Development of a Novel Electrochemical Sensor for Determination of Matrine in Sophora flavescens

A simple and sensitive electrochemical sensor fabricated with graphene nanosheets (GNs) and a hydroxyapatite (HA) nanocomposite–modified glassy carbon electrode (GCE) was developed for the determination of matrine (MT). The as-prepared electrode (GNs/HA/GCE) was verified to outperform bare a GCE and...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Junping, Wang, Yanchun, Zheng, Wei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6154538/
https://www.ncbi.nlm.nih.gov/pubmed/28368325
http://dx.doi.org/10.3390/molecules22040575
Descripción
Sumario:A simple and sensitive electrochemical sensor fabricated with graphene nanosheets (GNs) and a hydroxyapatite (HA) nanocomposite–modified glassy carbon electrode (GCE) was developed for the determination of matrine (MT). The as-prepared electrode (GNs/HA/GCE) was verified to outperform bare a GCE and GNs-modified electrode with increased oxidation peak currents and the decreased over-potential in the redox process of MT, indicating the great enhancement of electrocatalytic activity toward the oxidation of MT by the composite of GNs and HA. Under the optimized conditions, the oxidation peak currents were related linearly with the concentration of MT, ranging from 2 μM to 3 mM, and the detection limit (S/N = 3) was 1.2 μM. In addition, the proposed electrochemical sensor can be successfully applied in the quantitative determination of MT in Sophora flavescens extract.