Cargando…

Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices

AIMS: To evaluate whether patients with late complications of pacemakers or implantable cardioverter-defibrillators have hypersensitivity reactions to some of the materials used in generators or in electrodes, or to environmental metal burden. METHODS AND RESULTS: The cohort consisted of 20 men and...

Descripción completa

Detalles Bibliográficos
Autores principales: Maňoušek, Jan, Andršová, Irena, Stejskal, Vera, Vlašínová, Jitka, Sepši, Milan, Kuta, Jan, Klánová, Jana, Mazík, Michal, Jarkovský, Jiří, Šnajdrová, Lenka, Benešová, Klára, Novotný, Tomáš, Zadáková, Andrea, Špinar, Jindřich
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6123938/
https://www.ncbi.nlm.nih.gov/pubmed/29016950
http://dx.doi.org/10.1093/europace/eux227
_version_ 1783352933807554560
author Maňoušek, Jan
Andršová, Irena
Stejskal, Vera
Vlašínová, Jitka
Sepši, Milan
Kuta, Jan
Klánová, Jana
Mazík, Michal
Jarkovský, Jiří
Šnajdrová, Lenka
Benešová, Klára
Novotný, Tomáš
Zadáková, Andrea
Špinar, Jindřich
author_facet Maňoušek, Jan
Andršová, Irena
Stejskal, Vera
Vlašínová, Jitka
Sepši, Milan
Kuta, Jan
Klánová, Jana
Mazík, Michal
Jarkovský, Jiří
Šnajdrová, Lenka
Benešová, Klára
Novotný, Tomáš
Zadáková, Andrea
Špinar, Jindřich
author_sort Maňoušek, Jan
collection PubMed
description AIMS: To evaluate whether patients with late complications of pacemakers or implantable cardioverter-defibrillators have hypersensitivity reactions to some of the materials used in generators or in electrodes, or to environmental metal burden. METHODS AND RESULTS: The cohort consisted of 20 men and 4 women (mean age: 62.3 ± 17.2 years) who had a history of late complications of implanted devices. The control group involved 25 men and 8 women (mean age: 64.6 ± 14.0 years) who had comparable devices, but no history of late complications. Lymphocyte transformation test was used to evaluate hypersensitivity to eight metal pollutants (antimony, manganese, mercury, molybdenum, nickel, platinum, tin, and titanium) selected by results of questionnaires on environmental burden, and by material analysis of generators and electrode surfaces. Exposures to metal pollutants were approximately the same in patients and in controls. Titanium alloy used in generators contained at least 99.32% of titanium and trace levels of other metals; higher levels of tin and platinum were detected in electrode surfaces. Hypersensitivity reactions to mercury and tin were significantly more frequent in patients than in controls (patients and controls: mercury: 68.2 and 31.1%, respectively; P = 0.022; tin: 25.0 and 3.2%, respectively; P = 0.035). In contrast, hypersensitivity to manganese was significantly more frequent in controls than in patients (patients and controls: 13.6 and 50.0%, respectively; P = 0.008). CONCLUSION: Our findings suggest a possible relation between hypersensitivity to metals used in implantable devices or to environmental metal burden and the occurrence of their late complications.
format Online
Article
Text
id pubmed-6123938
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-61239382018-09-10 Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices Maňoušek, Jan Andršová, Irena Stejskal, Vera Vlašínová, Jitka Sepši, Milan Kuta, Jan Klánová, Jana Mazík, Michal Jarkovský, Jiří Šnajdrová, Lenka Benešová, Klára Novotný, Tomáš Zadáková, Andrea Špinar, Jindřich Europace Clinical Research AIMS: To evaluate whether patients with late complications of pacemakers or implantable cardioverter-defibrillators have hypersensitivity reactions to some of the materials used in generators or in electrodes, or to environmental metal burden. METHODS AND RESULTS: The cohort consisted of 20 men and 4 women (mean age: 62.3 ± 17.2 years) who had a history of late complications of implanted devices. The control group involved 25 men and 8 women (mean age: 64.6 ± 14.0 years) who had comparable devices, but no history of late complications. Lymphocyte transformation test was used to evaluate hypersensitivity to eight metal pollutants (antimony, manganese, mercury, molybdenum, nickel, platinum, tin, and titanium) selected by results of questionnaires on environmental burden, and by material analysis of generators and electrode surfaces. Exposures to metal pollutants were approximately the same in patients and in controls. Titanium alloy used in generators contained at least 99.32% of titanium and trace levels of other metals; higher levels of tin and platinum were detected in electrode surfaces. Hypersensitivity reactions to mercury and tin were significantly more frequent in patients than in controls (patients and controls: mercury: 68.2 and 31.1%, respectively; P = 0.022; tin: 25.0 and 3.2%, respectively; P = 0.035). In contrast, hypersensitivity to manganese was significantly more frequent in controls than in patients (patients and controls: 13.6 and 50.0%, respectively; P = 0.008). CONCLUSION: Our findings suggest a possible relation between hypersensitivity to metals used in implantable devices or to environmental metal burden and the occurrence of their late complications. Oxford University Press 2018-09 2017-08-02 /pmc/articles/PMC6123938/ /pubmed/29016950 http://dx.doi.org/10.1093/europace/eux227 Text en © The Author 2017. Published by Oxford University Press on behalf of the European Society of Cardiology http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Clinical Research
Maňoušek, Jan
Andršová, Irena
Stejskal, Vera
Vlašínová, Jitka
Sepši, Milan
Kuta, Jan
Klánová, Jana
Mazík, Michal
Jarkovský, Jiří
Šnajdrová, Lenka
Benešová, Klára
Novotný, Tomáš
Zadáková, Andrea
Špinar, Jindřich
Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title_full Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title_fullStr Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title_full_unstemmed Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title_short Hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
title_sort hypersensitivity to material and environmental burden as a possible cause of late complications of cardiac implantable electronic devices
topic Clinical Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6123938/
https://www.ncbi.nlm.nih.gov/pubmed/29016950
http://dx.doi.org/10.1093/europace/eux227
work_keys_str_mv AT manousekjan hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT andrsovairena hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT stejskalvera hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT vlasinovajitka hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT sepsimilan hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT kutajan hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT klanovajana hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT mazikmichal hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT jarkovskyjiri hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT snajdrovalenka hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT benesovaklara hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT novotnytomas hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT zadakovaandrea hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices
AT spinarjindrich hypersensitivitytomaterialandenvironmentalburdenasapossiblecauseoflatecomplicationsofcardiacimplantableelectronicdevices