Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators

OBJECTIVE(S): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators. METHODS: Data acquisition was performed using t...

Descripción completa

Detalles Bibliográficos
Autores principales: Morita, Keishin, Maebatake, Akira, Iwasaki, Rina, Shiotsuki, Yuki, Himuro, Kazuhiko, Baba, Shingo, Sasaki, Masayuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mashhad University of Medical Sciences 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6038970/
https://www.ncbi.nlm.nih.gov/pubmed/29998145
http://dx.doi.org/10.22038/aojnmb.2018.10330
_version_ 1783338599852277760
author Morita, Keishin
Maebatake, Akira
Iwasaki, Rina
Shiotsuki, Yuki
Himuro, Kazuhiko
Baba, Shingo
Sasaki, Masayuki
author_facet Morita, Keishin
Maebatake, Akira
Iwasaki, Rina
Shiotsuki, Yuki
Himuro, Kazuhiko
Baba, Shingo
Sasaki, Masayuki
author_sort Morita, Keishin
collection PubMed
description OBJECTIVE(S): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators. METHODS: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fan-beam (camera A-1 and B-1), low- and medium-energy general-purpose (camera A-2 and B-2), and low-energy high-resolution (camera A-3 and B-3) collimators. The SPECT images were reconstructed using filtered back projection (FBP) with Chang’s attenuation correction. However, the scatter correction was not performed. A pool phantom and a three-dimensional (3D) brain phantom were filled with (123)I solution to examine the reconstruction parameters. The optimal attenuation coefficient was based on the visual assessment of the profile curve, coefficient of variation (CV) [%], and summed difference from the reference activity of the pool phantom. The optimal Butterworth filter for the 3D-brain phantom was also determined based on a visual assessment. The anthropomorphic striatal phantom was filled with (123)I solution at striatum-to-background radioactivity ratios of 8, 6, 4, and 3. The specific binding ratio (SBR) of the striatum (calculated by the CT method) was used to compare the results with those of the parallel-hole collimators. RESULTS: The optimal attenuation coefficients were 0.09, 0.11, 0.05, 0.05, 0.11, and, 0.10 cm(-1) for cameras A-1, A-2, A-3, B-1, B-2, and B-3, respectively. The cutoff frequencies of the Butterworth filter were 0.32, 0.40, and 0.36 cycles/cm for camera A, and 0.46, 0.44, and 0.44 cycles/cm for camera B, respectively. The recovery rates of the SBR(mean) with camera A were 51.2%, 49.4%, and 45.6%, respectively. The difference was not statistically significant. The recovery rates of the SBR with camera B were 59.2%, 50.7%, and 50.8%, respectively. Camera B-1 showed significantly high SBR values. CONCLUSION: As the findings indicated, the optimal reconstruction parameters differed according to the devices and collimators. The fan beam collimator was found to provide promising results with each device.
format Online
Article
Text
id pubmed-6038970
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Mashhad University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-60389702018-07-11 Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators Morita, Keishin Maebatake, Akira Iwasaki, Rina Shiotsuki, Yuki Himuro, Kazuhiko Baba, Shingo Sasaki, Masayuki Asia Ocean J Nucl Med Biol Original Article OBJECTIVE(S): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators. METHODS: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fan-beam (camera A-1 and B-1), low- and medium-energy general-purpose (camera A-2 and B-2), and low-energy high-resolution (camera A-3 and B-3) collimators. The SPECT images were reconstructed using filtered back projection (FBP) with Chang’s attenuation correction. However, the scatter correction was not performed. A pool phantom and a three-dimensional (3D) brain phantom were filled with (123)I solution to examine the reconstruction parameters. The optimal attenuation coefficient was based on the visual assessment of the profile curve, coefficient of variation (CV) [%], and summed difference from the reference activity of the pool phantom. The optimal Butterworth filter for the 3D-brain phantom was also determined based on a visual assessment. The anthropomorphic striatal phantom was filled with (123)I solution at striatum-to-background radioactivity ratios of 8, 6, 4, and 3. The specific binding ratio (SBR) of the striatum (calculated by the CT method) was used to compare the results with those of the parallel-hole collimators. RESULTS: The optimal attenuation coefficients were 0.09, 0.11, 0.05, 0.05, 0.11, and, 0.10 cm(-1) for cameras A-1, A-2, A-3, B-1, B-2, and B-3, respectively. The cutoff frequencies of the Butterworth filter were 0.32, 0.40, and 0.36 cycles/cm for camera A, and 0.46, 0.44, and 0.44 cycles/cm for camera B, respectively. The recovery rates of the SBR(mean) with camera A were 51.2%, 49.4%, and 45.6%, respectively. The difference was not statistically significant. The recovery rates of the SBR with camera B were 59.2%, 50.7%, and 50.8%, respectively. Camera B-1 showed significantly high SBR values. CONCLUSION: As the findings indicated, the optimal reconstruction parameters differed according to the devices and collimators. The fan beam collimator was found to provide promising results with each device. Mashhad University of Medical Sciences 2018 /pmc/articles/PMC6038970/ /pubmed/29998145 http://dx.doi.org/10.22038/aojnmb.2018.10330 Text en © 2018 mums.ac.ir All rights reserved. This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Morita, Keishin
Maebatake, Akira
Iwasaki, Rina
Shiotsuki, Yuki
Himuro, Kazuhiko
Baba, Shingo
Sasaki, Masayuki
Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title_full Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title_fullStr Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title_full_unstemmed Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title_short Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
title_sort evaluation of the reconstruction parameters of brain dopamine transporter spect images obtained by a fan beam collimator: a comparison with parallel-hole collimators
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6038970/
https://www.ncbi.nlm.nih.gov/pubmed/29998145
http://dx.doi.org/10.22038/aojnmb.2018.10330
work_keys_str_mv AT moritakeishin evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT maebatakeakira evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT iwasakirina evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT shiotsukiyuki evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT himurokazuhiko evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT babashingo evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators
AT sasakimasayuki evaluationofthereconstructionparametersofbraindopaminetransporterspectimagesobtainedbyafanbeamcollimatoracomparisonwithparallelholecollimators