Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva
Rationale: Saliva as a sample matrix is rapidly gaining interest for disease diagnosis and point-of-care assays because it is easy to collect (non-invasive) and contains many health-related biomarkers. However, saliva poses particular problems relative to more common urine and blood matrices, which...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Ivyspring International Publisher
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6071535/ https://www.ncbi.nlm.nih.gov/pubmed/30083260 http://dx.doi.org/10.7150/thno.24941 |
_version_ | 1783343887930097664 |
---|---|
author | Noiphung, Julaluk Nguyen, Michael P. Punyadeera, Chamindie Wan, Yunxia Laiwattanapaisal, Wanida Henry, Charles S. |
author_facet | Noiphung, Julaluk Nguyen, Michael P. Punyadeera, Chamindie Wan, Yunxia Laiwattanapaisal, Wanida Henry, Charles S. |
author_sort | Noiphung, Julaluk |
collection | PubMed |
description | Rationale: Saliva as a sample matrix is rapidly gaining interest for disease diagnosis and point-of-care assays because it is easy to collect (non-invasive) and contains many health-related biomarkers. However, saliva poses particular problems relative to more common urine and blood matrices, which includes low analyte concentrations, lack of understanding of biomolecule transportation and inherent viscosity variability in human samples. While several studies have sought to improve assay sensitivity, few have addressed sample viscosity specifically. The goal of this study is to minimize the effect of sample viscosity on paper-based analytical devices (PADs) for the measurement of pH and nitrite in human saliva. Methods: PADs were used to measure salivary pH from 5.0 to 10.0 with a universal indicator consisting of chlorophenol red, phenol red and phenolphthalein. Nitrite determination was performed using the Griess reaction. Artificial saliva with viscosity values between 1.54 and 5.10 mPa∙s was tested on the proposed PAD. To ensure the proposed PADs can be tailored for use in-field analysis, the devices were shipped to Australia and tested with human specimens. Results: Initial experiments showed that viscosity had a significant impact on the calibration curve for nitrite; however, a more consistent curve could be generated when buffer was added after the sample, irrespective of sample viscosity. The linear range for nitrite detection was 0.1 to 2.4 mg/dL using the improved method. The nitrite measurement in artificial saliva also showed a good correlation with the standard spectrophotometry method (p=0.8484, paired sample t-test, n=20). Measured pH values from samples with varying viscosities correlated well with the results from our pH meter. Conclusions: The inherent variation of salivary viscosity that impacts nitrite and pH results can be addressed using a simple washing step on the PAD without the need for complex procedures. |
format | Online Article Text |
id | pubmed-6071535 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Ivyspring International Publisher |
record_format | MEDLINE/PubMed |
spelling | pubmed-60715352018-08-06 Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva Noiphung, Julaluk Nguyen, Michael P. Punyadeera, Chamindie Wan, Yunxia Laiwattanapaisal, Wanida Henry, Charles S. Theranostics Research Paper Rationale: Saliva as a sample matrix is rapidly gaining interest for disease diagnosis and point-of-care assays because it is easy to collect (non-invasive) and contains many health-related biomarkers. However, saliva poses particular problems relative to more common urine and blood matrices, which includes low analyte concentrations, lack of understanding of biomolecule transportation and inherent viscosity variability in human samples. While several studies have sought to improve assay sensitivity, few have addressed sample viscosity specifically. The goal of this study is to minimize the effect of sample viscosity on paper-based analytical devices (PADs) for the measurement of pH and nitrite in human saliva. Methods: PADs were used to measure salivary pH from 5.0 to 10.0 with a universal indicator consisting of chlorophenol red, phenol red and phenolphthalein. Nitrite determination was performed using the Griess reaction. Artificial saliva with viscosity values between 1.54 and 5.10 mPa∙s was tested on the proposed PAD. To ensure the proposed PADs can be tailored for use in-field analysis, the devices were shipped to Australia and tested with human specimens. Results: Initial experiments showed that viscosity had a significant impact on the calibration curve for nitrite; however, a more consistent curve could be generated when buffer was added after the sample, irrespective of sample viscosity. The linear range for nitrite detection was 0.1 to 2.4 mg/dL using the improved method. The nitrite measurement in artificial saliva also showed a good correlation with the standard spectrophotometry method (p=0.8484, paired sample t-test, n=20). Measured pH values from samples with varying viscosities correlated well with the results from our pH meter. Conclusions: The inherent variation of salivary viscosity that impacts nitrite and pH results can be addressed using a simple washing step on the PAD without the need for complex procedures. Ivyspring International Publisher 2018-06-13 /pmc/articles/PMC6071535/ /pubmed/30083260 http://dx.doi.org/10.7150/thno.24941 Text en © Ivyspring International Publisher This is an open access article distributed under the terms of the Creative Commons Attribution (CC BY-NC) license (https://creativecommons.org/licenses/by-nc/4.0/). See http://ivyspring.com/terms for full terms and conditions. |
spellingShingle | Research Paper Noiphung, Julaluk Nguyen, Michael P. Punyadeera, Chamindie Wan, Yunxia Laiwattanapaisal, Wanida Henry, Charles S. Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title | Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title_full | Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title_fullStr | Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title_full_unstemmed | Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title_short | Development of Paper-Based Analytical Devices for Minimizing the Viscosity Effect in Human Saliva |
title_sort | development of paper-based analytical devices for minimizing the viscosity effect in human saliva |
topic | Research Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6071535/ https://www.ncbi.nlm.nih.gov/pubmed/30083260 http://dx.doi.org/10.7150/thno.24941 |
work_keys_str_mv | AT noiphungjulaluk developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva AT nguyenmichaelp developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva AT punyadeerachamindie developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva AT wanyunxia developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva AT laiwattanapaisalwanida developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva AT henrycharless developmentofpaperbasedanalyticaldevicesforminimizingtheviscosityeffectinhumansaliva |