Cargando…

Magnetically Controlled Devices Parallel to the Spine in Children with Spinal Muscular Atrophy

BACKGROUND: Children with severe spinal deformity frequently are managed with growth-friendly implants. After initial surgery, externally controlled magnetic rods allow spinal deformity correction during growth without further surgical intervention. The ability to lengthen the spine without addition...

Descripción completa

Detalles Bibliográficos
Autores principales: Lorenz, Heiko M., Badwan, Batoul, Hecker, Marina M., Tsaknakis, Konstantinos, Groenefeld, Katharina, Braunschweig, Lena, Hell, Anna K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6133143/
https://www.ncbi.nlm.nih.gov/pubmed/30229231
http://dx.doi.org/10.2106/JBJS.OA.17.00036
Descripción
Sumario:BACKGROUND: Children with severe spinal deformity frequently are managed with growth-friendly implants. After initial surgery, externally controlled magnetic rods allow spinal deformity correction during growth without further surgical intervention. The ability to lengthen the spine without additional surgical procedures is especially beneficial in high-risk children, such as those with spinal muscular atrophy (SMA). The purpose of the present study was to assess the level of control of spinal deformity in a homogeneous group of patients with SMA who were managed with magnetically controlled implants for 2 years. METHODS: This prospective, nonrandomized study included 21 non-ambulatory children with type-II SMA and progressive scoliosis who were managed bilaterally with a magnetically controlled implant that was inserted parallel to the spine with use of rib-to-pelvis hook fixation. Radiographic measurements of scoliotic curves, kyphosis, lordosis, pelvic obliquity, and spinal length were performed before and after implantation of the magnetically controlled device and during external lengthening. The mean duration of follow-up was 2 years. RESULTS: The mean main curve of patients without prior vertical expandable prosthetic titanium rib (VEPTR) treatment decreased from 70° before implantation of the magnetically controlled device to 30° after implantation of the device. Correction was maintained during the follow-up period, with a mean curve of 31° at the time of the latest follow-up at 2.2 years. Pelvic obliquity was surgically corrected by 76% (from 17° to 4°) and remained stable during follow-up. Thoracic kyphosis could not be corrected within the follow-up period. Spinal length of children without prior spinal surgery increased by >50 mm immediately after device implantation and steadily increased at a rate of 13.5 mm/yr over the course of treatment. During treatment, 4 general complications occurred and 6 lengthening procedures failed, with 3 patients requiring surgical revision. CONCLUSIONS: Bilateral implantation of an externally controlled magnetic rod with rib-to-pelvis fixation represents a safe and efficient method to control spinal deformity in children with SMA, achieving sufficient and stable curve correction as well as increased spinal length. The complication rate was lower than those that have been described for VEPTR and other growing rod instrumentation strategies. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.