Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example

INTRODUCTION: Establishing analytical performance requirements for in vitro diagnostic (IVD) assays is a challenging process. Manufacturers try to optimize analytical performance by choosing amongst many combinations of different product performance characteristics. Sigma metrics and method decision...

Descripción completa

Detalles Bibliográficos
Autores principales: Petrides, Victoria, Schneider, Sharon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6039165/
https://www.ncbi.nlm.nih.gov/pubmed/30022889
http://dx.doi.org/10.11613/BM.2018.020903
_version_ 1783338632044609536
author Petrides, Victoria
Schneider, Sharon
author_facet Petrides, Victoria
Schneider, Sharon
author_sort Petrides, Victoria
collection PubMed
description INTRODUCTION: Establishing analytical performance requirements for in vitro diagnostic (IVD) assays is a challenging process. Manufacturers try to optimize analytical performance by choosing amongst many combinations of different product performance characteristics. Sigma metrics and method decision charts can be helpful aids in choosing appropriate analytical performance requirements. The objective of this research was to demonstrate the use of Sigma metrics and method decision charts to help establish analytical performance requirements and to optimize analytical performance at medical decision concentrations for an IVD assay. MATERIALS AND METHODS: A range of possible Sigma metrics were determined using three sources for total allowable error (TEa) and hypothetical total PSA assay results. Method decision charts were created for each TEa source and used to identify the maximum precision and bias that the assay could have to maintain sigma level performance of at least 3. RESULTS: To achieve a sigma performance level of at least 3 for a hypothetical total PSA assay, the maximum allowable coefficient of variation ranged from 5.0% to 11.2% depending on the TEa source. To achieve a sigma performance level of at least 6, the maximum allowable coefficient of variation ranged from 2.5% to 5.6% depending on the TEa source. CONCLUSIONS: Using Sigma metrics and method decision charts when establishing analytical performance requirements can help manufacturers choose product requirements that will optimize IVD assay product performance.
format Online
Article
Text
id pubmed-6039165
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-60391652018-07-18 Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example Petrides, Victoria Schneider, Sharon Biochem Med (Zagreb) Short Communications INTRODUCTION: Establishing analytical performance requirements for in vitro diagnostic (IVD) assays is a challenging process. Manufacturers try to optimize analytical performance by choosing amongst many combinations of different product performance characteristics. Sigma metrics and method decision charts can be helpful aids in choosing appropriate analytical performance requirements. The objective of this research was to demonstrate the use of Sigma metrics and method decision charts to help establish analytical performance requirements and to optimize analytical performance at medical decision concentrations for an IVD assay. MATERIALS AND METHODS: A range of possible Sigma metrics were determined using three sources for total allowable error (TEa) and hypothetical total PSA assay results. Method decision charts were created for each TEa source and used to identify the maximum precision and bias that the assay could have to maintain sigma level performance of at least 3. RESULTS: To achieve a sigma performance level of at least 3 for a hypothetical total PSA assay, the maximum allowable coefficient of variation ranged from 5.0% to 11.2% depending on the TEa source. To achieve a sigma performance level of at least 6, the maximum allowable coefficient of variation ranged from 2.5% to 5.6% depending on the TEa source. CONCLUSIONS: Using Sigma metrics and method decision charts when establishing analytical performance requirements can help manufacturers choose product requirements that will optimize IVD assay product performance. Croatian Society of Medical Biochemistry and Laboratory Medicine 2018-06-15 2018-06-15 /pmc/articles/PMC6039165/ /pubmed/30022889 http://dx.doi.org/10.11613/BM.2018.020903 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Short Communications
Petrides, Victoria
Schneider, Sharon
Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title_full Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title_fullStr Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title_full_unstemmed Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title_short Using Sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total PSA assay as an example
title_sort using sigma metrics to establish analytical product performance requirements and optimize analytical performance of an in vitro diagnostic assay using a theoretical total psa assay as an example
topic Short Communications
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6039165/
https://www.ncbi.nlm.nih.gov/pubmed/30022889
http://dx.doi.org/10.11613/BM.2018.020903
work_keys_str_mv AT petridesvictoria usingsigmametricstoestablishanalyticalproductperformancerequirementsandoptimizeanalyticalperformanceofaninvitrodiagnosticassayusingatheoreticaltotalpsaassayasanexample
AT schneidersharon usingsigmametricstoestablishanalyticalproductperformancerequirementsandoptimizeanalyticalperformanceofaninvitrodiagnosticassayusingatheoreticaltotalpsaassayasanexample