Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner
PURPOSE: The purpose of this study is to compare computed tomography (CT) radiation dose measurement methods proposed by TG111, International Electrotechnical Commission (IEC), and a direct dose profile integral (DPI) measurement method. METHODS: Pencil and Farmer ion chambers are used for integrati...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6020618/ https://www.ncbi.nlm.nih.gov/pubmed/29962685 http://dx.doi.org/10.4103/jmp.JMP_129_17 |
_version_ | 1783335336387018752 |
---|---|
author | Weir, Victor J. Zhang, Jie |
author_facet | Weir, Victor J. Zhang, Jie |
author_sort | Weir, Victor J. |
collection | PubMed |
description | PURPOSE: The purpose of this study is to compare computed tomography (CT) radiation dose measurement methods proposed by TG111, International Electrotechnical Commission (IEC), and a direct dose profile integral (DPI) measurement method. METHODS: Pencil and Farmer ion chambers are used for integrating dose profiles at different beam widths in a 60 cm long body phantom. Resulting DPI is used to calculate CT dose index (CTDI) at each beam width. Measurements are also done for a pencil chamber inserted into a 15 cm body phantom at the reference beam width. The reference measurement is scaled with pencil chamber measurements in air at different beam widths, according to the IEC approach. Finally, point dose measurements are done with a Farmer chamber under equilibrium conditions according to the TG111 method. All CTDIs calculated from measured data are compared to the scanner displayed CTDIs. RESULTS: Calculated CTDIs, at different beam widths, using the IEC approach are within 20% of CTDIs calculated from DPI measurements in a 60 cm long body phantom. Dose Length Integral (DLI) obtained from TG111 method is close to the results obtained from DPI measurements. Scanner displayed CTDIs are lower than all measured values by up to 38% at the techniques used. CONCLUSION: Although the IEC method is the easiest to use compared to the TG111 and direct DPI measurement method, it underestimates dose indices by about 20%. CTDIs displayed on the GE scanner are lower than those measured in this study by up to 38%. |
format | Online Article Text |
id | pubmed-6020618 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-60206182018-06-29 Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner Weir, Victor J. Zhang, Jie J Med Phys Original Article PURPOSE: The purpose of this study is to compare computed tomography (CT) radiation dose measurement methods proposed by TG111, International Electrotechnical Commission (IEC), and a direct dose profile integral (DPI) measurement method. METHODS: Pencil and Farmer ion chambers are used for integrating dose profiles at different beam widths in a 60 cm long body phantom. Resulting DPI is used to calculate CT dose index (CTDI) at each beam width. Measurements are also done for a pencil chamber inserted into a 15 cm body phantom at the reference beam width. The reference measurement is scaled with pencil chamber measurements in air at different beam widths, according to the IEC approach. Finally, point dose measurements are done with a Farmer chamber under equilibrium conditions according to the TG111 method. All CTDIs calculated from measured data are compared to the scanner displayed CTDIs. RESULTS: Calculated CTDIs, at different beam widths, using the IEC approach are within 20% of CTDIs calculated from DPI measurements in a 60 cm long body phantom. Dose Length Integral (DLI) obtained from TG111 method is close to the results obtained from DPI measurements. Scanner displayed CTDIs are lower than all measured values by up to 38% at the techniques used. CONCLUSION: Although the IEC method is the easiest to use compared to the TG111 and direct DPI measurement method, it underestimates dose indices by about 20%. CTDIs displayed on the GE scanner are lower than those measured in this study by up to 38%. Medknow Publications & Media Pvt Ltd 2018 /pmc/articles/PMC6020618/ /pubmed/29962685 http://dx.doi.org/10.4103/jmp.JMP_129_17 Text en Copyright: © 2018 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Weir, Victor J. Zhang, Jie Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title | Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title_full | Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title_fullStr | Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title_full_unstemmed | Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title_short | Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner |
title_sort | radiation dose measurements in a 256-slice computed tomography scanner |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6020618/ https://www.ncbi.nlm.nih.gov/pubmed/29962685 http://dx.doi.org/10.4103/jmp.JMP_129_17 |
work_keys_str_mv | AT weirvictorj radiationdosemeasurementsina256slicecomputedtomographyscanner AT zhangjie radiationdosemeasurementsina256slicecomputedtomographyscanner |