Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse

PURPOSE: Cardiac and respiratory motion derived image artefacts are reduced when data are acquired with cardiac and respiratory synchronisation. Where steady state imaging techniques are required in small animals, synchronisation is most commonly performed using retrospective gating techniques but t...

Descripción completa

Detalles Bibliográficos
Autores principales: Kinchesh, Paul, Gilchrist, Stuart, Beech, John S., Gomes, Ana L., Kersemans, Veerle, Newman, Robert G., Vojnovic, Borivoj, Allen, Philip D., Brady, Michael, Muschel, Ruth J., Smart, Sean C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6154312/
https://www.ncbi.nlm.nih.gov/pubmed/29964184
http://dx.doi.org/10.1016/j.mri.2018.06.017
_version_ 1783357671498317824
author Kinchesh, Paul
Gilchrist, Stuart
Beech, John S.
Gomes, Ana L.
Kersemans, Veerle
Newman, Robert G.
Vojnovic, Borivoj
Allen, Philip D.
Brady, Michael
Muschel, Ruth J.
Smart, Sean C.
author_facet Kinchesh, Paul
Gilchrist, Stuart
Beech, John S.
Gomes, Ana L.
Kersemans, Veerle
Newman, Robert G.
Vojnovic, Borivoj
Allen, Philip D.
Brady, Michael
Muschel, Ruth J.
Smart, Sean C.
author_sort Kinchesh, Paul
collection PubMed
description PURPOSE: Cardiac and respiratory motion derived image artefacts are reduced when data are acquired with cardiac and respiratory synchronisation. Where steady state imaging techniques are required in small animals, synchronisation is most commonly performed using retrospective gating techniques but these invoke an inherent time penalty. This paper reports the development of prospective gating techniques for cardiac and respiratory motion desensitised MRI with significantly reduced minimum scan time compared to retrospective gating. METHODS: Prospective gating incorporating the automatic reacquisition of data corrupted by motion at the entry to each breath was implemented in short TR 3D spoiled gradient echo imaging. Motion sensitivity was examined over the whole mouse body for scans performed without gating, with respiratory gating, and with cardio-respiratory gating. The gating methods were performed with and without automatic reacquisition of motion corrupted data immediately after completion of the same breath. Prospective cardio-respiratory gating, with acquisition of 64 k-space lines per cardiac R-wave, was used to enable whole body DCE-MRI in the mouse. RESULTS: Prospective cardio-respiratory gating enabled high fidelity steady state imaging of physiologically mobile organs such as the heart and lung. The automatic reacquisition of data corrupted by motion at the entry to each breath minimised respiratory motion artefact and enabled a highly efficient data capture that was adaptive to changes in the inter-breath interval. Prospective cardio-respiratory gating control enabled DCE-MRI to be performed over the whole mouse body with the acquisition of successive image volumes every 12–15 s at 422 μm isotropic resolution. CONCLUSIONS: Highly efficient cardio-respiratory motion desensitised steady state MRI can be performed in small animals with prospective synchronisation, centre-out phase-encode ordering, and the automatic reacquisition of data corrupted by motion at the entry to each breath. The method presented is robust against spontaneous changes in the breathing rate. Steady state imaging with prospective cardio-respiratory gating is much more efficient than with retrospective gating, and enables the examination of rapidly changing systems such as those found when using DCE-MRI.
format Online
Article
Text
id pubmed-6154312
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-61543122018-11-01 Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse Kinchesh, Paul Gilchrist, Stuart Beech, John S. Gomes, Ana L. Kersemans, Veerle Newman, Robert G. Vojnovic, Borivoj Allen, Philip D. Brady, Michael Muschel, Ruth J. Smart, Sean C. Magn Reson Imaging Article PURPOSE: Cardiac and respiratory motion derived image artefacts are reduced when data are acquired with cardiac and respiratory synchronisation. Where steady state imaging techniques are required in small animals, synchronisation is most commonly performed using retrospective gating techniques but these invoke an inherent time penalty. This paper reports the development of prospective gating techniques for cardiac and respiratory motion desensitised MRI with significantly reduced minimum scan time compared to retrospective gating. METHODS: Prospective gating incorporating the automatic reacquisition of data corrupted by motion at the entry to each breath was implemented in short TR 3D spoiled gradient echo imaging. Motion sensitivity was examined over the whole mouse body for scans performed without gating, with respiratory gating, and with cardio-respiratory gating. The gating methods were performed with and without automatic reacquisition of motion corrupted data immediately after completion of the same breath. Prospective cardio-respiratory gating, with acquisition of 64 k-space lines per cardiac R-wave, was used to enable whole body DCE-MRI in the mouse. RESULTS: Prospective cardio-respiratory gating enabled high fidelity steady state imaging of physiologically mobile organs such as the heart and lung. The automatic reacquisition of data corrupted by motion at the entry to each breath minimised respiratory motion artefact and enabled a highly efficient data capture that was adaptive to changes in the inter-breath interval. Prospective cardio-respiratory gating control enabled DCE-MRI to be performed over the whole mouse body with the acquisition of successive image volumes every 12–15 s at 422 μm isotropic resolution. CONCLUSIONS: Highly efficient cardio-respiratory motion desensitised steady state MRI can be performed in small animals with prospective synchronisation, centre-out phase-encode ordering, and the automatic reacquisition of data corrupted by motion at the entry to each breath. The method presented is robust against spontaneous changes in the breathing rate. Steady state imaging with prospective cardio-respiratory gating is much more efficient than with retrospective gating, and enables the examination of rapidly changing systems such as those found when using DCE-MRI. Elsevier 2018-11 /pmc/articles/PMC6154312/ /pubmed/29964184 http://dx.doi.org/10.1016/j.mri.2018.06.017 Text en © 2018 The Authors http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Kinchesh, Paul
Gilchrist, Stuart
Beech, John S.
Gomes, Ana L.
Kersemans, Veerle
Newman, Robert G.
Vojnovic, Borivoj
Allen, Philip D.
Brady, Michael
Muschel, Ruth J.
Smart, Sean C.
Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title_full Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title_fullStr Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title_full_unstemmed Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title_short Prospective gating control for highly efficient cardio-respiratory synchronised short and constant TR MRI in the mouse
title_sort prospective gating control for highly efficient cardio-respiratory synchronised short and constant tr mri in the mouse
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6154312/
https://www.ncbi.nlm.nih.gov/pubmed/29964184
http://dx.doi.org/10.1016/j.mri.2018.06.017
work_keys_str_mv AT kincheshpaul prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT gilchriststuart prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT beechjohns prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT gomesanal prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT kersemansveerle prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT newmanrobertg prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT vojnovicborivoj prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT allenphilipd prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT bradymichael prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT muschelruthj prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse
AT smartseanc prospectivegatingcontrolforhighlyefficientcardiorespiratorysynchronisedshortandconstanttrmriinthemouse