Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemaker
A 67-year old male with a dual-chamber pacemaker visited for a regular check-up. An unfamiliar message emerged on the display just after placing the programmer wand. We could recognize that the pacemaker had already been in the safe back-up mode of DDI, and the programmer prompted a re-initializatio...