Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?

BACKGROUND: In pacemaker‐dependent patients, the risk of asystole must be managed during device replacement. This study aimed to examine whether we could predict the indication for temporary pacing (TP) during the generator replacement. METHODS: We studied 105 consecutive patients who underwent pace...

Descripción completa

Detalles Bibliográficos
Autores principales: Kimura, Yuki, Sumiyoshi, Masataka, Inoue, Kenji, Shiozaki, Masayuki, Fukuda, Kentaro, Fujiwara, Yasumasa, Tabuchi, Haruna, Hayashi, Hidemori, Sekita, Gaku, Tokano, Takashi, Nakazato, Yuji, Daida, Hiroyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6111470/
https://www.ncbi.nlm.nih.gov/pubmed/30167017
http://dx.doi.org/10.1002/joa3.12072
_version_ 1783350663563968512
author Kimura, Yuki
Sumiyoshi, Masataka
Inoue, Kenji
Shiozaki, Masayuki
Fukuda, Kentaro
Fujiwara, Yasumasa
Tabuchi, Haruna
Hayashi, Hidemori
Sekita, Gaku
Tokano, Takashi
Nakazato, Yuji
Daida, Hiroyuki
author_facet Kimura, Yuki
Sumiyoshi, Masataka
Inoue, Kenji
Shiozaki, Masayuki
Fukuda, Kentaro
Fujiwara, Yasumasa
Tabuchi, Haruna
Hayashi, Hidemori
Sekita, Gaku
Tokano, Takashi
Nakazato, Yuji
Daida, Hiroyuki
author_sort Kimura, Yuki
collection PubMed
description BACKGROUND: In pacemaker‐dependent patients, the risk of asystole must be managed during device replacement. This study aimed to examine whether we could predict the indication for temporary pacing (TP) during the generator replacement. METHODS: We studied 105 consecutive patients who underwent pacemaker replacement due to battery depletion at Juntendo Nerima Hospital between September 2005 and December 2016. We examined the relationship between the stability of the intrinsic rhythm (IR) during pacemaker replacement and several clinical factors including age, gender, sick sinus syndrome or atrioventricular (AV) block, duration of pacing, structural heart disease, use of anti‐arrhythmic drugs, and the presence/absence of a stable IR (>40 bpm) at the outpatient clinic (OPC) just before pacemaker replacement. RESULTS: Of the 105 patients, we excluded 1 patient who required TP because of bradycardia‐dependent torsades de points. Therefore, we evaluated 104 patients for the indication for TP. TP was underwent in 19 patients (18%) because of an absence or instability of the IR during pacemaker replacement. The indication for TP was significantly correlated with AV block (84% vs 48%, P = .0044) and the absence of a stable IR at the last OPC visit (89% vs 24%, P < .0001). For predicting the indication for TP, the following values of no stable IR at the last OPC visit were obtained: 89% sensitivity, 77% specificity, 46% positive predictive value, and 97% negative predictive value. CONCLUSIONS: The presence of a stable IR at the last OPC visit was a good predictor (97%) of no indication for TP during pacemaker replacement.
format Online
Article
Text
id pubmed-6111470
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-61114702018-08-30 Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing? Kimura, Yuki Sumiyoshi, Masataka Inoue, Kenji Shiozaki, Masayuki Fukuda, Kentaro Fujiwara, Yasumasa Tabuchi, Haruna Hayashi, Hidemori Sekita, Gaku Tokano, Takashi Nakazato, Yuji Daida, Hiroyuki J Arrhythm Original Articles BACKGROUND: In pacemaker‐dependent patients, the risk of asystole must be managed during device replacement. This study aimed to examine whether we could predict the indication for temporary pacing (TP) during the generator replacement. METHODS: We studied 105 consecutive patients who underwent pacemaker replacement due to battery depletion at Juntendo Nerima Hospital between September 2005 and December 2016. We examined the relationship between the stability of the intrinsic rhythm (IR) during pacemaker replacement and several clinical factors including age, gender, sick sinus syndrome or atrioventricular (AV) block, duration of pacing, structural heart disease, use of anti‐arrhythmic drugs, and the presence/absence of a stable IR (>40 bpm) at the outpatient clinic (OPC) just before pacemaker replacement. RESULTS: Of the 105 patients, we excluded 1 patient who required TP because of bradycardia‐dependent torsades de points. Therefore, we evaluated 104 patients for the indication for TP. TP was underwent in 19 patients (18%) because of an absence or instability of the IR during pacemaker replacement. The indication for TP was significantly correlated with AV block (84% vs 48%, P = .0044) and the absence of a stable IR at the last OPC visit (89% vs 24%, P < .0001). For predicting the indication for TP, the following values of no stable IR at the last OPC visit were obtained: 89% sensitivity, 77% specificity, 46% positive predictive value, and 97% negative predictive value. CONCLUSIONS: The presence of a stable IR at the last OPC visit was a good predictor (97%) of no indication for TP during pacemaker replacement. John Wiley and Sons Inc. 2018-05-28 /pmc/articles/PMC6111470/ /pubmed/30167017 http://dx.doi.org/10.1002/joa3.12072 Text en © 2018 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Original Articles
Kimura, Yuki
Sumiyoshi, Masataka
Inoue, Kenji
Shiozaki, Masayuki
Fukuda, Kentaro
Fujiwara, Yasumasa
Tabuchi, Haruna
Hayashi, Hidemori
Sekita, Gaku
Tokano, Takashi
Nakazato, Yuji
Daida, Hiroyuki
Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title_full Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title_fullStr Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title_full_unstemmed Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title_short Stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?
title_sort stability of intrinsic rhythm in pacemaker‐dependent patients during pacemaker replacement: can we predict the need for temporary pacing?
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6111470/
https://www.ncbi.nlm.nih.gov/pubmed/30167017
http://dx.doi.org/10.1002/joa3.12072
work_keys_str_mv AT kimurayuki stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT sumiyoshimasataka stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT inouekenji stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT shiozakimasayuki stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT fukudakentaro stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT fujiwarayasumasa stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT tabuchiharuna stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT hayashihidemori stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT sekitagaku stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT tokanotakashi stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT nakazatoyuji stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing
AT daidahiroyuki stabilityofintrinsicrhythminpacemakerdependentpatientsduringpacemakerreplacementcanwepredicttheneedfortemporarypacing