Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements
BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between m...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6023104/ https://www.ncbi.nlm.nih.gov/pubmed/29953507 http://dx.doi.org/10.1371/journal.pone.0199732 |
_version_ | 1783335794668208128 |
---|---|
author | Horehledova, Barbora Mihl, Casper Schwemmer, Chris Hendriks, Babs M. F. Eijsvoogel, Nienke G. Kietselaer, Bastiaan L. J. H. Wildberger, Joachim E. Das, Marco |
author_facet | Horehledova, Barbora Mihl, Casper Schwemmer, Chris Hendriks, Babs M. F. Eijsvoogel, Nienke G. Kietselaer, Bastiaan L. J. H. Wildberger, Joachim E. Das, Marco |
author_sort | Horehledova, Barbora |
collection | PubMed |
description | BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time. METHODS: One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2(nd) and 3(rd) generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis. RESULTS: The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90–0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively. CONCLUSION: Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower. |
format | Online Article Text |
id | pubmed-6023104 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-60231042018-07-07 Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements Horehledova, Barbora Mihl, Casper Schwemmer, Chris Hendriks, Babs M. F. Eijsvoogel, Nienke G. Kietselaer, Bastiaan L. J. H. Wildberger, Joachim E. Das, Marco PLoS One Research Article BACKGROUND: Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time. METHODS: One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2(nd) and 3(rd) generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis. RESULTS: The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90–0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively. CONCLUSION: Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower. Public Library of Science 2018-06-28 /pmc/articles/PMC6023104/ /pubmed/29953507 http://dx.doi.org/10.1371/journal.pone.0199732 Text en © 2018 Horehledova et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Horehledova, Barbora Mihl, Casper Schwemmer, Chris Hendriks, Babs M. F. Eijsvoogel, Nienke G. Kietselaer, Bastiaan L. J. H. Wildberger, Joachim E. Das, Marco Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title | Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title_full | Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title_fullStr | Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title_full_unstemmed | Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title_short | Aortic root evaluation prior to transcatheter aortic valve implantation—Correlation of manual and semi-automatic measurements |
title_sort | aortic root evaluation prior to transcatheter aortic valve implantation—correlation of manual and semi-automatic measurements |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6023104/ https://www.ncbi.nlm.nih.gov/pubmed/29953507 http://dx.doi.org/10.1371/journal.pone.0199732 |
work_keys_str_mv | AT horehledovabarbora aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT mihlcasper aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT schwemmerchris aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT hendriksbabsmf aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT eijsvoogelnienkeg aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT kietselaerbastiaanljh aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT wildbergerjoachime aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements AT dasmarco aorticrootevaluationpriortotranscatheteraorticvalveimplantationcorrelationofmanualandsemiautomaticmeasurements |