Cargando…
Personalization of injection protocols to the individual patient’s blood volume and automated tube voltage selection (ATVS) in coronary CTA
PURPOSE: The aim was to assess personalised contrast media (CM) protocols—based on patient’s blood volume (BV) and automated tube voltage selection (ATVS)—in coronary computed tomography angiography (CCTA). METHODS: A total of 114 consecutive patients received an ECG-triggered or ECG-gated helical s...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6157854/ https://www.ncbi.nlm.nih.gov/pubmed/30256835 http://dx.doi.org/10.1371/journal.pone.0203682 |
Sumario: | PURPOSE: The aim was to assess personalised contrast media (CM) protocols—based on patient’s blood volume (BV) and automated tube voltage selection (ATVS)—in coronary computed tomography angiography (CCTA). METHODS: A total of 114 consecutive patients received an ECG-triggered or ECG-gated helical scan on a 3(rd)-generation dual-source CT with 70-120kV (ATVS) and 330mAs(qual.ref). CM was adapted to BV, scan time (s) and kV. Image quality (IQ) was assessed in a 17-segment coronary model using attenuation values (HU), contrast-to-noise (CNR), signal-to-noise ratio (SNR) (objective IQ) and a Likert scale (subjective IQ: 1 = poor/2 = sufficient/3 = good/4 = excellent). ig RESULTS: Patient distribution was: n = 60 for 70kV, n = 37 80kV and n = 17 90kV. Mean BV was 5.4±0.6L for men and 4.1±0.6L for women. Mean CM volume (300 mg I/mL) and flow rate were: 30.9±6.4mL and 3.3±0.5mL/s (70kV); 40.8±7.1mL and 4.5±0.6mL/s (80kV); 53.6±8.6mL and 5.7±0.6mL/s (90kV). Overall mean HU was >300HU in 98.2% (112/114) of patients. Overall mean attenuation was below 300HU in two scans (70kV) due to late scan timing. Of 1.661 segments, 95.4% was assessable. Mean CNR was 14±4(70kV), 13±3(80kV) and 14±4(90kV); mean SNR was 10±2(both 70kV+80kV) and 9±2(90kV). Objective IQ was comparable between kV settings, protocols and sex. Subjective IQ was diagnostic in all scans and excellent-sufficient in 95.4% of segments. CONCLUSIONS: Personalisation of CCTA CM injection protocols to BV and ATVS is a promising technique to tailor CM administration to the individual patient, while maintaining diagnostic IQ. |
---|