Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements

BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between m...

Descripción completa

Detalles Bibliográficos
Autores principales: Horehledova, Barbora, Mihl, Casper, Schwemmer, Chris, Hendriks, Babs M. F., Eijsvoogel, Nienke G., Kietselaer, Bastiaan L. J. H., Wildberger, Joachim E., Das, Marco
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6023104/
https://www.ncbi.nlm.nih.gov/pubmed/29953507
http://dx.doi.org/10.1371/journal.pone.0199732
_version_ 1783335794668208128
author Horehledova, Barbora
Mihl, Casper
Schwemmer, Chris
Hendriks, Babs M. F.
Eijsvoogel, Nienke G.
Kietselaer, Bastiaan L. J. H.
Wildberger, Joachim E.
Das, Marco
author_facet Horehledova, Barbora
Mihl, Casper
Schwemmer, Chris
Hendriks, Babs M. F.
Eijsvoogel, Nienke G.
Kietselaer, Bastiaan L. J. H.
Wildberger, Joachim E.
Das, Marco
author_sort Horehledova, Barbora
collection PubMed
description BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time. METHODS: One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2(nd) and 3(rd) generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis. RESULTS: The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90–0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively. CONCLUSION: Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower.
format Online
Article
Text
id pubmed-6023104
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-60231042018-07-07 Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements Horehledova, Barbora Mihl, Casper Schwemmer, Chris Hendriks, Babs M. F. Eijsvoogel, Nienke G. Kietselaer, Bastiaan L. J. H. Wildberger, Joachim E. Das, Marco PLoS One Research Article BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time. METHODS: One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2(nd) and 3(rd) generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis. RESULTS: The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90–0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively. CONCLUSION: Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower. Public Library of Science 2018-06-28 /pmc/articles/PMC6023104/ /pubmed/29953507 http://dx.doi.org/10.1371/journal.pone.0199732 Text en © 2018 Horehledova et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Horehledova, Barbora
Mihl, Casper
Schwemmer, Chris
Hendriks, Babs M. F.
Eijsvoogel, Nienke G.
Kietselaer, Bastiaan L. J. H.
Wildberger, Joachim E.
Das, Marco
Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title_full Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title_fullStr Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title_full_unstemmed Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title_short Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
title_sort aortic root evaluation prior to transcatheter aortic valve implantation—correlation of manual and semi-automatic measurements
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6023104/
https://www.ncbi.nlm.nih.gov/pubmed/29953507
http://dx.doi.org/10.1371/journal.pone.0199732
work_keys_str_mv AT horehledovabarbora aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT mihlcasper aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT schwemmerchris aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT hendriksbabsmf aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT eijsvoogelnienkeg aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT kietselaerbastiaanljh aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT wildbergerjoachime aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements
AT dasmarco aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements