Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study

BACKGROUND: Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current C...

Descripción completa

Detalles Bibliográficos
Autores principales: Mussmann, Bo, Andersen, Poul Erik, Torfing, Trine, Overgaard, Søren
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6117866/
https://www.ncbi.nlm.nih.gov/pubmed/30186626
http://dx.doi.org/10.1177/2058460118796539
Descripción
Sumario:BACKGROUND: Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current CT workstations PURPOSE: To test the between-scan agreement and reliability of custom segmentation software for BMD measurements adjacent to cemented and uncemented acetabular cups in dual-energy CT (DECT). MATERIAL AND METHODS: Twenty-four male patients with total hip arthroplasty were scanned and rescanned using 130-keV virtual monochromatic DECT images. Hemispherical regions of interest were defined slice-by-slice and BMD was calculated around the acetabular cup using custom segmentation software. RESULTS: In the uncemented cup, the mean BMD was 153 mg/cm(3) with a between-scan difference of 10 mg/cm(3) (P < 0.0001). In the cemented cup, the mean BMD was 186 mg/cm(3) with a between-scan difference of 6 mg/cm(3) (P = 0.15). In both uncemented and cemented cups the intraclass correlation coefficient between repeated measurements was >0.95 and narrow Bland–Altman Limits of Agreement. CONCLUSION: BMD can be measured with high absolute between-scan agreement and good reliability adjacent to acetabular cemented and uncemented cups using DECT and segmentation software.