Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device

The InterStim II (Medtronic, Inc., Minneapolis, MN) sacral nerve stimulator has been approved for MRI scanning of the head only. All other body areas are contraindicated by the manufacturer. This report presents the successful MRI examination of the left hand in a patient with an InterStim II device...

Descripción completa

Detalles Bibliográficos
Autores principales: Fry, Andrew, Kotnis, Nikhil, Edward, Russel, Wright, Peter J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6159144/
https://www.ncbi.nlm.nih.gov/pubmed/30363155
http://dx.doi.org/10.1259/bjrcr.20170064
_version_ 1783358569694887936
author Fry, Andrew
Kotnis, Nikhil
Edward, Russel
Wright, Peter J
author_facet Fry, Andrew
Kotnis, Nikhil
Edward, Russel
Wright, Peter J
author_sort Fry, Andrew
collection PubMed
description The InterStim II (Medtronic, Inc., Minneapolis, MN) sacral nerve stimulator has been approved for MRI scanning of the head only. All other body areas are contraindicated by the manufacturer. This report presents the successful MRI examination of the left hand in a patient with an InterStim II device. Following an assessment of the risks and benefits of proceeding with the scan it was shown that there were minimal additional risks, which could be easily managed with appropriate patient positioning, coil selection and other established techniques. Informed consent was obtained and the scan completed without incident. Following the scan the patient reported full functioning of the device. MRI of the hand is feasible in patients with InterStim II implants using transmit/receive coils with appropriate risk controls in place. Further study of the safety of MRI of other body regions in InterStim II patients is appropriate.
format Online
Article
Text
id pubmed-6159144
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher The British Institute of Radiology
record_format MEDLINE/PubMed
spelling pubmed-61591442018-10-25 Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device Fry, Andrew Kotnis, Nikhil Edward, Russel Wright, Peter J BJR Case Rep Case Report The InterStim II (Medtronic, Inc., Minneapolis, MN) sacral nerve stimulator has been approved for MRI scanning of the head only. All other body areas are contraindicated by the manufacturer. This report presents the successful MRI examination of the left hand in a patient with an InterStim II device. Following an assessment of the risks and benefits of proceeding with the scan it was shown that there were minimal additional risks, which could be easily managed with appropriate patient positioning, coil selection and other established techniques. Informed consent was obtained and the scan completed without incident. Following the scan the patient reported full functioning of the device. MRI of the hand is feasible in patients with InterStim II implants using transmit/receive coils with appropriate risk controls in place. Further study of the safety of MRI of other body regions in InterStim II patients is appropriate. The British Institute of Radiology 2017-10-13 /pmc/articles/PMC6159144/ /pubmed/30363155 http://dx.doi.org/10.1259/bjrcr.20170064 Text en © 2017 The Authors. Published by the British Institute of Radiology http://creativecommons.org/licenses/by/4.0/ This is an open access article under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution and reproduction in any medium, provided the original author and source are credited.
spellingShingle Case Report
Fry, Andrew
Kotnis, Nikhil
Edward, Russel
Wright, Peter J
Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title_full Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title_fullStr Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title_full_unstemmed Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title_short Off-label magnetic resonance imaging of an InterStim II sacral nerve stimulator device
title_sort off-label magnetic resonance imaging of an interstim ii sacral nerve stimulator device
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6159144/
https://www.ncbi.nlm.nih.gov/pubmed/30363155
http://dx.doi.org/10.1259/bjrcr.20170064
work_keys_str_mv AT fryandrew offlabelmagneticresonanceimagingofaninterstimiisacralnervestimulatordevice
AT kotnisnikhil offlabelmagneticresonanceimagingofaninterstimiisacralnervestimulatordevice
AT edwardrussel offlabelmagneticresonanceimagingofaninterstimiisacralnervestimulatordevice
AT wrightpeterj offlabelmagneticresonanceimagingofaninterstimiisacralnervestimulatordevice