivery consistency of a Varian Edge linear accelerator over the entire course of treatment for nasopharynx carcinoma (NPC) and prostate cancer intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment plans was investigated using four different approaches. Three NPCs and three prostate plans were delivered in 34 and 29 consecutive days, respectively, using a Varian Edge equipped with a 120 high-definition (HD) multileaf collimator (MLC). All deliveries were measured with an electronic portal imaging device (EPID), and MapCheck2 and ArcCheck commercial systems with gamma analysis used to compare the results of all daily measurements against the pretreatment patient-specific quality assurance. The daily log files generated were also assessed for differences between the actual and planned doses using an in-house program to replace the original values in the DICOM plan files with the delivered parameter values from the log file, and then exporting the plans back to the treatment planning system for reconstruction of the actual dose delivered. The trajectory log file and EPID methods showed very good agreement, with minimal deviations between the daily delivered and reference doses. However, comparisons of the MapCheck2 and ArcCheck with the EPID revealed statistically significant differences (P < 0.001, one-tailed) with greater daily fluctuations, raising concerns over the performance, and reliability of the MapCheck2 and ArcCheck systems when being used to identify IMRT and VMAT plans with poor dosimetric accuracy. We conclude that the Varian Edge linear accelerator equipped with a 120 HD MLC can consistently deliver IMRT and VMAT plans over the entire treatment course.
format Online
Article
Text
id pubmed-6020617
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-60206172018-06-29 Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans Woon, Wui Ann Ravindran, Paul B. Ekayanake, Piyasiri Lim, Yivonne Yih Fang J Med Phys Technical Note The delivery consistency of a Varian Edge linear accelerator over the entire course of treatment for nasopharynx carcinoma (NPC) and prostate cancer intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment plans was investigated using four different approaches. Three NPCs and three prostate plans were delivered in 34 and 29 consecutive days, respectively, using a Varian Edge equipped with a 120 high-definition (HD) multileaf collimator (MLC). All deliveries were measured with an electronic portal imaging device (EPID), and MapCheck2 and ArcCheck commercial systems with gamma analysis used to compare the results of all daily measurements against the pretreatment patient-specific quality assurance. The daily log files generated were also assessed for differences between the actual and planned doses using an in-house program to replace the original values in the DICOM plan files with the delivered parameter values from the log file, and then exporting the plans back to the treatment planning system for reconstruction of the actual dose delivered. The trajectory log file and EPID methods showed very good agreement, with minimal deviations between the daily delivered and reference doses. However, comparisons of the MapCheck2 and ArcCheck with the EPID revealed statistically significant differences (P < 0.001, one-tailed) with greater daily fluctuations, raising concerns over the performance, and reliability of the MapCheck2 and ArcCheck systems when being used to identify IMRT and VMAT plans with poor dosimetric accuracy. We conclude that the Varian Edge linear accelerator equipped with a 120 HD MLC can consistently deliver IMRT and VMAT plans over the entire treatment course. Medknow Publications & Media Pvt Ltd 2018 /pmc/articles/PMC6020617/ /pubmed/29962690 http://dx.doi.org/10.4103/jmp.JMP_109_17 Text en Copyright: © 2018 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Technical Note
Woon, Wui Ann
Ravindran, Paul B.
Ekayanake, Piyasiri
Lim, Yivonne Yih Fang
Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title_full Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title_fullStr Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title_full_unstemmed Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title_short Validation of Delivery Consistency for Intensity-Modulated Radiation Therapy and Volumetric-Modulated Arc Therapy Plans
title_sort validation of delivery consistency for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6020617/
https://www.ncbi.nlm.nih.gov/pubmed/29962690
http://dx.doi.org/10.4103/jmp.JMP_109_17
work_keys_str_mv AT woonwuiann validationofdeliveryconsistencyforintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapyplans
AT ravindranpaulb validationofdeliveryconsistencyforintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapyplans
AT ekayanakepiyasiri validationofdeliveryconsistencyforintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapyplans
AT limyivonneyihfang validationofdeliveryconsistencyforintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapyplans