Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio

BACKGROUND: Although reverse sequence algorithms (RSA) for syphilis screening are performing well, they still have to rely on treponemal confirmatory tests at least for sera reactive by enzyme immunoassay/chemiluminescence immunoassay (EIA/CIA) and unreactive by rapid plasma reagin (RPR). Quebec’s l...

Descripción completa

Detalles Bibliográficos
Autores principales: Serhir, Bouchra, Labbé, Annie-Claude, Doualla-Bell, Florence, Simard, Marc, Lambert, Gilles, Trudelle, Annick, Longtin, Jean, Tremblay, Cécile, Fortin, Claude
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6136794/
https://www.ncbi.nlm.nih.gov/pubmed/30212544
http://dx.doi.org/10.1371/journal.pone.0204001
_version_ 1783355072355237888
author Serhir, Bouchra
Labbé, Annie-Claude
Doualla-Bell, Florence
Simard, Marc
Lambert, Gilles
Trudelle, Annick
Longtin, Jean
Tremblay, Cécile
Fortin, Claude
author_facet Serhir, Bouchra
Labbé, Annie-Claude
Doualla-Bell, Florence
Simard, Marc
Lambert, Gilles
Trudelle, Annick
Longtin, Jean
Tremblay, Cécile
Fortin, Claude
author_sort Serhir, Bouchra
collection PubMed
description BACKGROUND: Although reverse sequence algorithms (RSA) for syphilis screening are performing well, they still have to rely on treponemal confirmatory tests at least for sera reactive by enzyme immunoassay/chemiluminescence immunoassay (EIA/CIA) and unreactive by rapid plasma reagin (RPR). Quebec’s laboratory network previously showed that 3.3% of EIA/CIA reactive and weakly-reactive RPR samples (RPR titer of 1 to 4) would have been misclassified as syphilis cases if a treponemal confirmatory test had not been performed. OBJECTIVES: To correlate the magnitude of signal-to-cutoff (S/CO) ratios of the 4 most used commercial first-line EIA/CIA kits in Quebec with syphilis confirmation results and establish a S/CO value above which treponemal confirmation would not be required. METHODS: Serum samples from previously undiagnosed individuals (n = 7 404) obtained between January 2014 and February 2017 that were reactive by EIA/CIA and either negative by RPR or reactive with a low titer (1 to 4) were included in the study. All samples were tested with Treponema pallidum particle agglutination (TP-PA) and, if negative or inconclusive, with a line immunoassay (LIA). Syphilis infection confirmation was defined by a reactive TP-PA or LIA. Logistic regression analysis was used to determine S/CO values (95% CI lower bound = 0.98) above which confirmation would not be required. The four kits studied were Architect TP, BioPlex IgG, Syphilis EIA II, and Trep-Sure. RESULTS: Of 2609 reactive EIA/CIA specimens tested for the determination of S/CO values, 1730 (66%) were confirmed as true syphilis cases. Confirmation rate was significantly higher in samples with low-titer positive RPR (92%) than with negative RPR samples (54%); p<0.01. A linear probability model (95% CI lower bound = 0.98) predicted the S/CO value above which a confirmation would no longer be needed for the Architect TP (16.4), Bioplex IgG (7.4) and Trep-Sure (24.6). No linearity was observed between the S/CO value of Syphilis EIA II and the confirmation rate. The validity of the predicted S/CO values was investigated using 4 795 specimens. The use of an S/CO value of 16.4 with the Architect TP kit and of 24.6 for the Trep-Sure kit would obviate the need for confirmation of 18.5% and 13.2% of sera from the all RPR subgroup, respectively. For the BioPlex IgG kit, 81.1% of sera would not require confirmation when using the S/CO value of 7.4 in the low titer RPR subgroup. CONCLUSION: Signal-to-cut-off values could be used to identify sera that do not require extra treponemal confirmation for 3 of the 4 most used first-line EIA/CIA kits in Quebec. Using these values in our current reverse screening algorithm (RSA) would avoid the need for confirmatory tests in 14 to 20% of sera, a proportion that could reach 75% among low-titer RPR.
format Online
Article
Text
id pubmed-6136794
