Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes

PURPOSE: To evaluate the intrasession repeatability of refractive and ocular aberrometric measurements obtained using a new multidiagnostic device in healthy eyes. PATIENTS AND METHODS: A total of 107 eyes of 107 patients, age ranging from 23 to 65 years, were enrolled in this study. A complete eye...

Descripción completa

Detalles Bibliográficos
Autores principales: Piñero, David P, López-Navarro, Alberto, Cabezos, Inmaculada, de Fez, Dolores, Caballero, María T, Camps, Vicent J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6095565/
https://www.ncbi.nlm.nih.gov/pubmed/30214365
http://dx.doi.org/10.2147/OPTO.S132572
_version_ 1783347960539512832
author Piñero, David P
López-Navarro, Alberto
Cabezos, Inmaculada
de Fez, Dolores
Caballero, María T
Camps, Vicent J
author_facet Piñero, David P
López-Navarro, Alberto
Cabezos, Inmaculada
de Fez, Dolores
Caballero, María T
Camps, Vicent J
author_sort Piñero, David P
collection PubMed
description PURPOSE: To evaluate the intrasession repeatability of refractive and ocular aberrometric measurements obtained using a new multidiagnostic device in healthy eyes. PATIENTS AND METHODS: A total of 107 eyes of 107 patients, age ranging from 23 to 65 years, were enrolled in this study. A complete eye examination was performed in all eyes, including an ocular examination using the VX120 system. Three consecutive measurements were obtained using this device to assess the intrasession repeatability of different refractive and ocular aber-rometric parameters. The within-subject standard deviation (S(w)), intrasubject precision (1.96×S(w)), and intraclass correlation coefficient (ICC) were calculated. RESULTS: S(w) and intrasubject precision for refractive data were below 0.12 and 0.20 D, respectively, in all cases. The ICC ranged from 0.947 for the J(45) power vector component to 0.997 for the sphere. Concerning aberrometric measurements S(w) and intrasubject precision values were below 0.05 µm and 0.10 µm, respectively. Likewise, the ICC ranged from 0.805 for the quadrafoil root mean square to 0.954 for the primary spherical aberration. Poor correlations were found between most of the refractive parameters and their S(w) (–0.033≤r≤0.053, p≥0.064). Moderate and significant positive correlations were found between the magnitude of the aberrometric parameters evaluated and their S(w) (r≥0.446, p<0.001). CONCLUSION: The new multidiagnostic device evaluated is able to provide consistent measurements of refraction and ocular aberrations in healthy eyes. Future studies should confirm if this consistency is also observed in highly aberrated eyes.
format Online
Article
Text
id pubmed-6095565
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-60955652018-09-13 Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes Piñero, David P López-Navarro, Alberto Cabezos, Inmaculada de Fez, Dolores Caballero, María T Camps, Vicent J Clin Optom (Auckl) Case Series PURPOSE: To evaluate the intrasession repeatability of refractive and ocular aberrometric measurements obtained using a new multidiagnostic device in healthy eyes. PATIENTS AND METHODS: A total of 107 eyes of 107 patients, age ranging from 23 to 65 years, were enrolled in this study. A complete eye examination was performed in all eyes, including an ocular examination using the VX120 system. Three consecutive measurements were obtained using this device to assess the intrasession repeatability of different refractive and ocular aber-rometric parameters. The within-subject standard deviation (S(w)), intrasubject precision (1.96×S(w)), and intraclass correlation coefficient (ICC) were calculated. RESULTS: S(w) and intrasubject precision for refractive data were below 0.12 and 0.20 D, respectively, in all cases. The ICC ranged from 0.947 for the J(45) power vector component to 0.997 for the sphere. Concerning aberrometric measurements S(w) and intrasubject precision values were below 0.05 µm and 0.10 µm, respectively. Likewise, the ICC ranged from 0.805 for the quadrafoil root mean square to 0.954 for the primary spherical aberration. Poor correlations were found between most of the refractive parameters and their S(w) (–0.033≤r≤0.053, p≥0.064). Moderate and significant positive correlations were found between the magnitude of the aberrometric parameters evaluated and their S(w) (r≥0.446, p<0.001). CONCLUSION: The new multidiagnostic device evaluated is able to provide consistent measurements of refraction and ocular aberrations in healthy eyes. Future studies should confirm if this consistency is also observed in highly aberrated eyes. Dove Medical Press 2017-05-04 /pmc/articles/PMC6095565/ /pubmed/30214365 http://dx.doi.org/10.2147/OPTO.S132572 Text en © 2017 Piñero et al. This work is published and licensed by Dove Medical Press Limited The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Case Series
Piñero, David P
López-Navarro, Alberto
Cabezos, Inmaculada
de Fez, Dolores
Caballero, María T
Camps, Vicent J
Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title_full Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title_fullStr Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title_full_unstemmed Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title_short Intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
title_sort intrasession repeatability of refractive and ocular aberrometric measurements obtained using a multidiagnostic device in healthy eyes
topic Case Series
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6095565/
https://www.ncbi.nlm.nih.gov/pubmed/30214365
http://dx.doi.org/10.2147/OPTO.S132572
work_keys_str_mv AT pinerodavidp intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes
AT lopeznavarroalberto intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes
AT cabezosinmaculada intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes
AT defezdolores intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes
AT caballeromariat intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes
AT campsvicentj intrasessionrepeatabilityofrefractiveandocularaberrometricmeasurementsobtainedusingamultidiagnosticdeviceinhealthyeyes