Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing scheme...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6032685/ https://www.ncbi.nlm.nih.gov/pubmed/29941712 http://dx.doi.org/10.4103/0366-6999.235109 |
_version_ | 1783337546616406016 |
---|---|
author | Zhao, Hai-Jian Ge, Meng-Lei Yan, Yin Zhang, Tian-Jiao Zeng, Jie Zhou, Wei-Yan Wang, Yu-Fei Meng, Qing-Hui Zhang, Chuan-Bao |
author_facet | Zhao, Hai-Jian Ge, Meng-Lei Yan, Yin Zhang, Tian-Jiao Zeng, Jie Zhou, Wei-Yan Wang, Yu-Fei Meng, Qing-Hui Zhang, Chuan-Bao |
author_sort | Zhao, Hai-Jian |
collection | PubMed |
description | BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials. METHODS: Inductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays. RESULTS: All assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from −0.13 mmol/L to 0.00 mmol/L (−5.61–0.01%), and the ranges for the three concentrations were −0.10–0.04 mmol/L (−5.71–2.35%), −0.14–−0.01 mmol/L (−5.80–−0.30%), and −0.19–0.04 mmol/L (−6.24–1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays. CONCLUSIONS: Biases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect. |
format | Online Article Text |
id | pubmed-6032685 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-60326852018-07-20 Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements Zhao, Hai-Jian Ge, Meng-Lei Yan, Yin Zhang, Tian-Jiao Zeng, Jie Zhou, Wei-Yan Wang, Yu-Fei Meng, Qing-Hui Zhang, Chuan-Bao Chin Med J (Engl) Original Article BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials. METHODS: Inductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays. RESULTS: All assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from −0.13 mmol/L to 0.00 mmol/L (−5.61–0.01%), and the ranges for the three concentrations were −0.10–0.04 mmol/L (−5.71–2.35%), −0.14–−0.01 mmol/L (−5.80–−0.30%), and −0.19–0.04 mmol/L (−6.24–1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays. CONCLUSIONS: Biases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect. Medknow Publications & Media Pvt Ltd 2018-07-05 /pmc/articles/PMC6032685/ /pubmed/29941712 http://dx.doi.org/10.4103/0366-6999.235109 Text en Copyright: © 2018 Chinese Medical Journal http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Zhao, Hai-Jian Ge, Meng-Lei Yan, Yin Zhang, Tian-Jiao Zeng, Jie Zhou, Wei-Yan Wang, Yu-Fei Meng, Qing-Hui Zhang, Chuan-Bao Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title | Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title_full | Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title_fullStr | Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title_full_unstemmed | Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title_short | Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements |
title_sort | inductively coupled plasma mass spectrometry as a reference method to evaluate serum calcium measurement bias and the commutability of processed materials during routine measurements |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6032685/ https://www.ncbi.nlm.nih.gov/pubmed/29941712 http://dx.doi.org/10.4103/0366-6999.235109 |
work_keys_str_mv | AT zhaohaijian inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT gemenglei inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT yanyin inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT zhangtianjiao inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT zengjie inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT zhouweiyan inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT wangyufei inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT mengqinghui inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements AT zhangchuanbao inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements |