Cargando…

Contrast media-assisted in-treatment cone beam CT during single-isocentre volumetric-modulated arc therapy for multiple brain metastases: a case study

A direct visualization technique for verifying intrafractional localization accuracy of multiple brain metastases during single-isocentre volumetric-modulated arc therapy has been proposed using contrast media-assisted in-treatment cone beam CT (CBCT). Contrast-enhanced planning CT images were acqui...

Descripción completa

Detalles Bibliográficos
Autores principales: Nomoto, Akihiro K, Takahashi, Wataru, Yamashita, Hideomi, Haga, Akihiro, Yoda, Kiyoshi, Nakagawa, Keiichi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6159277/
https://www.ncbi.nlm.nih.gov/pubmed/30363248
http://dx.doi.org/10.1259/bjrcr.20160088
Descripción
Sumario:A direct visualization technique for verifying intrafractional localization accuracy of multiple brain metastases during single-isocentre volumetric-modulated arc therapy has been proposed using contrast media-assisted in-treatment cone beam CT (CBCT). Contrast-enhanced planning CT images were acquired immediately after intravenous bolus administration of iodized contrast media at a dose of 2 ml kg(–1). Out of 41 nodules detected on the images, 8 lesions were contoured as high-risk gross tumour volumes (GTVs). Prior to each treatment, CBCT imaging was performed to match the skull structures with the planning CT images. Immediately after another intravenous bolus injection of the iodized contrast media at the same dose as administered for the planning CT imaging, contrast-enhanced CBCT images were acquired during volumetric-modulated arc therapy delivery, thereby providing direct verification of time-averaged tumour position during treatment. The planning target volume contours were overlaid with the in-treatment CBCT images, thereby allowing us to directly visualize the localization accuracy of each GTV when the beam delivery was completed. It was visually confirmed that each GTV was accurately localized inside each planning target volume during beam delivery.