Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system

PURPOSE: To report a novel 3D‐printed device (“SH phantom”) that is designed for routine accuracy check of the Gamma Knife Icon High Definition Motion Management (HDMM) system. METHODS: SH phantom was designed using tinkerCAD software and printed on a commercial 3D printer. We evaluated the SH phant...

Descripción completa

Detalles Bibliográficos
Autores principales: Wu, Chuan, Radevic, Marlyn B., Glass, Jennifer S., Skubic, Stan E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036381/
https://www.ncbi.nlm.nih.gov/pubmed/29797396
http://dx.doi.org/10.1002/acm2.12339
_version_ 1783338156630736896
author Wu, Chuan
Radevic, Marlyn B.
Glass, Jennifer S.
Skubic, Stan E.
author_facet Wu, Chuan
Radevic, Marlyn B.
Glass, Jennifer S.
Skubic, Stan E.
author_sort Wu, Chuan
collection PubMed
description PURPOSE: To report a novel 3D‐printed device (“SH phantom”) that is designed for routine accuracy check of the Gamma Knife Icon High Definition Motion Management (HDMM) system. METHODS: SH phantom was designed using tinkerCAD software and printed on a commercial 3D printer. We evaluated the SH phantom on our Gamma Knife Icon unit regarding its usability and accuracy for routine HDMM QA. RESULTS: Single‐axis and multiple‐axis measurements validated the SH phantom design and implementation. An HDMM QA accuracy of 0.22 mm or better along single axis was found using SH phantom. CONCLUSIONS: The SH phantom proved to be a quick and simple tool to use to perform the HDMM system QA. The SH phantom was tested successfully and adopted by us as part of monthly QA for the Gamma Knife Icon.
format Online
Article
Text
id pubmed-6036381
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-60363812018-07-12 Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system Wu, Chuan Radevic, Marlyn B. Glass, Jennifer S. Skubic, Stan E. J Appl Clin Med Phys Technical Notes PURPOSE: To report a novel 3D‐printed device (“SH phantom”) that is designed for routine accuracy check of the Gamma Knife Icon High Definition Motion Management (HDMM) system. METHODS: SH phantom was designed using tinkerCAD software and printed on a commercial 3D printer. We evaluated the SH phantom on our Gamma Knife Icon unit regarding its usability and accuracy for routine HDMM QA. RESULTS: Single‐axis and multiple‐axis measurements validated the SH phantom design and implementation. An HDMM QA accuracy of 0.22 mm or better along single axis was found using SH phantom. CONCLUSIONS: The SH phantom proved to be a quick and simple tool to use to perform the HDMM system QA. The SH phantom was tested successfully and adopted by us as part of monthly QA for the Gamma Knife Icon. John Wiley and Sons Inc. 2018-05-23 /pmc/articles/PMC6036381/ /pubmed/29797396 http://dx.doi.org/10.1002/acm2.12339 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Wu, Chuan
Radevic, Marlyn B.
Glass, Jennifer S.
Skubic, Stan E.
Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title_full Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title_fullStr Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title_full_unstemmed Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title_short Technical note: A 3D‐printed phantom for routine accuracy check of Gamma Knife Icon HDMM system
title_sort technical note: a 3d‐printed phantom for routine accuracy check of gamma knife icon hdmm system
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6036381/
https://www.ncbi.nlm.nih.gov/pubmed/29797396
http://dx.doi.org/10.1002/acm2.12339
work_keys_str_mv AT wuchuan technicalnotea3dprintedphantomforroutineaccuracycheckofgammaknifeiconhdmmsystem
AT radevicmarlynb technicalnotea3dprintedphantomforroutineaccuracycheckofgammaknifeiconhdmmsystem
AT glassjennifers technicalnotea3dprintedphantomforroutineaccuracycheckofgammaknifeiconhdmmsystem
AT skubicstane technicalnotea3dprintedphantomforroutineaccuracycheckofgammaknifeiconhdmmsystem