Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker

A 67-year old male with a dual-chamber pacemaker visited for a regular check-up. An unfamiliar message emerged on the display just after placing the programmer wand. We could recognize that the pacemaker had already been in the safe back-up mode of DDI, and the programmer prompted a re-initializatio...

Descripción completa

Detalles Bibliográficos
Autores principales: Yazaki, Kyoichiro, Watarai, Masahiro, Kahata, Mitsuru, Kumagai, Asako, Inoue, Koji, Koganei, Hiroshi, Enta, Kenji, Otsuka, Masato, Ishii, Yasuhiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6090071/
https://www.ncbi.nlm.nih.gov/pubmed/29477310
http://dx.doi.org/10.1016/j.ipej.2018.02.005
_version_ 1783347135810371584
author Yazaki, Kyoichiro
Watarai, Masahiro
Kahata, Mitsuru
Kumagai, Asako
Inoue, Koji
Koganei, Hiroshi
Enta, Kenji
Otsuka, Masato
Ishii, Yasuhiro
author_facet Yazaki, Kyoichiro
Watarai, Masahiro
Kahata, Mitsuru
Kumagai, Asako
Inoue, Koji
Koganei, Hiroshi
Enta, Kenji
Otsuka, Masato
Ishii, Yasuhiro
author_sort Yazaki, Kyoichiro
collection PubMed
description A 67-year old male with a dual-chamber pacemaker visited for a regular check-up. An unfamiliar message emerged on the display just after placing the programmer wand. We could recognize that the pacemaker had already been in the safe back-up mode of DDI, and the programmer prompted a re-initialization request. We are so surprised because there was no indication of device malfunction the day before in daily monitoring and a 12-lead electrocardiogram revealed normally working in the DDD mode just before checking the device. The pacemaker was immediately re-programmed to the former setting. This phenomenon has not recurred for 12 months.
format Online
Article
Text
id pubmed-6090071
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-60900712018-08-16 Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker Yazaki, Kyoichiro Watarai, Masahiro Kahata, Mitsuru Kumagai, Asako Inoue, Koji Koganei, Hiroshi Enta, Kenji Otsuka, Masato Ishii, Yasuhiro Indian Pacing Electrophysiol J Case Report A 67-year old male with a dual-chamber pacemaker visited for a regular check-up. An unfamiliar message emerged on the display just after placing the programmer wand. We could recognize that the pacemaker had already been in the safe back-up mode of DDI, and the programmer prompted a re-initialization request. We are so surprised because there was no indication of device malfunction the day before in daily monitoring and a 12-lead electrocardiogram revealed normally working in the DDD mode just before checking the device. The pacemaker was immediately re-programmed to the former setting. This phenomenon has not recurred for 12 months. Elsevier 2018-02-22 /pmc/articles/PMC6090071/ /pubmed/29477310 http://dx.doi.org/10.1016/j.ipej.2018.02.005 Text en © 2018 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Yazaki, Kyoichiro
Watarai, Masahiro
Kahata, Mitsuru
Kumagai, Asako
Inoue, Koji
Koganei, Hiroshi
Enta, Kenji
Otsuka, Masato
Ishii, Yasuhiro
Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title_full Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title_fullStr Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title_full_unstemmed Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title_short Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
title_sort cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6090071/
https://www.ncbi.nlm.nih.gov/pubmed/29477310
http://dx.doi.org/10.1016/j.ipej.2018.02.005
work_keys_str_mv AT yazakikyoichiro causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT wataraimasahiro causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT kahatamitsuru causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT kumagaiasako causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT inouekoji causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT koganeihiroshi causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT entakenji causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT otsukamasato causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker
AT ishiiyasuhiro causeofthepoweronresetphenomenonotherthanelectricmagneticinterferenceinapatientwithapacemaker