Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure
A novel phantomless, EPID‐based method of measuring the beam focal spot offset of a linear accelerator was proposed and validated for Varian machines. In this method, one set of jaws and the MLC were utilized to form a symmetric field and then a 180(o) collimator rotation was utilized to determine t...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036348/ https://www.ncbi.nlm.nih.gov/pubmed/29761625 http://dx.doi.org/10.1002/acm2.12344 |
_version_ | 1783338149067358208 |
---|---|
author | Chojnowski, Jacek M. Taylor, Lee M. Sykes, Jonathan R. Thwaites, David I. |
author_facet | Chojnowski, Jacek M. Taylor, Lee M. Sykes, Jonathan R. Thwaites, David I. |
author_sort | Chojnowski, Jacek M. |
collection | PubMed |
description | A novel phantomless, EPID‐based method of measuring the beam focal spot offset of a linear accelerator was proposed and validated for Varian machines. In this method, one set of jaws and the MLC were utilized to form a symmetric field and then a 180(o) collimator rotation was utilized to determine the radiation isocenter defined by the jaws and the MLC, respectively. The difference between these two isocentres is directly correlated with the beam focal spot offset of the linear accelerator. In the current work, the method has been considered for Elekta linacs. An Elekta linac with the Agility(®) head does not have two set of jaws, therefore, a modified method is presented making use of one set of diaphragms, the MLC and a full 360(o) collimator rotation. The modified method has been tested on two Elekta Synergy(®) linacs with Agility(®) heads and independently validated. A practical guide with instructions and a MATLAB (®) code is attached for easy implementation. |
format | Online Article Text |
id | pubmed-6036348 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-60363482018-07-12 Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure Chojnowski, Jacek M. Taylor, Lee M. Sykes, Jonathan R. Thwaites, David I. J Appl Clin Med Phys Radiation Oncology Physics A novel phantomless, EPID‐based method of measuring the beam focal spot offset of a linear accelerator was proposed and validated for Varian machines. In this method, one set of jaws and the MLC were utilized to form a symmetric field and then a 180(o) collimator rotation was utilized to determine the radiation isocenter defined by the jaws and the MLC, respectively. The difference between these two isocentres is directly correlated with the beam focal spot offset of the linear accelerator. In the current work, the method has been considered for Elekta linacs. An Elekta linac with the Agility(®) head does not have two set of jaws, therefore, a modified method is presented making use of one set of diaphragms, the MLC and a full 360(o) collimator rotation. The modified method has been tested on two Elekta Synergy(®) linacs with Agility(®) heads and independently validated. A practical guide with instructions and a MATLAB (®) code is attached for easy implementation. John Wiley and Sons Inc. 2018-05-14 /pmc/articles/PMC6036348/ /pubmed/29761625 http://dx.doi.org/10.1002/acm2.12344 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Chojnowski, Jacek M. Taylor, Lee M. Sykes, Jonathan R. Thwaites, David I. Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title | Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title_full | Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title_fullStr | Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title_full_unstemmed | Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title_short | Beam focal spot position determination for an Elekta linac with the Agility(®) head; practical guide with a ready‐to‐go procedure |
title_sort | beam focal spot position determination for an elekta linac with the agility(®) head; practical guide with a ready‐to‐go procedure |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036348/ https://www.ncbi.nlm.nih.gov/pubmed/29761625 http://dx.doi.org/10.1002/acm2.12344 |
work_keys_str_mv | AT chojnowskijacekm beamfocalspotpositiondeterminationforanelektalinacwiththeagilityheadpracticalguidewithareadytogoprocedure AT taylorleem beamfocalspotpositiondeterminationforanelektalinacwiththeagilityheadpracticalguidewithareadytogoprocedure AT sykesjonathanr beamfocalspotpositiondeterminationforanelektalinacwiththeagilityheadpracticalguidewithareadytogoprocedure AT thwaitesdavidi beamfocalspotpositiondeterminationforanelektalinacwiththeagilityheadpracticalguidewithareadytogoprocedure |