Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom

BACKGROUND: Stereotactic Body Radiotherapy (SBRT) is an ablative dose delivery technique which requires the highest levels of precision and accuracy. Modeling dose to a lung treatment volume has remained a complex and challenging endeavor due to target motion and the low density of the surrounding m...

Descripción completa

Detalles Bibliográficos
Autores principales: Mohatt, Dennis J., Ma, Tianjun, Wiant, David B., Islam, Naveed M., Gomez, Jorge, Singh, Anurag K., Malhotra, Harish K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6124019/
https://www.ncbi.nlm.nih.gov/pubmed/30180894
http://dx.doi.org/10.1186/s13014-018-1110-2
_version_ 1783352952960843776
author Mohatt, Dennis J.
Ma, Tianjun
Wiant, David B.
Islam, Naveed M.
Gomez, Jorge
Singh, Anurag K.
Malhotra, Harish K.
author_facet Mohatt, Dennis J.
Ma, Tianjun
Wiant, David B.
Islam, Naveed M.
Gomez, Jorge
Singh, Anurag K.
Malhotra, Harish K.
author_sort Mohatt, Dennis J.
collection PubMed
description BACKGROUND: Stereotactic Body Radiotherapy (SBRT) is an ablative dose delivery technique which requires the highest levels of precision and accuracy. Modeling dose to a lung treatment volume has remained a complex and challenging endeavor due to target motion and the low density of the surrounding media. When coupled together, these factors give rise to pulmonary induced tissue heterogeneities which can lead to inaccuracies in dose computation. This investigation aims to determine which combination of imaging techniques and computational algorithms best compensates for time dependent lung target displacements. METHODS: A Quasar phantom was employed to simulate respiratory motion for target ranges up to 3 cm. 4DCT imaging was used to generate Average Intensity Projection (AIP), Free Breathing (FB), and Maximum Intensity Projection (MIP) image sets. In addition, we introduce and compare a fourth dataset for dose computation based on a novel phase weighted density (PWD) technique. All plans were created using Eclipse version 13.6 treatment planning system and calculated using the Analytical Anisotropic Algorithm and Acuros XB. Dose delivery was performed using Truebeam STx linear accelerator where radiochromic film measurements were accessed using gamma analysis to compare planned versus delivered dose. RESULTS: In the most extreme case scenario, the mean CT difference between FB and MIP datasets was found to be greater than 200 HU. The near maximum dose discrepancies between AAA and AXB algorithms were determined to be marginal (< 2.2%), with a greater variability occurring within the near minimum dose regime (< 7%). Radiochromatic film verification demonstrated all AIP and FB based computations exceeded 98% passing rates under conventional radiotherapy tolerances (gamma 3%, 3 mm). Under more stringent SBRT tolerances (gamma 3%, 1 mm), the AIP and FB based treatment plans exhibited higher pass rates (> 85%) when compared to MIP and PWD (< 85%) for AAA computations. For AXB, however, the delivery accuracy for all datasets were greater than 85% (gamma 3%,1 mm), with a corresponding reduction in overall lung irradiation. CONCLUSIONS: Despite the substantial density variations between computational datasets over an extensive range of target movement, the dose difference between CT datasets is small and could not be quantified with ion chamber. Radiochromatic film analysis suggests the optimal CT dataset is dependent on the dose algorithm used for evaluation. With AAA, AIP and FB resulted in the best conformance between measured versus calculated dose for target motion ranging up to 3 cm under both conventional and SBRT tolerance criteria. With AXB, pass rates improved for all datasets with the PWD technique demonstrating slightly better conformity over AIP and FB based computations (gamma 3%, 1 mm). As verified in previous studies, our results confirm a clear advantage in delivery accuracy along with a relative decrease in calculated dose to the lung when using Acuros XB over AAA.
format Online
Article
Text
id pubmed-6124019
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-61240192018-09-10 Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom Mohatt, Dennis J. Ma, Tianjun Wiant, David B. Islam, Naveed M. Gomez, Jorge Singh, Anurag K. Malhotra, Harish K. Radiat Oncol Research BACKGROUND: Stereotactic Body Radiotherapy (SBRT) is an ablative dose delivery technique which requires the highest levels of precision and accuracy. Modeling dose to a lung treatment volume has remained a complex and challenging endeavor due to target motion and the low density of the surrounding media. When coupled together, these factors give rise to pulmonary induced tissue heterogeneities which can lead to inaccuracies in dose computation. This investigation aims to determine which combination of imaging techniques and computational algorithms best compensates for time dependent lung target displacements. METHODS: A Quasar phantom was employed to simulate respiratory motion for target ranges up to 3 cm. 4DCT imaging was used to generate Average Intensity Projection (AIP), Free Breathing (FB), and Maximum Intensity Projection (MIP) image sets. In addition, we introduce and compare a fourth dataset for dose computation based on a novel phase weighted density (PWD) technique. All plans were created using Eclipse version 13.6 treatment planning system and calculated using the Analytical Anisotropic Algorithm and Acuros XB. Dose delivery was performed using Truebeam STx linear accelerator where radiochromic film measurements were accessed using gamma analysis to compare planned versus delivered dose. RESULTS: In the most extreme case scenario, the mean CT difference between FB and MIP datasets was found to be greater than 200 HU. The near maximum dose discrepancies between AAA and AXB algorithms were determined to be marginal (< 2.2%), with a greater variability occurring within the near minimum dose regime (< 7%). Radiochromatic film verification demonstrated all AIP and FB based computations exceeded 98% passing rates under conventional radiotherapy tolerances (gamma 3%, 3 mm). Under more stringent SBRT tolerances (gamma 3%, 1 mm), the AIP and FB based treatment plans exhibited higher pass rates (> 85%) when compared to MIP and PWD (< 85%) for AAA computations. For AXB, however, the delivery accuracy for all datasets were greater than 85% (gamma 3%,1 mm), with a corresponding reduction in overall lung irradiation. CONCLUSIONS: Despite the substantial density variations between computational datasets over an extensive range of target movement, the dose difference between CT datasets is small and could not be quantified with ion chamber. Radiochromatic film analysis suggests the optimal CT dataset is dependent on the dose algorithm used for evaluation. With AAA, AIP and FB resulted in the best conformance between measured versus calculated dose for target motion ranging up to 3 cm under both conventional and SBRT tolerance criteria. With AXB, pass rates improved for all datasets with the PWD technique demonstrating slightly better conformity over AIP and FB based computations (gamma 3%, 1 mm). As verified in previous studies, our results confirm a clear advantage in delivery accuracy along with a relative decrease in calculated dose to the lung when using Acuros XB over AAA. BioMed Central 2018-09-04 /pmc/articles/PMC6124019/ /pubmed/30180894 http://dx.doi.org/10.1186/s13014-018-1110-2 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Mohatt, Dennis J.
Ma, Tianjun
Wiant, David B.
Islam, Naveed M.
Gomez, Jorge
Singh, Anurag K.
Malhotra, Harish K.
Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title_full Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title_fullStr Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title_full_unstemmed Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title_short Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
title_sort technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6124019/
https://www.ncbi.nlm.nih.gov/pubmed/30180894
http://dx.doi.org/10.1186/s13014-018-1110-2
work_keys_str_mv AT mohattdennisj technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT matianjun technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT wiantdavidb technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT islamnaveedm technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT gomezjorge technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT singhanuragk technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom
AT malhotraharishk technicalanddosimetricimplicationsofrespiratoryinduceddensityvariationsinaheterogeneouslungphantom