Collimator scatter factor: Monte Carlo and in-air measurements approaches
BACKGROUND: Linac output as a function of field sizes has a phantom and a head scatter component. This last term can be measured in-air with appropriate build-up ensuring a complete electron equilibrium and the absence of the contaminant electrons. Equilibrium conditions could be achieved using a bu...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6042423/ https://www.ncbi.nlm.nih.gov/pubmed/29996873 http://dx.doi.org/10.1186/s13014-018-1070-6 |
_version_ | 1783339150763622400 |
---|---|
author | Fogliata, A. Stravato, A. Reggiori, G. Tomatis, S. Würfel, J. Scorsetti, M. Cozzi, L. |
author_facet | Fogliata, A. Stravato, A. Reggiori, G. Tomatis, S. Würfel, J. Scorsetti, M. Cozzi, L. |
author_sort | Fogliata, A. |
collection | PubMed |
description | BACKGROUND: Linac output as a function of field sizes has a phantom and a head scatter component. This last term can be measured in-air with appropriate build-up ensuring a complete electron equilibrium and the absence of the contaminant electrons. Equilibrium conditions could be achieved using a build-up cap or a mini-phantom. Monte Carlo simulations in a virtual phantom mimicking a mini-phantom were analysed with the aim of better understanding the setup conditions for measuring the collimator scatter factor that is the head scatter component of the linac output factors. METHODS: Beams of 6 and 15 MV from a TrueBeam, with size from 4 × 4 to 40 × 40 cm(2) were simulated in cylindrical acrylic phantoms 20 cm long, of different diameters, from 0.5 to 4 cm, with the cylinder axis coincident with the beam central axis. The PRIMO package, based on PENELOPE Monte Carlo code, was used. The phase-space files for a Varian TrueBeam linac, provided by the linac vendor, were used for the linac head simulation. Depth dose curves were analysed, and collimator scatter factors estimated at different depth in the different phantom conditions. Additionally, in-air measurements using acyrilic and brass build-up caps, as well as acrylic mini-phantom were acquired for 6 and 18 MV beams from a Varian Clinac DHX. RESULTS: The depth dose curves along the cylinders were compared, showing, in each phantom, very similar curves for all analysed field sizes, proving the correctness in estimating the collimator scatter factor in the mini-phantom, provided to position the detector to a sufficient depth to exclude electron contamination. The results were confirmed by the measurements, where the acrylic build-up cap showed to be inadequate to properly estimate the collimator scatter factors, while the mini-phantom and the brass caps gave reasonable measurements. CONCLUSION: A better understanding of the beam characteristics inside a virtual mini-phantom through the analysis of depth dose curves, showed the critical points of using the acrylic build-up cap, and suggested the use of the mini-phantom for the collimator scatter factor measurements in the medium-large field size range. |
format | Online Article Text |
id | pubmed-6042423 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-60424232018-07-13 Collimator scatter factor: Monte Carlo and in-air measurements approaches Fogliata, A. Stravato, A. Reggiori, G. Tomatis, S. Würfel, J. Scorsetti, M. Cozzi, L. Radiat Oncol Research BACKGROUND: Linac output as a function of field sizes has a phantom and a head scatter component. This last term can be measured in-air with appropriate build-up ensuring a complete electron equilibrium and the absence of the contaminant electrons. Equilibrium conditions could be achieved using a build-up cap or a mini-phantom. Monte Carlo simulations in a virtual phantom mimicking a mini-phantom were analysed with the aim of better understanding the setup conditions for measuring the collimator scatter factor that is the head scatter component of the linac output factors. METHODS: Beams of 6 and 15 MV from a TrueBeam, with size from 4 × 4 to 40 × 40 cm(2) were simulated in cylindrical acrylic phantoms 20 cm long, of different diameters, from 0.5 to 4 cm, with the cylinder axis coincident with the beam central axis. The PRIMO package, based on PENELOPE Monte Carlo code, was used. The phase-space files for a Varian TrueBeam linac, provided by the linac vendor, were used for the linac head simulation. Depth dose curves were analysed, and collimator scatter factors estimated at different depth in the different phantom conditions. Additionally, in-air measurements using acyrilic and brass build-up caps, as well as acrylic mini-phantom were acquired for 6 and 18 MV beams from a Varian Clinac DHX. RESULTS: The depth dose curves along the cylinders were compared, showing, in each phantom, very similar curves for all analysed field sizes, proving the correctness in estimating the collimator scatter factor in the mini-phantom, provided to position the detector to a sufficient depth to exclude electron contamination. The results were confirmed by the measurements, where the acrylic build-up cap showed to be inadequate to properly estimate the collimator scatter factors, while the mini-phantom and the brass caps gave reasonable measurements. CONCLUSION: A better understanding of the beam characteristics inside a virtual mini-phantom through the analysis of depth dose curves, showed the critical points of using the acrylic build-up cap, and suggested the use of the mini-phantom for the collimator scatter factor measurements in the medium-large field size range. BioMed Central 2018-07-11 /pmc/articles/PMC6042423/ /pubmed/29996873 http://dx.doi.org/10.1186/s13014-018-1070-6 Text en © The Author(s). 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Fogliata, A. Stravato, A. Reggiori, G. Tomatis, S. Würfel, J. Scorsetti, M. Cozzi, L. Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title | Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title_full | Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title_fullStr | Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title_full_unstemmed | Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title_short | Collimator scatter factor: Monte Carlo and in-air measurements approaches |
title_sort | collimator scatter factor: monte carlo and in-air measurements approaches |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6042423/ https://www.ncbi.nlm.nih.gov/pubmed/29996873 http://dx.doi.org/10.1186/s13014-018-1070-6 |
work_keys_str_mv | AT fogliataa collimatorscatterfactormontecarloandinairmeasurementsapproaches AT stravatoa collimatorscatterfactormontecarloandinairmeasurementsapproaches AT reggiorig collimatorscatterfactormontecarloandinairmeasurementsapproaches AT tomatiss collimatorscatterfactormontecarloandinairmeasurementsapproaches AT wurfelj collimatorscatterfactormontecarloandinairmeasurementsapproaches AT scorsettim collimatorscatterfactormontecarloandinairmeasurementsapproaches AT cozzil collimatorscatterfactormontecarloandinairmeasurementsapproaches |