Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements

BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing scheme...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhao, Hai-Jian, Ge, Meng-Lei, Yan, Yin, Zhang, Tian-Jiao, Zeng, Jie, Zhou, Wei-Yan, Wang, Yu-Fei, Meng, Qing-Hui, Zhang, Chuan-Bao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6032685/
https://www.ncbi.nlm.nih.gov/pubmed/29941712
http://dx.doi.org/10.4103/0366-6999.235109
_version_ 1783337546616406016
author Zhao, Hai-Jian
Ge, Meng-Lei
Yan, Yin
Zhang, Tian-Jiao
Zeng, Jie
Zhou, Wei-Yan
Wang, Yu-Fei
Meng, Qing-Hui
Zhang, Chuan-Bao
author_facet Zhao, Hai-Jian
Ge, Meng-Lei
Yan, Yin
Zhang, Tian-Jiao
Zeng, Jie
Zhou, Wei-Yan
Wang, Yu-Fei
Meng, Qing-Hui
Zhang, Chuan-Bao
author_sort Zhao, Hai-Jian
collection PubMed
description BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials. METHODS: Inductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays. RESULTS: All assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from −0.13 mmol/L to 0.00 mmol/L (−5.61–0.01%), and the ranges for the three concentrations were −0.10–0.04 mmol/L (−5.71–2.35%), −0.14–−0.01 mmol/L (−5.80–−0.30%), and −0.19–0.04 mmol/L (−6.24–1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays. CONCLUSIONS: Biases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect.
format Online
Article
Text
id pubmed-6032685
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-60326852018-07-20 Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements Zhao, Hai-Jian Ge, Meng-Lei Yan, Yin Zhang, Tian-Jiao Zeng, Jie Zhou, Wei-Yan Wang, Yu-Fei Meng, Qing-Hui Zhang, Chuan-Bao Chin Med J (Engl) Original Article BACKGROUND: Measuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials. METHODS: Inductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays. RESULTS: All assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from −0.13 mmol/L to 0.00 mmol/L (−5.61–0.01%), and the ranges for the three concentrations were −0.10–0.04 mmol/L (−5.71–2.35%), −0.14–−0.01 mmol/L (−5.80–−0.30%), and −0.19–0.04 mmol/L (−6.24–1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays. CONCLUSIONS: Biases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect. Medknow Publications & Media Pvt Ltd 2018-07-05 /pmc/articles/PMC6032685/ /pubmed/29941712 http://dx.doi.org/10.4103/0366-6999.235109 Text en Copyright: © 2018 Chinese Medical Journal http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Zhao, Hai-Jian
Ge, Meng-Lei
Yan, Yin
Zhang, Tian-Jiao
Zeng, Jie
Zhou, Wei-Yan
Wang, Yu-Fei
Meng, Qing-Hui
Zhang, Chuan-Bao
Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title_full Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title_fullStr Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title_full_unstemmed Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title_short Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
title_sort inductively coupled plasma mass spectrometry as a reference method to evaluate serum calcium measurement bias and the commutability of processed materials during routine measurements
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6032685/
https://www.ncbi.nlm.nih.gov/pubmed/29941712
http://dx.doi.org/10.4103/0366-6999.235109
work_keys_str_mv AT zhaohaijian inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT gemenglei inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT yanyin inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT zhangtianjiao inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT zengjie inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT zhouweiyan inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT wangyufei inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT mengqinghui inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements
AT zhangchuanbao inductivelycoupledplasmamassspectrometryasareferencemethodtoevaluateserumcalciummeasurementbiasandthecommutabilityofprocessedmaterialsduringroutinemeasurements