Comparison of manual & automated analysis methods for corneal endothelial cell density measurements by specular microscopy
PURPOSE: To determine the reliability of corneal endothelial cell density (ECD) obtained by automated specular microscopy versus that of validated manual methods and factors that predict such reliability. METHODS: Sharp central images from 94 control and 106 glaucomatous eyes were captured with Kona...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6039579/ https://www.ncbi.nlm.nih.gov/pubmed/28797649 http://dx.doi.org/10.1016/j.optom.2017.06.001 |
Sumario: | PURPOSE: To determine the reliability of corneal endothelial cell density (ECD) obtained by automated specular microscopy versus that of validated manual methods and factors that predict such reliability. METHODS: Sharp central images from 94 control and 106 glaucomatous eyes were captured with Konan specular microscope NSP-9900. All images were analyzed by trained graders using Konan CellChek Software, employing the fully- and semi-automated methods as well as Center Method. Images with low cell count (input cells number <100) and/or guttata were compared with the Center and Flex-Center Methods. ECDs were compared and absolute error was used to assess variation. The effect on ECD of age, cell count, cell size, and cell size variation was evaluated. RESULTS: No significant difference was observed between the Center and Flex-Center Methods in corneas with guttata (p = 0.48) or low ECD (p = 0.11). No difference (p = 0.32) was observed in ECD of normal controls <40 yrs old between the fully-automated method and manual Center Method. However, in older controls and glaucomatous eyes, ECD was overestimated by the fully-automated method (p = 0.034) and semi-automated method (p = 0.025) as compared to manual method. CONCLUSION: Our findings show that automated analysis significantly overestimates ECD in the eyes with high polymegathism and/or large cell size, compared to the manual method. Therefore, we discourage reliance upon the fully-automated method alone to perform specular microscopy analysis, particularly if an accurate ECD value is imperative. |
---|