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-61367942018-09-27 Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio Serhir, Bouchra Labbé, Annie-Claude Doualla-Bell, Florence Simard, Marc Lambert, Gilles Trudelle, Annick Longtin, Jean Tremblay, Cécile Fortin, Claude PLoS One Research Article BACKGROUND: Although reverse sequence algorithms (RSA) for syphilis screening are performing well, they still have to rely on treponemal confirmatory tests at least for sera reactive by enzyme immunoassay/chemiluminescence immunoassay (EIA/CIA) and unreactive by rapid plasma reagin (RPR). Quebec’s laboratory network previously showed that 3.3% of EIA/CIA reactive and weakly-reactive RPR samples (RPR titer of 1 to 4) would have been misclassified as syphilis cases if a treponemal confirmatory test had not been performed. OBJECTIVES: To correlate the magnitude of signal-to-cutoff (S/CO) ratios of the 4 most used commercial first-line EIA/CIA kits in Quebec with syphilis confirmation results and establish a S/CO value above which treponemal confirmation would not be required. METHODS: Serum samples from previously undiagnosed individuals (n = 7 404) obtained between January 2014 and February 2017 that were reactive by EIA/CIA and either negative by RPR or reactive with a low titer (1 to 4) were included in the study. All samples were tested with Treponema pallidum particle agglutination (TP-PA) and, if negative or inconclusive, with a line immunoassay (LIA). Syphilis infection confirmation was defined by a reactive TP-PA or LIA. Logistic regression analysis was used to determine S/CO values (95% CI lower bound = 0.98) above which confirmation would not be required. The four kits studied were Architect TP, BioPlex IgG, Syphilis EIA II, and Trep-Sure. RESULTS: Of 2609 reactive EIA/CIA specimens tested for the determination of S/CO values, 1730 (66%) were confirmed as true syphilis cases. Confirmation rate was significantly higher in samples with low-titer positive RPR (92%) than with negative RPR samples (54%); p<0.01. A linear probability model (95% CI lower bound = 0.98) predicted the S/CO value above which a confirmation would no longer be needed for the Architect TP (16.4), Bioplex IgG (7.4) and Trep-Sure (24.6). No linearity was observed between the S/CO value of Syphilis EIA II and the confirmation rate. The validity of the predicted S/CO values was investigated using 4 795 specimens. The use of an S/CO value of 16.4 with the Architect TP kit and of 24.6 for the Trep-Sure kit would obviate the need for confirmation of 18.5% and 13.2% of sera from the all RPR subgroup, respectively. For the BioPlex IgG kit, 81.1% of sera would not require confirmation when using the S/CO value of 7.4 in the low titer RPR subgroup. CONCLUSION: Signal-to-cut-off values could be used to identify sera that do not require extra treponemal confirmation for 3 of the 4 most used first-line EIA/CIA kits in Quebec. Using these values in our current reverse screening algorithm (RSA) would avoid the need for confirmatory tests in 14 to 20% of sera, a proportion that could reach 75% among low-titer RPR. Public Library of Science 2018-09-13 /pmc/articles/PMC6136794/ /pubmed/30212544 http://dx.doi.org/10.1371/journal.pone.0204001 Text en © 2018 Serhir et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Serhir, Bouchra
Labbé, Annie-Claude
Doualla-Bell, Florence
Simard, Marc
Lambert, Gilles
Trudelle, Annick
Longtin, Jean
Tremblay, Cécile
Fortin, Claude
Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title_full Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title_fullStr Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title_full_unstemmed Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title_short Improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
title_sort improvement of reverse sequence algorithm for syphilis diagnosis using optimal treponemal screening assay signal-to-cutoff ratio
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6136794/
https://www.ncbi.nlm.nih.gov/pubmed/30212544
http://dx.doi.org/10.1371/journal.pone.0204001
work_keys_str_mv AT serhirbouchra improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT labbeannieclaude improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT douallabellflorence improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT simardmarc improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT lambertgilles improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT trudelleannick improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT longtinjean improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT tremblaycecile improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio
AT fortinclaude improvementofreversesequencealgorithmforsyphilisdiagnosisusingoptimaltreponemalscreeningassaysignaltocutoffratio