Cargando…

Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture

BACKGROUND: The Poly Implant Prothèse (PIP) implants were withdrawn from the market in 2010 due to the use of low-grade silicone, causing a high risk for implant rupture. The aim of this study was to investigate the implant dynamics of PIP breast implants, as well as to determine the rate and predic...

Descripción completa

Detalles Bibliográficos
Autores principales: Bachour, Y., Heinze, Z. C. M., Dormaar, T. S., van Selms, W. G., Ritt, M. J. P. F., Niessen, F. B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6153863/
https://www.ncbi.nlm.nih.gov/pubmed/30294070
http://dx.doi.org/10.1007/s00238-018-1427-y
_version_ 1783357578829365248
author Bachour, Y.
Heinze, Z. C. M.
Dormaar, T. S.
van Selms, W. G.
Ritt, M. J. P. F.
Niessen, F. B.
author_facet Bachour, Y.
Heinze, Z. C. M.
Dormaar, T. S.
van Selms, W. G.
Ritt, M. J. P. F.
Niessen, F. B.
author_sort Bachour, Y.
collection PubMed
description BACKGROUND: The Poly Implant Prothèse (PIP) implants were withdrawn from the market in 2010 due to the use of low-grade silicone, causing a high risk for implant rupture. The aim of this study was to investigate the implant dynamics of PIP breast implants, as well as to determine the rate and predictors of implant gel bleeding, rupture, and capsular contracture in PIP implants. METHODS: Eighty women with a total of 152 PIP implants who underwent a reoperation in 2012 were enrolled in this study. Physical investigation included assessing the Baker score and demographics were retrospectively traced in medical records. The pre- and post-operative volumes of the implants were calculated and their state was determined intraoperatively by the surgeon. RESULTS: The implants were removed after a mean implant duration of 11 ± 2.1 years. Gel bleed and implant rupture occurred in respectively 42 and 25% of the implants. Intact implants had post-operative volume increase as well as decrease. There was a correlation between gel bleeding and more post-operative implant volume increase (P ≤ 0.05). Capsular contracture had a protective effect against post-operative implant volume increase (P ≤ 0.05), while a post-operative implant volume increase provided a protective influence in developing capsular contracture (P ≤ 0.05). Additionally, implant rupture led to a higher risk of capsular contracture (P ≤ 0.05). CONCLUSIONS: We managed to illustrate that PIP implant shells were too permeable and that there is a correlation between gel bleeding and the increase of the post-operative implant volume. Implant rupture led to a higher risk for developing capsular contracture. Level of evidence: Level III, risk / prognostic study. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00238-018-1427-y) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-6153863
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-61538632018-10-04 Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture Bachour, Y. Heinze, Z. C. M. Dormaar, T. S. van Selms, W. G. Ritt, M. J. P. F. Niessen, F. B. Eur J Plast Surg Original Paper BACKGROUND: The Poly Implant Prothèse (PIP) implants were withdrawn from the market in 2010 due to the use of low-grade silicone, causing a high risk for implant rupture. The aim of this study was to investigate the implant dynamics of PIP breast implants, as well as to determine the rate and predictors of implant gel bleeding, rupture, and capsular contracture in PIP implants. METHODS: Eighty women with a total of 152 PIP implants who underwent a reoperation in 2012 were enrolled in this study. Physical investigation included assessing the Baker score and demographics were retrospectively traced in medical records. The pre- and post-operative volumes of the implants were calculated and their state was determined intraoperatively by the surgeon. RESULTS: The implants were removed after a mean implant duration of 11 ± 2.1 years. Gel bleed and implant rupture occurred in respectively 42 and 25% of the implants. Intact implants had post-operative volume increase as well as decrease. There was a correlation between gel bleeding and more post-operative implant volume increase (P ≤ 0.05). Capsular contracture had a protective effect against post-operative implant volume increase (P ≤ 0.05), while a post-operative implant volume increase provided a protective influence in developing capsular contracture (P ≤ 0.05). Additionally, implant rupture led to a higher risk of capsular contracture (P ≤ 0.05). CONCLUSIONS: We managed to illustrate that PIP implant shells were too permeable and that there is a correlation between gel bleeding and the increase of the post-operative implant volume. Implant rupture led to a higher risk for developing capsular contracture. Level of evidence: Level III, risk / prognostic study. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00238-018-1427-y) contains supplementary material, which is available to authorized users. Springer Berlin Heidelberg 2018-06-01 2018 /pmc/articles/PMC6153863/ /pubmed/30294070 http://dx.doi.org/10.1007/s00238-018-1427-y Text en © The Author(s) 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Paper
Bachour, Y.
Heinze, Z. C. M.
Dormaar, T. S.
van Selms, W. G.
Ritt, M. J. P. F.
Niessen, F. B.
Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title_full Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title_fullStr Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title_full_unstemmed Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title_short Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture
title_sort poly implant prothèse silicone breast implants: implant dynamics and capsular contracture
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6153863/
https://www.ncbi.nlm.nih.gov/pubmed/30294070
http://dx.doi.org/10.1007/s00238-018-1427-y
work_keys_str_mv AT bachoury polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture
AT heinzezcm polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture
AT dormaarts polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture
AT vanselmswg polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture
AT rittmjpf polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture
AT niessenfb polyimplantprothesesiliconebreastimplantsimplantdynamicsandcapsularcontracture