Internal quality assurance of HIL indices on Roche Cobas c702
Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6034854/ https://www.ncbi.nlm.nih.gov/pubmed/29979722 http://dx.doi.org/10.1371/journal.pone.0200088 |
_version_ | 1783337949242327040 |
---|---|
author | Lippi, Giuseppe Cadamuro, Janne Danese, Elisa Gelati, Matteo Montagnana, Martina von Meyer, Alexander Salvagno, Gian Luca Simundic, Ana-Maria |
author_facet | Lippi, Giuseppe Cadamuro, Janne Danese, Elisa Gelati, Matteo Montagnana, Martina von Meyer, Alexander Salvagno, Gian Luca Simundic, Ana-Maria |
author_sort | Lippi, Giuseppe |
collection | PubMed |
description | Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices. Pools containing different values of each of the three HIL indices were prepared from routine plasma samples, divided in aliquots and frozen at -20°C. Stability of frozen materials was assessed by thawing one aliquot of each pool after different days of freezing (1, 4, 8, 15, 22 and 29), and by measuring HIL indices on baseline fresh samples and frozen-thawed aliquots with Roche Cobas c702. Five fresh liquid IQCs materials were also measured at the same time points. Intra-assay and inter-assay imprecision of HIL indices calculated with commercial IQC materials ranged between 1.1–2.0% and 1.6–3.3%, respectively. When target values of HIL indices were calculated using frozen-thawed aliquots, the inter-assay imprecision of in-house prepared materials was optimal, even better than that of commercial liquid IQCs (H-index, 0.8% versus 1.6%; L-index, 2.2% versus 2.5%; I-index, 0.8% versus 3.3%). In conclusion, in-house prepared IQC materials are cost-effective alternatives to commercial liquid IQCs for HIL quality assurance. |
format | Online Article Text |
id | pubmed-6034854 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-60348542018-07-19 Internal quality assurance of HIL indices on Roche Cobas c702 Lippi, Giuseppe Cadamuro, Janne Danese, Elisa Gelati, Matteo Montagnana, Martina von Meyer, Alexander Salvagno, Gian Luca Simundic, Ana-Maria PLoS One Research Article Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices. Pools containing different values of each of the three HIL indices were prepared from routine plasma samples, divided in aliquots and frozen at -20°C. Stability of frozen materials was assessed by thawing one aliquot of each pool after different days of freezing (1, 4, 8, 15, 22 and 29), and by measuring HIL indices on baseline fresh samples and frozen-thawed aliquots with Roche Cobas c702. Five fresh liquid IQCs materials were also measured at the same time points. Intra-assay and inter-assay imprecision of HIL indices calculated with commercial IQC materials ranged between 1.1–2.0% and 1.6–3.3%, respectively. When target values of HIL indices were calculated using frozen-thawed aliquots, the inter-assay imprecision of in-house prepared materials was optimal, even better than that of commercial liquid IQCs (H-index, 0.8% versus 1.6%; L-index, 2.2% versus 2.5%; I-index, 0.8% versus 3.3%). In conclusion, in-house prepared IQC materials are cost-effective alternatives to commercial liquid IQCs for HIL quality assurance. Public Library of Science 2018-07-06 /pmc/articles/PMC6034854/ /pubmed/29979722 http://dx.doi.org/10.1371/journal.pone.0200088 Text en © 2018 Lippi et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Lippi, Giuseppe Cadamuro, Janne Danese, Elisa Gelati, Matteo Montagnana, Martina von Meyer, Alexander Salvagno, Gian Luca Simundic, Ana-Maria Internal quality assurance of HIL indices on Roche Cobas c702 |
title | Internal quality assurance of HIL indices on Roche Cobas c702 |
title_full | Internal quality assurance of HIL indices on Roche Cobas c702 |
title_fullStr | Internal quality assurance of HIL indices on Roche Cobas c702 |
title_full_unstemmed | Internal quality assurance of HIL indices on Roche Cobas c702 |
title_short | Internal quality assurance of HIL indices on Roche Cobas c702 |
title_sort | internal quality assurance of hil indices on roche cobas c702 |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6034854/ https://www.ncbi.nlm.nih.gov/pubmed/29979722 http://dx.doi.org/10.1371/journal.pone.0200088 |
work_keys_str_mv | AT lippigiuseppe internalqualityassuranceofhilindicesonrochecobasc702 AT cadamurojanne internalqualityassuranceofhilindicesonrochecobasc702 AT daneseelisa internalqualityassuranceofhilindicesonrochecobasc702 AT gelatimatteo internalqualityassuranceofhilindicesonrochecobasc702 AT montagnanamartina internalqualityassuranceofhilindicesonrochecobasc702 AT vonmeyeralexander internalqualityassuranceofhilindicesonrochecobasc702 AT salvagnogianluca internalqualityassuranceofhilindicesonrochecobasc702 AT simundicanamaria internalqualityassuranceofhilindicesonrochecobasc702 |