Cargando…
Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy
Monitoring nitric oxide at the trace level is required in a large range of applications. We report on a trace gas analyzer optimized for nitric oxide measurements by Optical Feedback Cavity Enhanced Absorption Spectroscopy with an interband cascade laser at 5.3 µm. The short response time of the ins...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6068816/ https://www.ncbi.nlm.nih.gov/pubmed/29932103 http://dx.doi.org/10.3390/s18071997 |
_version_ | 1783343353016877056 |
---|---|
author | Richard, Lucile Romanini, Daniele Ventrillard, Irène |
author_facet | Richard, Lucile Romanini, Daniele Ventrillard, Irène |
author_sort | Richard, Lucile |
collection | PubMed |
description | Monitoring nitric oxide at the trace level is required in a large range of applications. We report on a trace gas analyzer optimized for nitric oxide measurements by Optical Feedback Cavity Enhanced Absorption Spectroscopy with an interband cascade laser at 5.3 µm. The short response time of the instrument allows for reaching the level of 50 ppt in only 180 ms. Its stability enables averaging up to 12 min to reach a detection limit of 0.9 ppt. Absolute concentration calibration requires to account for the optical saturation effect that results from the intense absorption line intensity addressed here, in the mid infrared region, in contrast to instruments that are operating in the near infrared region. |
format | Online Article Text |
id | pubmed-6068816 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-60688162018-08-07 Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy Richard, Lucile Romanini, Daniele Ventrillard, Irène Sensors (Basel) Article Monitoring nitric oxide at the trace level is required in a large range of applications. We report on a trace gas analyzer optimized for nitric oxide measurements by Optical Feedback Cavity Enhanced Absorption Spectroscopy with an interband cascade laser at 5.3 µm. The short response time of the instrument allows for reaching the level of 50 ppt in only 180 ms. Its stability enables averaging up to 12 min to reach a detection limit of 0.9 ppt. Absolute concentration calibration requires to account for the optical saturation effect that results from the intense absorption line intensity addressed here, in the mid infrared region, in contrast to instruments that are operating in the near infrared region. MDPI 2018-06-22 /pmc/articles/PMC6068816/ /pubmed/29932103 http://dx.doi.org/10.3390/s18071997 Text en © 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Richard, Lucile Romanini, Daniele Ventrillard, Irène Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title | Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title_full | Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title_fullStr | Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title_full_unstemmed | Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title_short | Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy |
title_sort | nitric oxide analysis down to ppt levels by optical-feedback cavity-enhanced absorption spectroscopy |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6068816/ https://www.ncbi.nlm.nih.gov/pubmed/29932103 http://dx.doi.org/10.3390/s18071997 |
work_keys_str_mv | AT richardlucile nitricoxideanalysisdowntopptlevelsbyopticalfeedbackcavityenhancedabsorptionspectroscopy AT romaninidaniele nitricoxideanalysisdowntopptlevelsbyopticalfeedbackcavityenhancedabsorptionspectroscopy AT ventrillardirene nitricoxideanalysisdowntopptlevelsbyopticalfeedbackcavityenhancedabsorptionspectroscopy |