Cargando…
Malfunction of a central venous multilumen access catheter caused by kinking: A case report
RATIONALE: A multilumen access catheter (MAC) is a large-bore catheter that prevents kinking or collapsing. Misplacement is a frequently reported complication. PATIENT CONCERNS: A 59-year-old man who was diagnosed with a hemoperitoneum due to a liver laceration after a fall. DIAGNOSIS: After cathete...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer Health
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6078719/ https://www.ncbi.nlm.nih.gov/pubmed/30045303 http://dx.doi.org/10.1097/MD.0000000000011622 |
Sumario: | RATIONALE: A multilumen access catheter (MAC) is a large-bore catheter that prevents kinking or collapsing. Misplacement is a frequently reported complication. PATIENT CONCERNS: A 59-year-old man who was diagnosed with a hemoperitoneum due to a liver laceration after a fall. DIAGNOSIS: After catheterization, we were able to aspirate blood through the 9 Fr, but not the 12-gauge line. Thus, we assumed that the catheter was misplaced. Nevertheless, ultrasonographic findings did not reveal the reason for the malfunction. We found kinking in the distal part of the catheter after removing it. INTERVENTION: The MAC was removed soon after stopping the active bleeding. OUTCOMES: A hemihepatectomy was successfully performed, and the patient was transferred to the intensive care unit. LESSONS: Anesthesiologists should consider kinking of large-bore catheters, including MACs. |
---|