Cargando…

Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment

BACKGROUND: Modern radiotherapy offers the possibility of highly accurate tumor treatment. To benefit from this precision at its best, regular positioning verification is necessary. By the use of image-guided radiotherapy and the application of safety margins the influence of positioning inaccuracie...

Descripción completa

Detalles Bibliográficos
Autores principales: Bell, Katharina, Licht, Norbert, Rübe, Christian, Dzierma, Yvonne
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6167812/
https://www.ncbi.nlm.nih.gov/pubmed/30285806
http://dx.doi.org/10.1186/s13014-018-1141-8
_version_ 1783360263819362304
author Bell, Katharina
Licht, Norbert
Rübe, Christian
Dzierma, Yvonne
author_facet Bell, Katharina
Licht, Norbert
Rübe, Christian
Dzierma, Yvonne
author_sort Bell, Katharina
collection PubMed
description BACKGROUND: Modern radiotherapy offers the possibility of highly accurate tumor treatment. To benefit from this precision at its best, regular positioning verification is necessary. By the use of image-guided radiotherapy and the application of safety margins the influence of positioning inaccuracies can be counteracted. In this study the effect of additional imaging dose by set-up verification is compared with the effect of dose smearing by positioning inaccuracies for a collective of head-and-neck cancer patients. METHODS: This study is based on treatment plans of 40 head-and-neck cancer patients. To evaluate the imaging dose several image guidance scenarios with different energies, techniques and frequencies were simulated and added to the original plan. The influence of the positioning inaccuracies was assessed by the use of real applied table shifts for positioning. The isocenters were shifted back appropriately to these values to simulate that no positioning correction had been performed. For the single fractions the shifted plans were summed considering three different scenarios: The summation of only shifted plans, the consideration of the original plan for the fractions with set-up verification, and the addition of the extra imaging dose to the latter. For both effects (additional imaging dose and dose smearing), plans were analyzed and compared considering target coverage, sparing of organs at risk (OAR) and normal tissue complication probability (NTCP). RESULTS: Daily verification of the patient positioning using 3D imaging with MV energies result in non-negligible high doses. kV imaging has only marginal influence on plan quality, primarily related to sparing of organs at risk, even with daily 3D imaging. For this collective, sparing of organs at risk and NTCP are worse due to potential positioning errors. CONCLUSION: Regular set-up verification is essential for precise radiation treatment. Relating to the additional dose, the use of kV modalities is uncritical for any frequency and technique. Dose smearing due to positioning errors for this collective mainly resulted in a decrease of OAR sparing. Target coverage also suffered from the positioning inaccuracies, especially for individual patients. Taking into account both examined effects the relevance of an extensive IGRT is clearly present, even at the expense of additional imaging dose and time expenditure.
format Online
Article
Text
id pubmed-6167812
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-61678122018-10-09 Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment Bell, Katharina Licht, Norbert Rübe, Christian Dzierma, Yvonne Radiat Oncol Research BACKGROUND: Modern radiotherapy offers the possibility of highly accurate tumor treatment. To benefit from this precision at its best, regular positioning verification is necessary. By the use of image-guided radiotherapy and the application of safety margins the influence of positioning inaccuracies can be counteracted. In this study the effect of additional imaging dose by set-up verification is compared with the effect of dose smearing by positioning inaccuracies for a collective of head-and-neck cancer patients. METHODS: This study is based on treatment plans of 40 head-and-neck cancer patients. To evaluate the imaging dose several image guidance scenarios with different energies, techniques and frequencies were simulated and added to the original plan. The influence of the positioning inaccuracies was assessed by the use of real applied table shifts for positioning. The isocenters were shifted back appropriately to these values to simulate that no positioning correction had been performed. For the single fractions the shifted plans were summed considering three different scenarios: The summation of only shifted plans, the consideration of the original plan for the fractions with set-up verification, and the addition of the extra imaging dose to the latter. For both effects (additional imaging dose and dose smearing), plans were analyzed and compared considering target coverage, sparing of organs at risk (OAR) and normal tissue complication probability (NTCP). RESULTS: Daily verification of the patient positioning using 3D imaging with MV energies result in non-negligible high doses. kV imaging has only marginal influence on plan quality, primarily related to sparing of organs at risk, even with daily 3D imaging. For this collective, sparing of organs at risk and NTCP are worse due to potential positioning errors. CONCLUSION: Regular set-up verification is essential for precise radiation treatment. Relating to the additional dose, the use of kV modalities is uncritical for any frequency and technique. Dose smearing due to positioning errors for this collective mainly resulted in a decrease of OAR sparing. Target coverage also suffered from the positioning inaccuracies, especially for individual patients. Taking into account both examined effects the relevance of an extensive IGRT is clearly present, even at the expense of additional imaging dose and time expenditure. BioMed Central 2018-10-01 /pmc/articles/PMC6167812/ /pubmed/30285806 http://dx.doi.org/10.1186/s13014-018-1141-8 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Bell, Katharina
Licht, Norbert
Rübe, Christian
Dzierma, Yvonne
Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title_full Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title_fullStr Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title_full_unstemmed Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title_short Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
title_sort image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6167812/
https://www.ncbi.nlm.nih.gov/pubmed/30285806
http://dx.doi.org/10.1186/s13014-018-1141-8
work_keys_str_mv AT bellkatharina imageguidanceandpositioningaccuracyinclinicalpracticeinfluenceofpositioningerrorsandimagingdoseontherealdosedistributionforheadandneckcancertreatment
AT lichtnorbert imageguidanceandpositioningaccuracyinclinicalpracticeinfluenceofpositioningerrorsandimagingdoseontherealdosedistributionforheadandneckcancertreatment
AT rubechristian imageguidanceandpositioningaccuracyinclinicalpracticeinfluenceofpositioningerrorsandimagingdoseontherealdosedistributionforheadandneckcancertreatment
AT dziermayvonne imageguidanceandpositioningaccuracyinclinicalpracticeinfluenceofpositioningerrorsandimagingdoseontherealdosedistributionforheadandneckcancertreatment