Clinical Utility and Cross-Reactivity of Insulin and C-Peptide Assays by the Lumipulse G1200 System
BACKGROUND: Measurement of insulin and C-peptide concentrations is important for deciding whether insulin treatment is required in diabetic patients. We aimed to investigate the analytical performance of insulin and C-peptide assays using the Lumipulse G1200 system (Fujirebio Inc., Tokyo, Japan). ME...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society for Laboratory Medicine
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6056384/ https://www.ncbi.nlm.nih.gov/pubmed/30027696 http://dx.doi.org/10.3343/alm.2018.38.6.530 |
Sumario: | BACKGROUND: Measurement of insulin and C-peptide concentrations is important for deciding whether insulin treatment is required in diabetic patients. We aimed to investigate the analytical performance of insulin and C-peptide assays using the Lumipulse G1200 system (Fujirebio Inc., Tokyo, Japan). METHODS: We examined the precision, linearity, and cross-reactivity of insulin and C-peptide using five insulin analogues and purified proinsulin. A method comparison was conducted between the Lumipulse G1200 and Roche E170 (Roche Diagnostics, Mannheim, Germany) systems in 200 diabetic patients on insulin treatment. Reference intervals for insulin and C-peptide concentrations were determined in 279 healthy individuals. RESULTS: For insulin and C-peptide assays, within-laboratory precision (% CV) was 3.78–4.14 and 2.89–3.35%, respectively. The linearity of the insulin assay in the range of 0–2,778 pmol/L was R(2)=0.9997, and that of the C-peptide assay in the range of 0–10 nmol/L was R(2)=0.9996. The correlation coefficient (r) between the Roche E170 and Lumipulse G1200 results was 0.943 (P<0.001) for insulin and 0.996 (P<0.001) for C-peptide. The mean differences in insulin and C-peptide between Lumipulse G1200 and the Roche E170 were 19.4 pmol/L and 0.2 nmol/L, respectively. None of the insulin analogues or proinsulin showed significant cross-reactivity with the Lumipulse G1200. Reference intervals of insulin and C-peptide were 7.64–70.14 pmol/L and 0.17–0.85 nmol/L, respectively. CONCLUSIONS: Insulin and C-peptide tests on the Lumipulse G1200 show adequate analytical performance and are expected to be acceptable for use in clinical areas. |
---|