Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines

Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clin...

Descripción completa

Detalles Bibliográficos
Autores principales: Du, Weiliang, Gao, Song, Jiang, Wei, Kudchadker, Rajat J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6123125/
https://www.ncbi.nlm.nih.gov/pubmed/29956463
http://dx.doi.org/10.1002/acm2.12402
_version_ 1783352794048102400
author Du, Weiliang
Gao, Song
Jiang, Wei
Kudchadker, Rajat J.
author_facet Du, Weiliang
Gao, Song
Jiang, Wei
Kudchadker, Rajat J.
author_sort Du, Weiliang
collection PubMed
description Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clinac series linacs from Varian, installation of IsoCal is optional and the effects of IsoCal on the imaging systems can be turned on or off after the IsoCal procedure is performed. In this study, we report on the effectiveness of IsoCal in improving the coincidence of the image centers with the radiation isocenter, using an independent Winston‐Lutz (WL) method to locate the radiation isocenter. A ball‐bearing phantom was imaged with 2D MV, 2D kV, and cone beam computed radiography systems on two TrueBeam and two Clinac machines. Using the same phantom, digital WL tests with 16 combinations of gantry and collimator angles were performed to locate the radiation isocenter. The offsets between the IsoCal‐calibrated image centers and the WL radiation isocenter were found to be within 0.4 mm on the four linacs in this study. When IsoCal was turned off, the maximal offsets of the image centers were greater than 1.0 mm on the two Clinac machines. The method developed in this study can be used as a vendor‐independent quality assurance tool to assess the isocentricity of the image centers and radiation central axes.
format Online
Article
Text
id pubmed-6123125
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-61231252018-09-10 Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines Du, Weiliang Gao, Song Jiang, Wei Kudchadker, Rajat J. J Appl Clin Med Phys Radiation Oncology Physics Modern medical linear accelerators (linacs) are often equipped with image guidance systems that are capable of megavolt (MV), kilovolt (kV), planar, or volumetric imaging. On Varian TrueBeam linacs, the isocenter accuracies of the imaging systems are calibrated with a procedure named IsoCal. On Clinac series linacs from Varian, installation of IsoCal is optional and the effects of IsoCal on the imaging systems can be turned on or off after the IsoCal procedure is performed. In this study, we report on the effectiveness of IsoCal in improving the coincidence of the image centers with the radiation isocenter, using an independent Winston‐Lutz (WL) method to locate the radiation isocenter. A ball‐bearing phantom was imaged with 2D MV, 2D kV, and cone beam computed radiography systems on two TrueBeam and two Clinac machines. Using the same phantom, digital WL tests with 16 combinations of gantry and collimator angles were performed to locate the radiation isocenter. The offsets between the IsoCal‐calibrated image centers and the WL radiation isocenter were found to be within 0.4 mm on the four linacs in this study. When IsoCal was turned off, the maximal offsets of the image centers were greater than 1.0 mm on the two Clinac machines. The method developed in this study can be used as a vendor‐independent quality assurance tool to assess the isocentricity of the image centers and radiation central axes. John Wiley and Sons Inc. 2018-06-29 /pmc/articles/PMC6123125/ /pubmed/29956463 http://dx.doi.org/10.1002/acm2.12402 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Du, Weiliang
Gao, Song
Jiang, Wei
Kudchadker, Rajat J.
Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title_full Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title_fullStr Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title_full_unstemmed Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title_short Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines
title_sort independent evaluation of the effectiveness of isocal in improving image center accuracy on varian truebeam and clinac machines
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6123125/
https://www.ncbi.nlm.nih.gov/pubmed/29956463
http://dx.doi.org/10.1002/acm2.12402
work_keys_str_mv AT duweiliang independentevaluationoftheeffectivenessofisocalinimprovingimagecenteraccuracyonvariantruebeamandclinacmachines
AT gaosong independentevaluationoftheeffectivenessofisocalinimprovingimagecenteraccuracyonvariantruebeamandclinacmachines
AT jiangwei independentevaluationoftheeffectivenessofisocalinimprovingimagecenteraccuracyonvariantruebeamandclinacmachines
AT kudchadkerrajatj independentevaluationoftheeffectivenessofisocalinimprovingimagecenteraccuracyonvariantruebeamandclinacmachines