Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy

PURPOSE: Radiation treatments delivered with real‐time multileaf collimator (MLC) tracking currently lack fast pretreatment or real‐time quality assurance. The purpose of this study is to test a 2D silicon detector, MagicPlate‐512 (MP512), in a complex clinical environment involving real‐time reconf...

Descripción completa

Detalles Bibliográficos
Autores principales: Duncan, Mitchell, Newall, Matthew K., Caillet, Vincent, Booth, Jeremy T., Keall, Paul J., Lerch, Michael, Perevertaylo, Vladimir, Rosenfeld, Anatoly B., Petasecca, Marco
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036363/
https://www.ncbi.nlm.nih.gov/pubmed/29873185
http://dx.doi.org/10.1002/acm2.12364
_version_ 1783338152323186688
author Duncan, Mitchell
Newall, Matthew K.
Caillet, Vincent
Booth, Jeremy T.
Keall, Paul J.
Lerch, Michael
Perevertaylo, Vladimir
Rosenfeld, Anatoly B.
Petasecca, Marco
author_facet Duncan, Mitchell
Newall, Matthew K.
Caillet, Vincent
Booth, Jeremy T.
Keall, Paul J.
Lerch, Michael
Perevertaylo, Vladimir
Rosenfeld, Anatoly B.
Petasecca, Marco
author_sort Duncan, Mitchell
collection PubMed
description PURPOSE: Radiation treatments delivered with real‐time multileaf collimator (MLC) tracking currently lack fast pretreatment or real‐time quality assurance. The purpose of this study is to test a 2D silicon detector, MagicPlate‐512 (MP512), in a complex clinical environment involving real‐time reconfiguration of the MLC leaves during target tracking. METHODS: MP512 was placed in the center of a solid water phantom and mounted on a motion platform used to simulate three different patient motions. Electromagnetic target tracking was implemented using the Calypso system (Varian Medical Systems, Palo Alto, CA, USA) and an MLC tracking software. A two‐arc VMAT plan was delivered and 2D dose distributions were reconstructed by MP512, EBT3 film, and the Eclipse treatment planning system (TPS). Dose maps were compared using gamma analysis with 2%/2 mm and 3%/3 mm acceptance criteria. Dose profiles were generated in sup‐inf and lateral directions to show the agreement of MP512 to EBT3 and to highlight the efficacy of the MLC tracking system in mitigating the effect of the simulated patient motion. RESULTS: Using a 3%/3 mm acceptance criterion for 2D gamma analysis, MP512 to EBT3 film agreement was 99% and MP512 to TPS agreement was 100%. For a 2%/2 mm criterion, the agreement was 95% and 98%, respectively. Full width at half maximum and 80%/20% penumbral width of the MP512 and EBT3 dose profiles agreed within 1 mm and 0.5 mm, respectively. Patient motion increased the measured dose profile penumbral width by nearly 2 mm (with respect to the no‐motion case); however, the MLC tracking strategy was able to mitigate 80% of this effect. CONCLUSIONS: MP512 is capable of high spatial resolution 2D dose reconstruction during adaptive MLC tracking, including arc deliveries. It shows potential as an effective tool for 2D small field dosimetry and pretreatment quality assurance for MLC tracking modalities. These results provide confidence that detector‐based pretreatment dosimetry is clinically feasible despite fast real‐time MLC reconfigurations.
format Online
Article
Text
id pubmed-6036363
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-60363632018-07-12 Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy Duncan, Mitchell Newall, Matthew K. Caillet, Vincent Booth, Jeremy T. Keall, Paul J. Lerch, Michael Perevertaylo, Vladimir Rosenfeld, Anatoly B. Petasecca, Marco J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: Radiation treatments delivered with real‐time multileaf collimator (MLC) tracking currently lack fast pretreatment or real‐time quality assurance. The purpose of this study is to test a 2D silicon detector, MagicPlate‐512 (MP512), in a complex clinical environment involving real‐time reconfiguration of the MLC leaves during target tracking. METHODS: MP512 was placed in the center of a solid water phantom and mounted on a motion platform used to simulate three different patient motions. Electromagnetic target tracking was implemented using the Calypso system (Varian Medical Systems, Palo Alto, CA, USA) and an MLC tracking software. A two‐arc VMAT plan was delivered and 2D dose distributions were reconstructed by MP512, EBT3 film, and the Eclipse treatment planning system (TPS). Dose maps were compared using gamma analysis with 2%/2 mm and 3%/3 mm acceptance criteria. Dose profiles were generated in sup‐inf and lateral directions to show the agreement of MP512 to EBT3 and to highlight the efficacy of the MLC tracking system in mitigating the effect of the simulated patient motion. RESULTS: Using a 3%/3 mm acceptance criterion for 2D gamma analysis, MP512 to EBT3 film agreement was 99% and MP512 to TPS agreement was 100%. For a 2%/2 mm criterion, the agreement was 95% and 98%, respectively. Full width at half maximum and 80%/20% penumbral width of the MP512 and EBT3 dose profiles agreed within 1 mm and 0.5 mm, respectively. Patient motion increased the measured dose profile penumbral width by nearly 2 mm (with respect to the no‐motion case); however, the MLC tracking strategy was able to mitigate 80% of this effect. CONCLUSIONS: MP512 is capable of high spatial resolution 2D dose reconstruction during adaptive MLC tracking, including arc deliveries. It shows potential as an effective tool for 2D small field dosimetry and pretreatment quality assurance for MLC tracking modalities. These results provide confidence that detector‐based pretreatment dosimetry is clinically feasible despite fast real‐time MLC reconfigurations. John Wiley and Sons Inc. 2018-06-05 /pmc/articles/PMC6036363/ /pubmed/29873185 http://dx.doi.org/10.1002/acm2.12364 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Duncan, Mitchell
Newall, Matthew K.
Caillet, Vincent
Booth, Jeremy T.
Keall, Paul J.
Lerch, Michael
Perevertaylo, Vladimir
Rosenfeld, Anatoly B.
Petasecca, Marco
Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title_full Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title_fullStr Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title_full_unstemmed Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title_short Real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
title_sort real‐time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036363/
https://www.ncbi.nlm.nih.gov/pubmed/29873185
http://dx.doi.org/10.1002/acm2.12364
work_keys_str_mv AT duncanmitchell realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT newallmatthewk realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT cailletvincent realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT boothjeremyt realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT keallpaulj realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT lerchmichael realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT perevertaylovladimir realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT rosenfeldanatolyb realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy
AT petaseccamarco realtimehighspatialresolutiondoseverificationinstereotacticmotionadaptivearcradiotherapy