Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry

BACKGROUND: The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom usin...

Descripción completa

Detalles Bibliográficos
Autores principales: Radojcic, Đeni Smilovic, Rajlic, David, Casar, Bozidar, Kolacio, Manda Svabic, Obajdin, Nevena, Faj, Dario, Jurkovic, Slaven
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sciendo 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6137356/
https://www.ncbi.nlm.nih.gov/pubmed/30210046
http://dx.doi.org/10.2478/raon-2018-0019
_version_ 1783355170857418752
author Radojcic, Đeni Smilovic
Rajlic, David
Casar, Bozidar
Kolacio, Manda Svabic
Obajdin, Nevena
Faj, Dario
Jurkovic, Slaven
author_facet Radojcic, Đeni Smilovic
Rajlic, David
Casar, Bozidar
Kolacio, Manda Svabic
Obajdin, Nevena
Faj, Dario
Jurkovic, Slaven
author_sort Radojcic, Đeni Smilovic
collection PubMed
description BACKGROUND: The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom using gamma method. Nevertheless, this method does not replicate the realistic conditions present when a patient is undergoing therapy. Therefore, to more accurately evaluate the treatment planning system (TPS) capabilities, gamma passing rates were examined for beams of different complexity passing through inhomogeneous phantoms. MATERIALS AND METHODS: The research was performed using Siemens Oncor Expression linear accelerator, Siemens Somatom Open CT simulator and Elekta Monaco TPS. A 2D detector array was used to evaluate dose distribution accuracy in homogeneous, semi-anthropomorphic and anthropomorphic phantoms. Validation was based on gamma analysis with 3%/3mm and 2%/2mm criteria, respectively. RESULTS: Passing rates of the complex dose distributions degrade depending on the thickness of non-water equivalent material. They also depend on dose reporting mode used. It is observed that the passing rate decreases with plan complexity. Comparison of the data for all set-ups of semi-anthropomorphic and anthropomorphic phantoms shows that passing rates are higher in the anthropomorphic phantom. CONCLUSIONS: Presented results raise a question of possible limits of dose distribution verification in assessment of plan delivery quality. Consequently, good results obtained using standard patient specific dosimetry methodology do not guarantee the accuracy of delivered dose distribution in real clinical cases.
format Online
Article
Text
id pubmed-6137356
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Sciendo
record_format MEDLINE/PubMed
spelling pubmed-61373562018-09-14 Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry Radojcic, Đeni Smilovic Rajlic, David Casar, Bozidar Kolacio, Manda Svabic Obajdin, Nevena Faj, Dario Jurkovic, Slaven Radiol Oncol Research Article BACKGROUND: The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom using gamma method. Nevertheless, this method does not replicate the realistic conditions present when a patient is undergoing therapy. Therefore, to more accurately evaluate the treatment planning system (TPS) capabilities, gamma passing rates were examined for beams of different complexity passing through inhomogeneous phantoms. MATERIALS AND METHODS: The research was performed using Siemens Oncor Expression linear accelerator, Siemens Somatom Open CT simulator and Elekta Monaco TPS. A 2D detector array was used to evaluate dose distribution accuracy in homogeneous, semi-anthropomorphic and anthropomorphic phantoms. Validation was based on gamma analysis with 3%/3mm and 2%/2mm criteria, respectively. RESULTS: Passing rates of the complex dose distributions degrade depending on the thickness of non-water equivalent material. They also depend on dose reporting mode used. It is observed that the passing rate decreases with plan complexity. Comparison of the data for all set-ups of semi-anthropomorphic and anthropomorphic phantoms shows that passing rates are higher in the anthropomorphic phantom. CONCLUSIONS: Presented results raise a question of possible limits of dose distribution verification in assessment of plan delivery quality. Consequently, good results obtained using standard patient specific dosimetry methodology do not guarantee the accuracy of delivered dose distribution in real clinical cases. Sciendo 2018-04-30 /pmc/articles/PMC6137356/ /pubmed/30210046 http://dx.doi.org/10.2478/raon-2018-0019 Text en © 2018 Đeni Smilovic Radojcic, David Rajlic, Bozidar Casar, Manda Svabic Kolacio, Nevena Obajdin, Dario Faj, Slaven Jurkovic, published by Sciendo http://creativecommons.org/licenses/by-nc-nd/3.0 This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
spellingShingle Research Article
Radojcic, Đeni Smilovic
Rajlic, David
Casar, Bozidar
Kolacio, Manda Svabic
Obajdin, Nevena
Faj, Dario
Jurkovic, Slaven
Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title_full Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title_fullStr Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title_full_unstemmed Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title_short Evaluation of Two-dimensional Dose Distributions for Pre-treatment Patient-specific IMRT Dosimetry
title_sort evaluation of two-dimensional dose distributions for pre-treatment patient-specific imrt dosimetry
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6137356/
https://www.ncbi.nlm.nih.gov/pubmed/30210046
http://dx.doi.org/10.2478/raon-2018-0019
work_keys_str_mv AT radojcicđenismilovic evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT rajlicdavid evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT casarbozidar evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT kolaciomandasvabic evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT obajdinnevena evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT fajdario evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry
AT jurkovicslaven evaluationoftwodimensionaldosedistributionsforpretreatmentpatientspecificimrtdosimetry