Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma

PURPOSE: The aim of this study was to determine whether fluctuations in intraocular pressure (IOP) occur as a result of the order of IOP measurements or successive IOP measurements in patients with glaucoma and, if so, identify the factors causing these fluctuations. PATIENTS AND METHODS: Four hundr...

Descripción completa

Detalles Bibliográficos
Autores principales: Yaoeda, Kiyoshi, Fukushima, Atsushi, Shirakashi, Motohiro, Miki, Atsushi, Fukuchi, Takeo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6103609/
https://www.ncbi.nlm.nih.gov/pubmed/30154644
http://dx.doi.org/10.2147/OPTH.S174277
_version_ 1783349371890302976
author Yaoeda, Kiyoshi
Fukushima, Atsushi
Shirakashi, Motohiro
Miki, Atsushi
Fukuchi, Takeo
author_facet Yaoeda, Kiyoshi
Fukushima, Atsushi
Shirakashi, Motohiro
Miki, Atsushi
Fukuchi, Takeo
author_sort Yaoeda, Kiyoshi
collection PubMed
description PURPOSE: The aim of this study was to determine whether fluctuations in intraocular pressure (IOP) occur as a result of the order of IOP measurements or successive IOP measurements in patients with glaucoma and, if so, identify the factors causing these fluctuations. PATIENTS AND METHODS: Four hundred twenty-eight eyes of 214 Japanese patients with primary open-angle glaucoma (POAG) were enrolled. Patients treated with beta-blockers or prostaglandin analogs alone were included. Additionally, in the IOP measurements by noncontact tonometer, the same cases of IOP of the right and left eyes prior to this study were included in this study. Four successive IOP measurements were carried out using a Goldmann applanation tonometer as follows: IOP was measured in the first eye (right or left) and then in the fellow eye and IOP was again measured in the first eye and then in the fellow eye. Repeated-measures analysis of variance was used to test the differences in IOP between successive measurements. Generalized linear mixed models were used to test differences in IOP measurements between the right and the left eyes on repeated applanation tonometry and according to the order of measurement. Conditional binomial logistic regression analysis was used to identify factors associated with fluctuating repeated applanation tonometry measurements. A P-value of <0.05 was considered statistically significant. RESULTS: IOP values decreased significantly according to the number of measurements (13.8–13.0; P<0.001–0.036, respectively). There was no significant difference in IOP measurements between the right and left eyes. The first IOP measurement was significantly higher than the fourth measurement (P=0.038); however, there was no significant difference between other combinations. The use of a prostaglandin analog was the only significant contributor to fluctuating IOP measurements (P=0.002). CONCLUSION: IOP measured in the first eye, either right or left, was higher than that measured in the fellow eye in Japanese patients with POAG. The use of a prostaglandin analog may be associated with fluctuating IOP on repeated applanation tonometry.
format Online
Article
Text
id pubmed-6103609
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-61036092018-08-28 Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma Yaoeda, Kiyoshi Fukushima, Atsushi Shirakashi, Motohiro Miki, Atsushi Fukuchi, Takeo Clin Ophthalmol Original Research PURPOSE: The aim of this study was to determine whether fluctuations in intraocular pressure (IOP) occur as a result of the order of IOP measurements or successive IOP measurements in patients with glaucoma and, if so, identify the factors causing these fluctuations. PATIENTS AND METHODS: Four hundred twenty-eight eyes of 214 Japanese patients with primary open-angle glaucoma (POAG) were enrolled. Patients treated with beta-blockers or prostaglandin analogs alone were included. Additionally, in the IOP measurements by noncontact tonometer, the same cases of IOP of the right and left eyes prior to this study were included in this study. Four successive IOP measurements were carried out using a Goldmann applanation tonometer as follows: IOP was measured in the first eye (right or left) and then in the fellow eye and IOP was again measured in the first eye and then in the fellow eye. Repeated-measures analysis of variance was used to test the differences in IOP between successive measurements. Generalized linear mixed models were used to test differences in IOP measurements between the right and the left eyes on repeated applanation tonometry and according to the order of measurement. Conditional binomial logistic regression analysis was used to identify factors associated with fluctuating repeated applanation tonometry measurements. A P-value of <0.05 was considered statistically significant. RESULTS: IOP values decreased significantly according to the number of measurements (13.8–13.0; P<0.001–0.036, respectively). There was no significant difference in IOP measurements between the right and left eyes. The first IOP measurement was significantly higher than the fourth measurement (P=0.038); however, there was no significant difference between other combinations. The use of a prostaglandin analog was the only significant contributor to fluctuating IOP measurements (P=0.002). CONCLUSION: IOP measured in the first eye, either right or left, was higher than that measured in the fellow eye in Japanese patients with POAG. The use of a prostaglandin analog may be associated with fluctuating IOP on repeated applanation tonometry. Dove Medical Press 2018-08-17 /pmc/articles/PMC6103609/ /pubmed/30154644 http://dx.doi.org/10.2147/OPTH.S174277 Text en © 2018 Yaoeda et al. This work is published and licensed by Dove Medical Press Limited The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Original Research
Yaoeda, Kiyoshi
Fukushima, Atsushi
Shirakashi, Motohiro
Miki, Atsushi
Fukuchi, Takeo
Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title_full Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title_fullStr Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title_full_unstemmed Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title_short Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma
title_sort factors associated with fluctuations in repeated measurements of intraocular pressure using the goldmann applanation tonometer in japanese patients with primary open-angle glaucoma
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6103609/
https://www.ncbi.nlm.nih.gov/pubmed/30154644
http://dx.doi.org/10.2147/OPTH.S174277
work_keys_str_mv AT yaoedakiyoshi factorsassociatedwithfluctuationsinrepeatedmeasurementsofintraocularpressureusingthegoldmannapplanationtonometerinjapanesepatientswithprimaryopenangleglaucoma
AT fukushimaatsushi factorsassociatedwithfluctuationsinrepeatedmeasurementsofintraocularpressureusingthegoldmannapplanationtonometerinjapanesepatientswithprimaryopenangleglaucoma
AT shirakashimotohiro factorsassociatedwithfluctuationsinrepeatedmeasurementsofintraocularpressureusingthegoldmannapplanationtonometerinjapanesepatientswithprimaryopenangleglaucoma
AT mikiatsushi factorsassociatedwithfluctuationsinrepeatedmeasurementsofintraocularpressureusingthegoldmannapplanationtonometerinjapanesepatientswithprimaryopenangleglaucoma
AT fukuchitakeo factorsassociatedwithfluctuationsinrepeatedmeasurementsofintraocularpressureusingthegoldmannapplanationtonometerinjapanesepatientswithprimaryopenangleglaucoma